KISS 110A Immersion Circulator

Sommaire

  1. Overview p. 1
    Product introduction and model identification
  2. Spec p. 2
    Technical data according to DIN 12876
  3. p. 1 table, 36 rows

KISS 110A Technical data according to DIN 12876

ParameterValue
Operating temperature range25...100 °C
with water cooling20...100 °C
with refrigerator15...100 °C
Temperature stability at 70°C0,05 K
temperature set point / displaydigital
Absolute accuracysetup for calibration
Internal temperature sensorPt100
Interface digitalUSB (Device), RS232
Interface
Alarm messageoptic, acoustic
Safety classificationIII / FL
Heating power at 240V2,1 kW
Heating power at 230V2 kW
Heating power at 220V1,8 kW
Heating power at 208V1,6 kW
Heating power at 200V1,5 kW
max. delivery14 l/min
max. delivery pressure0,25 bar
max. delivery (suction)10,5 l/min
max. delivery pressure (suction)0,17 bar
Pump connenction (optional)M16x1 male
Bath volume7,5 l
min. filling capacity4,4 l
Height of bath opening160 mm
Width bath opening WxD/ bath depth130x310/ 150 mm
Overall dimensions WxDxH **147x507x330 mm
Net weight6 kg
Power supply requirement200-240V 1~/2~ 50/60Hz
max. current10 A
min. Fuse10A
max. Fuse16A
Pressure equipment categoryN/A
Degree of ProtectionIP20
min. ambient temperature5 °C
max. ambient temperature40 °C
from Serial-No.404515

** Overall dimensions WxDxH

Engineering Facts

14 facts · every claim linked to a page
Selection

The usable temperature range depends on the cooling method: 25–100 °C without cooling, 20–100 °C with water cooling, and 15–100 °C with a refrigerator.

Determines whether the bath can reach sub-ambient or near-ambient setpoints and what cooling utility must be provided.

Source
“Operating temperature range 25...100 °C with water cooling 20...100 °C with refrigerator 15...100 °C”
Selection

The bath holds 7.5 l but requires a minimum filling of 4.4 l, so the working volume for samples and fixtures is limited.

Needed to size the bath for the sample load and to avoid running below the minimum fill level.

Source
“Bath volume 7,5 l min. filling capacity 4,4 l”
Installation

The unit draws up to 10 A and requires branch-circuit fusing between 10 A and 16 A on a 200–240 V single/two-phase supply.

Dictates the electrical supply and protective device rating at the installation site.

Source
“max. current 10 A min. Fuse 10A max. Fuse 16A”
Installation

The device is rated IP20, so it must be installed in a dry location protected from water ingress.

Prevents placing the circulator where splashing or condensation could cause electrical hazards.

Source
“Degree of Protection IP20”
Installation

Ambient temperature at the installation site must stay between 5 °C and 40 °C.

Ensures the unit operates within its specified environmental envelope and avoids derating or damage.

Source
“min. ambient temperature 5 °C max. ambient temperature 40 °C”
Safety

The unit is Safety Class III / FL with adjustable overtemperature protection per DIN 12876, providing fail-safe protection for flammable liquids.

Confirms suitability for flammable bath fluids and the required safety category for the application.

Source
“Safety classification III / FL”
Operation

The plastic pump is not resistant to acetone or polyglycols, so these fluids must be avoided or a more resistant plastic must be ordered.

Prevents chemical attack and premature pump failure when selecting the thermofluid.

Source
“Special Case: Acetone and Polyglycol: The plastic pump is not resistant against acetone and polyglycols (depending on the manufacturer).”
Operation

For freeze protection, water should be mixed with glysantine or ethylene glycol rather than used pure.

Avoids freezing and damage when operating near or below ambient with water-based fluid.

Source
“It is recommended that water is mixed with either glysantine or ethylene glycol for freeze protection.”
Installation

An external branch circuit protection according to UL 489 is required.

Mandates a UL-listed breaker at the installation for compliance and safe fault clearing.

Source
“External branch circuit protection according to UL 489 required.”
Installation

Single/two-phase devices are supplied with a power cable and country-specific plug that must be specified at ordering.

Ensures the correct plug type is delivered for the site's electrical outlet.

Source
“Single / two-phase devices (100V to 240V) --> with power cable and country-specific plug (please specify when ordering)”
Performance

The pressure pump delivers up to 14 l/min at 0.25 bar, while the suction pump delivers up to 10.5 l/min at 0.17 bar.

Used to verify the pump can overcome the pressure drop of external circuits and hoses.

Source
“max. delivery 14 l/min max. delivery pressure 0,25 bar max. delivery (suction) 10,5 l/min max. delivery pressure (suction) 0,17 bar”
Performance

Heating power scales with supply voltage from 1.5 kW at 200 V up to 2.1 kW at 240 V.

Affects heat-up time and confirms the actual power available at the site's voltage.

Source
“Heating power at 240V 2,1 kW Heating power at 230V 2 kW Heating power at 220V 1,8 kW Heating power at 208V 1,6 kW Heating power at 200V 1,5 kW”
Conformité

Voltage and frequency tolerances of ±10% are permitted only if they do not act in opposite directions.

Defines the acceptable supply quality for reliable operation and warranty compliance.

Source
“Voltage + / - 10%, as long as the frequency tolerance does not run in the opposite direction.”
Selection

The optional Pt100 sensor connection only displays process temperature and cannot be used for control.

Prevents specifying the external sensor expecting closed-loop control of the process.

Source
“Option: Pt100 sensor connection #10688 to display (not control) e.g. of the process temperature (only available factory fitted, additional charge).”