KISS 202C Immersion Circulator Datasheet

Sommaire

  1. Overview p. 1
    Product introduction and key features of KISS 202C
  2. Spec p. 2
    Technical data according to DIN 12876 standard
  3. p. 1 table, 37 rows

Technical data according to DIN 12876

ParameterKISS 202C
Operating temperature range45...200 °C
with water cooling20...200 °C
with refrigerator-30...200 °C
Temperature stability at 70°C0,05 K
temperature set point / displaydigital
Internal temperature sensorPt100
Interface digitalUSB (Device), RS232
Interface
Alarm messageoptic, acoustic
Safety classificationIII / FL
Heating power at 240V2,1 kW
Heating power at 230V2 kW
Heating power at 220V1,8 kW
Heating power at 208V1,6 kW
Heating power at 200V1,5 kW
Pressure / Suction pump
max. delivery14 l/min
max. delivery pressure0,25 bar
max. delivery (suction)10,5 l/min
max. delivery pressure (suction)0,17 bar
Pump connectionM16x1 male
Bath volume3,5 l
Filling capacity2 l
Bath opening diameter25 mm
Bath depth150 mm
Height of bath opening190 mm
Overall dimensions WxDxH **178x260x355 mm
Net weight8 kg
Power supply requirement200-240V 1~/2~ 50/60Hz
max. current10 A
min. Fuse10A
max. Fuse16A
Pressure equipment categoryN/A
Degree of ProtectionIP20
min. ambient temperature5 °C
max. ambient temperature40 °C
from Serial-No.S393769

** Overall dimensions WxDxH

Engineering Facts

14 facts · every claim linked to a page
Selection

The operating temperature range depends on the cooling method chosen, so the required range dictates whether water cooling or a refrigerator is needed.

Determines whether the base unit suffices or a water-cooling loop or refrigerator option must be specified.

Source
“Operating temperature range 45...200 °C with water cooling 20...200 °C with refrigerator -30...200 °C”
Selection

The pump delivers up to 14 l/min with 0.25 bar pressure, which constrains the external circuit resistance the unit can drive.

Needed to check that the external loop pressure drop is compatible with the pump curve.

Source
“max. delivery 14 l/min max. delivery pressure 0,25 bar”
Installation

The unit requires a 200-240 V single- or two-phase supply at 50/60 Hz with a 10 A minimum fuse.

Defines the electrical supply and branch protection that must be provided at the installation site.

Source
“Power supply requirement 200-240V 1~/2~ 50/60Hz max. current 10 A min. Fuse 10A”
Installation

An external branch circuit protection according to UL 489 is mandatory, so the panel must include a listed breaker.

Affects the electrical installation design and compliance for UL jurisdictions.

Source
“External branch circuit protection according to UL 489 required.”
Installation

Voltage and frequency tolerances are coupled: a -10% voltage deviation is only allowed if frequency does not deviate in the opposite direction.

Prevents specifying the unit on a supply where combined voltage and frequency excursions exceed the allowed envelope.

Source
“Voltage + / - 10%, as long as the frequency tolerance does not run in the opposite direction.”
Operation

The plastic pump is not resistant to acetone or polyglycols, so these media must be avoided or a more resistant plastic specified.

Media compatibility must be checked before selecting the unit for a given heat transfer fluid.

Source
“The plastic pump is not resistant against acetone and polyglycols (depending on the manufacturer).”
Operation

For freeze protection, water should be mixed with glysantine or ethylene glycol rather than relying on pure water.

Guides the choice of heat transfer fluid to avoid freezing and damage at low temperatures.

Source
“It is recommended that water is mixed with either glysantine or ethylene glycol for freeze protection.”
Safety

The unit has adjustable overtemperature protection per DIN 12876 and safety class III/FL, which must be configured for the application.

The overtemperature cut-out is a key safety function that must be set correctly during commissioning.

Source
“With adjustable overtemperature protection according to DIN 12876.”
Safety

The autostart function restarts the unit after a power failure, which may be undesirable for unattended or sensitive processes.

Engineers must decide whether automatic restart after mains loss is acceptable for their process.

Source
“Autostart function for power failure”
Installation

The degree of protection is IP20, so the unit must be installed in a dry, protected location away from water spray.

Restricts the installation environment and enclosure requirements.

Source
“Degree of Protection IP20”
Installation

Ambient temperature must stay between 5 °C and 40 °C for correct operation.

Defines the allowable room conditions at the installation site.

Source
“min. ambient temperature 5 °C max. ambient temperature 40 °C”
Selection

The bath holds 3.5 l with a 2 l filling capacity and a 25 mm opening, limiting access and immersion depth.

Determines whether the bath opening and volume suit the samples or external circuit to be connected.

Source
“Bath volume 3,5 l Filling capacity 2 l Bath opening diameter 25 mm”
Conformité

The unit is classified as EN55011 Class A Group 1, which is industrial EMC classification and may not suit residential environments.

Affects EMC compliance and suitability for the intended electromagnetic environment.

Source
“Classification (disturbance) to EN55011: Class A, Group 1”
Installation

Single- and two-phase devices are supplied with a country-specific plug that must be specified at order time.

The plug type must be stated on the order to match the site's socket.

Source
“Single / two-phase devices (100V to 240V) --> with power cable and country-specific plug (please specify when ordering)”