The usable temperature range depends on the cooling method: 25–100 °C without cooling, 20–100 °C with water cooling, and 15–100 °C with a refrigerator.
Determines whether the bath can reach sub-ambient or near-ambient setpoints and what cooling utility must be provided.
Source
“Operating temperature range 25...100 °C with water cooling 20...100 °C with refrigerator 15...100 °C”
Selection
The bath holds 7.5 l but requires a minimum filling of 4.4 l, so the working volume for samples and fixtures is limited.
Needed to size the bath for the sample load and to avoid running below the minimum fill level.
Source
“Bath volume 7,5 l min. filling capacity 4,4 l”
Installation
The unit draws up to 10 A and requires branch-circuit fusing between 10 A and 16 A on a 200–240 V single/two-phase supply.
Dictates the electrical supply and protective device rating at the installation site.
Source
“max. current 10 A min. Fuse 10A max. Fuse 16A”
Installation
The device is rated IP20, so it must be installed in a dry location protected from water ingress.
Prevents placing the circulator where splashing or condensation could cause electrical hazards.
Source
“Degree of Protection IP20”
Installation
Ambient temperature at the installation site must stay between 5 °C and 40 °C.
Ensures the unit operates within its specified environmental envelope and avoids derating or damage.
Source
“min. ambient temperature 5 °C max. ambient temperature 40 °C”
Safety
The unit is Safety Class III / FL with adjustable overtemperature protection per DIN 12876, providing fail-safe protection for flammable liquids.
Confirms suitability for flammable bath fluids and the required safety category for the application.
Source
“Safety classification III / FL”
Operation
The plastic pump is not resistant to acetone or polyglycols, so these fluids must be avoided or a more resistant plastic must be ordered.
Prevents chemical attack and premature pump failure when selecting the thermofluid.
Source
“Special Case: Acetone and Polyglycol: The plastic pump is not resistant against acetone and polyglycols (depending on the manufacturer).”
Operation
For freeze protection, water should be mixed with glysantine or ethylene glycol rather than used pure.
Avoids freezing and damage when operating near or below ambient with water-based fluid.
Source
“It is recommended that water is mixed with either glysantine or ethylene glycol for freeze protection.”
Installation
An external branch circuit protection according to UL 489 is required.
Mandates a UL-listed breaker at the installation for compliance and safe fault clearing.
Source
“External branch circuit protection according to UL 489 required.”
Installation
Single/two-phase devices are supplied with a power cable and country-specific plug that must be specified at ordering.
Ensures the correct plug type is delivered for the site's electrical outlet.
Source
“Single / two-phase devices (100V to 240V) --> with power cable and country-specific plug (please specify when ordering)”
Performance
The pressure pump delivers up to 14 l/min at 0.25 bar, while the suction pump delivers up to 10.5 l/min at 0.17 bar.
Used to verify the pump can overcome the pressure drop of external circuits and hoses.