KISS 112A Immersion Circulator Datasheet

Sommaire

  1. Overview p. 1
    Product introduction and main features of KISS 112A
  2. Spec p. 2
    Technical data according to DIN 12876
  3. p. 1 table, 37 rows

Technical data according to DIN 12876

ParameterKISS 112A
Order-No.2052.0003.98
Operating temperature range25...100 °C
with water cooling20...100 °C
with refrigerator15...100 °C
Temperature stability at 70°C0,05 K
temperature set point / displaydigital
Absolute accuracysetup for calibration
Internal temperature sensorPt100
Interface digitalUSB (Device), RS232
InterfaceAlarm message
optic, acoustic
Safety classificationIII / FL
Heating power at 240V2,1 kW
Heating power at 230V2 kW
Heating power at 220V1,8 kW
Heating power at 208V1,6 kW
Heating power at 200V1,5 kW
max. delivery14 l/min
max. delivery pressure0,25 bar
max. delivery (suction)10,5 l/min
max. delivery pressure (suction)0,17 bar
Pump connenction (optional)M16x1 male
Bath volume12 l
min. filling capacity7,3 l
Height of bath opening165 mm
Width bath opening WxD/ bath depth275x161/ 150 mm
Overall dimensions WxDxH **333x360x335 mm
Net weight8 kg
Power supply requirement200-240V 1~/2~ 50/60Hz
max. current10 A
min. Fuse10A
max. Fuse16A
Pressure equipment categoryN/A
Degree of ProtectionIP20
min. ambient temperature5 °C
max. ambient temperature40 °C
from Serial-No.401722

** Overall dimensions WxDxH

Engineering Facts

14 facts · every claim linked to a page
Selection

The usable temperature range depends on the cooling method: 25–100 °C without cooling, 20–100 °C with water cooling, and 15–100 °C with a refrigerator.

Determines whether the bath can reach the required low temperature with the available cooling utility.

Source
“Operating temperature range 25...100 °C with water cooling 20...100 °C with refrigerator 15...100 °C”
Installation

The unit requires a 200–240 V single- or two-phase supply at 50/60 Hz and draws up to 10 A, so the branch circuit must be sized accordingly.

Needed to confirm site power availability and breaker sizing before installation.

Source
“Power supply requirement 200-240V 1~/2~ 50/60Hz max. current 10 A”
Installation

An external branch circuit protection device according to UL 489 is mandatory.

Affects the electrical panel design and compliance of the installation.

Source
“External branch circuit protection according to UL 489 required.”
Installation

Voltage and frequency tolerances are ±10% only if they do not oppose each other, e.g. -10% voltage with +3% frequency is not allowed.

Prevents selecting a site with unstable supply that would violate the allowed operating envelope.

Source
“Voltage + / - 10%, as long as the frequency tolerance does not run in the opposite direction.”
Installation

Single- and two-phase devices are delivered with a power cable and country-specific plug, which must be specified at ordering.

Ensures the correct plug type is supplied for the destination country.

Source
“Single / two-phase devices (100V to 240V) --> with power cable and country-specific plug (please specify when ordering)”
Operation

The plastic pump is not resistant to acetone or polyglycols, so these fluids must be avoided or a more resistant plastic option ordered.

Prevents premature pump failure when choosing the heat transfer fluid.

Source
“The plastic pump is not resistant against acetone and polyglycols (depending on the manufacturer).”
Operation

For freeze protection, water should be mixed with glysantine or ethylene glycol rather than using pure water.

Guides fluid selection to avoid freezing damage in low-temperature operation.

Source
“It is recommended that water is mixed with either glysantine or ethylene glycol for freeze protection.”
Safety

The unit is safety class III / FL with adjustable overtemperature protection according to DIN 12876.

Confirms the safety concept and the need to set the overtemperature limit during commissioning.

Source
“With adjustable overtemperature protection according to DIN 12876.”
Safety

The autostart function restarts the unit automatically after a power failure, which must be considered in the risk assessment.

Unattended restart after mains interruption can be hazardous and must be evaluated for the process.

Source
“Autostart function for power failure”
Installation

The enclosure is rated IP20, so it must be installed in a dry location protected from water ingress.

Restricts the installation environment and prevents placing the unit where splashing or condensation occurs.

Source
“Degree of Protection IP20”
Operation

Ambient temperature must stay between 5 °C and 40 °C for specified performance.

Defines the allowable room conditions at the installation site.

Source
“min. ambient temperature 5 °C max. ambient temperature 40 °C”
Performance

The pressure/suction pump delivers up to 14 l/min at 0.25 bar pressure and 10.5 l/min at 0.17 bar suction.

Used to check whether the external circuit hydraulic resistance is compatible with the pump.

Source
“max. delivery 14 l/min max. delivery pressure 0,25 bar”
Selection

The bath holds 12 l total with a minimum filling of 7.3 l, and the bath opening is 275 x 161 mm with 150 mm depth.

Determines fluid volume needed and whether the sample or immersion cooler fits the opening.

Source
“Bath volume 12 l min. filling capacity 7,3 l”
Conformité

The equipment is classified as EN 55011 Class A Group 1, which is intended for industrial environments, not residential.

Affects EMC compliance when installing in non-industrial or sensitive environments.

Source
“Classification (disturbance) to EN55011: Class A, Group 1”