KISS K6s Technical Data

Sommaire

  1. Overview p. 1
    Product introduction and general description of KISS K6s
  2. Spec p. 2
    Technical data according to DIN 12876 standard
  3. p. 2 table, 9 rows
  4. p. 1 table, 37 rows

Technical data according to DIN 12876

ParameterValue
Power supply requirement208-240V 1~/2~ 50/60Hz
max. current immersion thermostat10 A
max. current refrigerated bath1,3 A
min. Fuse10A
max. Fuse16A
Degree of ProtectionIP20
min. ambient temperature5 °C
max. ambient temperature40 °C
from Serial-No.:S395982

1.2/20

Technical data according to DIN 12876 — KISS K6s (Order-No.: 2008.0044.98)

CharacteristicValue
Operating temperature range-25...200 °C
Temperature stability at 70°C0,05 K
temperature set point / displaydigital
Absolute accuracysetup for calibration
Internal temperature sensorPt100
Interface digitalUSB (Device), RS232
InterfaceAlarm message
Alarm messageoptic, acoustic
Safety classificationIII / FL
Heating power at 240V2,1 kW
Heating power at 230V2 kW
Heating power at 220V1,8 kW
Heating power at 208V1,6 kW
Cooling power at 20°C0,26 kW
Cooling power at 0°C0,21 kW
Cooling power at -10°C0,15 kW
Cooling power at -20°C0,05 kW
Refrigeration machineair-cooled, natural
refrigerantRefrigerant (ASHRAE, GHS)
Refrigerant (ASHRAE, GHS)R-290 (A3, H220)
Global Warming Potential (GWP)0,02
Refrigerant quantity0,047 kg
CO2 equivalent0,0 t
UN-numberUN 3358
Pressure / Suction pump max. delivery14 l/min
max. delivery pressure0,25 bar
max. delivery (suction)10,5 l/min
max. delivery pressure (suction)0,17 bar
Pump connectionM16x1 male
Pump connectionNW8/12
Bath volume4,5 l
Width bath opening WxD140x120 mm
Bath depth150 mm
Height of bath opening376 mm
Overall dimensions WxDxH **210x400x546 mm
Net weight25 kg
sound pressure level +/- 4 dB(A)57 dB(A)

** Overall dimensions WxDxH — см. сноску в оригинале; значения приведены дословно.

Engineering Facts

14 facts · every claim linked to a page
Safety

The unit uses flammable refrigerant R-290 (A3, H220), so installation must respect flammable-refrigerant requirements and avoid ignition sources.

Flammable refrigerant classification affects site safety review, ventilation, and ignition-source control during installation.

Source
“Refrigerant (ASHRAE, GHS) R-290 (A3, H220)”
Installation

The refrigerated bath is air-cooled, so it needs adequate air flow and space around the unit; the document explicitly says to respect space requirements.

Air-cooled condensers lose capacity or overheat if installed in tight or poorly ventilated spaces.

Source
“** Please respect space requirements. See operating conditions at www.huber-online.com”
Performance

Cooling capacity is only 0.05 kW at -20 °C, so the unit is not suitable for high heat loads near the low end of its range.

Low-temperature cooling capacity determines whether the bath can pull down and hold temperature with the actual process load.

Source
“at -20°C 0,05 kW”
Performance

Cooling capacity is rated at 20 °C room temperature and may drop if ambient temperature rises.

Selection must account for real lab ambient conditions, not just the catalog cooling curve.

Source
“Output data valid for: Room temperature 20°C. If the ambient temperature rises, the cooling capacity may drop.”
Installation

The allowable ambient temperature range is 5 to 40 °C, so the installation location must stay within this range.

Ambient limits define where the unit can be installed and whether additional room conditioning is needed.

Source
“min. ambient temperature 5 °C max. ambient temperature 40 °C”
Installation

The power supply must be 208-240 V single-phase or two-phase at 50/60 Hz, with a minimum 10 A and maximum 16 A fuse.

Electrical supply and protection sizing must be confirmed before installation.

Source
“Power supply requirement 208-240V 1~/2~ 50/60Hz”
Installation

Voltage and frequency tolerances are linked: ±5% voltage is allowed only with ±2% frequency tolerance, and some combinations are not allowed.

Site power quality must be checked because an apparently acceptable voltage deviation can still be disallowed with the wrong frequency deviation.

Source
“-5% voltage and +2% frequency -> not allowed!”
Operation

The plastic pump is not resistant to acetone or polyglycols, so these fluids must be avoided unless a more resistant plastic option is ordered.

Fluid compatibility determines whether the standard pump can be used or a special version is required.

Source
“The plastic pump is not resistant against acetone and polyglycols (depending on the manufacturer).”
Operation

For freeze protection, the document recommends mixing water with glysantine or ethylene glycol.

This guides heat-transfer-fluid selection and avoids freezing damage during low-temperature operation.

Source
“It is recommended that water is mixed with either glysantine or ethylene glycol for freeze protection.”
Safety

The unit has adjustable overtemperature protection according to DIN 12876 and safety class III / FL.

Overtemperature protection is a key safety requirement for unattended or routine operation.

Source
“With adjustable overtemperature protection according to DIN 12876.”
Operation

The controller has an autostart function for power failure, which means the unit may restart automatically after power returns.

Autostart affects risk assessment and procedures after power interruptions.

Source
“Autostart function for power failure”
Installation

The pump connections are M16x1 male and NW8/12, and included accessories include hose connector NW12 and sleeve nuts thread M16x1.

Connection geometry must match hoses and fittings before installation.

Source
“Pump connection M16x1 male”
Conformité

The equipment is classified for electromagnetic disturbance as EN55011 Class A, Group 1.

Class A equipment may require EMC consideration in residential or sensitive laboratory environments.

Source
“Classification (disturbance) to EN55011: Class A, Group 1”
Installation

The degree of protection is IP20, so the unit is not protected against water ingress and must be kept dry.

IP20 limits where the unit can be installed and rules out washdown or wet environments.

Source
“Degree of Protection IP20”