Huber CS 550-H OD Circulating Bath — Datasheet

Sommaire

  1. Overview p. 1
    Introduces the CS 550-H OD unit and its main features
  2. Spec pp. 1-2
    Lists performance, temperature, and dimensional data per DIN 12876
  3. p. 2 table, 15 rows
  4. p. 1 table, 40 rows

Technical data according to DIN 12876

ParameterValue
max. delivery125 l/min
max. delivery pressure4,7 bar
Pump connectionG1 male
min. filling capacity160 l
Bath volume advance warning175 l
max. filling capacity195 l
Overall dimensions WxDxH **1400x1000x1800 mm
Net weight715 kg
sound pressure level +/- 4 dB(A)79 dB(A)
Power supply (3 Phase)460V 3~ 60Hz
max. current (3 Phase)73 A
Fuse (3 phase)3x90A
min. ambient temperature-20 °C
max. ambient temperature50 °C
from Serial-No.:236124

** Please respect space requirements. See operating conditions at www.huber-online.com

Technical data according to DIN 12876

ParameterValue
Operating temperature range-20...85 °C
temperature set point / displaycolour LED Touchscreen
Internal temperature sensorPt100
Temperature stability at -10°C2 K
Interface digitalRS232
Safety classificationI / NFL
Heating power17 kW
Kälteleistung bei Umgebungstemperatur 5°C
at 15°C61 kW
at 10°C54 kW
at 0°C29 kW
at -10°C15 kW
at -20°C9 kW
Cooling power at ambient temperature 20°C
at 15°C70 kW
at 10°C60 kW
at 0°C33 kW
at -10°C20 kW
at -20°C12 kW
Kälteleistung bei Umgebungstemperatur 35°C
at 15°C63 kW
at 10°C55 kW
at 0°C29 kW
at -10°C16 kW
at -20°C9 kW
Kälteleistung bei Umgebungstemperatur 40°C
at 15°C55 kW
at 10°C49 kW
at 0°C25 kW
at -10°C13 kW
at -20°C7 kW
Kälteleistung bei Umgebungstemperatur 50°C
at 15°C44 kW
at 10°C41 kW
at 0°C19 kW
at -10°C9 kW
Refrigeration machineair-cooled, CFC- and HCFC-free
Refrigerant (ASHRAE, GHS)R-449A (A1, H280)
Refrigerant quantity23 kg
CO2 equivalent32,13 t

Engineering Facts

14 facts · every claim linked to a page
Selection

The chiller's cooling capacity is strongly ambient-dependent, so it must be sized against the actual site ambient temperature rather than a single nominal figure.

Prevents undersizing when the unit is installed in a warm room or outdoors.

Source
“If the ambient temperature rises, the cooling capacity may drop.”
Installation

The unit is rated for outdoor mounting down to -20 °C ambient, which affects enclosure, frost protection and siting decisions.

Allows outdoor installation but requires planning for low ambient conditions.

Source
“Suitable for outside mounting (min. environmental temperature -20°C)”
Installation

Above +35 °C ambient the cooling power is reduced, so the installation location must be checked against the 50 °C maximum.

Defines the ambient threshold above which performance derating begins.

Source
“Reduction of cooling power above +35°C environmental temperature”
Installation

The electrical components are IP54, which sets the minimum protection level expected at the installation site.

Confirms suitability for dusty or splash-exposed outdoor locations.

Source
“Protection classification of the electrical components IP54”
Installation

Three-phase units above 63 A are delivered without cable and without plug, so the customer must provide the full power connection.

Affects electrical installation scope and cost before delivery.

Source
“Three-phase devices with current consumption greater than 63A --> without cable, without plug”
Installation

Voltage and frequency tolerances are coupled, so a -5 % voltage combined with +2 % frequency is not acceptable.

Prevents connecting the chiller to a supply outside the permitted combined tolerance.

Source
“-5% voltage and +2% frequency -> not allowed!”
Installation

The remote control panel is connected via a 20 m cable, which constrains where the operating panel can be located.

Determines cable routing and panel placement during site planning.

Source
“Delivered with remote control: operating panel connected via 20m cable”
Operation

The flow rate is set by a manual bypass valve on the rear of the chiller, so access to the back is needed for adjustment.

Requires rear clearance for commissioning and flow balancing.

Source
“The flow rate can be adjusted via the manual bypass valve on the backside of the chiller.”
Operation

The recommended heat transfer fluid is a 50:50 water-monoethylene glycol mix, which must be used to match the -20 °C range.

Ensures freeze protection and correct viscosity over the operating range.

Source
“Recommended thermofluid: Water - Monoethylene Glycol 50:50”
Operation

The bath has a minimum filling of 160 l and a maximum of 195 l, with a warning at 175 l, defining the usable fluid inventory.

Sets the fluid volume to procure and the level limits for safe operation.

Source
“min. filling capacity”
Safety

The refrigerant is R-449A classified A1 with GHS hazard H280, so leak and pressure safety measures apply.

Informs refrigerant handling, labeling and ventilation requirements.

Source
“R-449A (A1, H280)”
Safety

The unit carries 23 kg of R-449A equal to 32.13 t CO2 equivalent, which triggers refrigerant reporting obligations.

Needed for F-gas/CO2-equivalent registration and leak-check scheduling.

Source
“CO2 equivalent”
Conformité

The equipment is EMC Class A Group 1, so it is intended for industrial environments and may disturb residential areas.

Restricts installation to industrial EMC environments.

Source
“Classification (disturbance) to EN55011: Class A, Group 1”
Performance

The pump delivers up to 125 l/min at 4.7 bar through a G1 male connection, which sizes the external circuit.

Defines the hydraulic capability and connection size for the process loop.

Source
“max. delivery pressure”