  1. Overview p. 1
    Product introduction and main features of the Minichiller
  2. Spec pp. 1-2
    Technical data, performance figures and operating limits
  3. Spec p. 2
    Compliance data and parameters according to DIN 12876
  4. p. 2 table, 18 rows
  5. p. 1 table, 23 rows

Technical data according to DIN 12876

ParameterValue
max. delivery (suction)18 l/min
max. delivery pressure (suction)0,4 bar
Pump connectionM16x1 male
Cooling water connectionG1/2 male
min. cooling water differential pressure3 bar
max. cooling water pressure6 bar
min. filling capacity2,8 l
Volume of expansion2,2 l
Overall dimensions WxDxH **280x490x424 mm
Net weight40 kg
Power supply requirement208-240V 1~/2~ 50/60Hz
max. current12 A
Fuse16 A
Pressure equipment categoryArt. 4.3 PED
Degree of ProtectionIP20
min. ambient temperature5 °C
max. ambient temperature40 °C
from Serial-No.:S574334

** Please respect space requirements. See operating conditions at www.huber-online.com

Technical data according to DIN 12876 — Minichiller 1000w-H (Order-No.: 3080.0012.01)

CharacteristicValue
Operating temperature range-20...100 °C
temperature set point / display5,7" colour Touchscreen
Internal temperature sensorPt100
Sensor external connectionPt100
Temperature stability at -10°C0,2 K
Interface digitalEthernet, USB (Host u. Device), RS232
Safety classificationIII / FL
Heating power at 240V2,1 kW
Heating power at 230V2 kW
Heating power at 220V1,8 kW
Heating power at 208V1,6 kW
Cooling power at ambient temperature +20°C at 15°C1 kW
Cooling power at ambient temperature +20°C at 0°C0,8 kW
Cooling power at ambient temperature +20°C at -10°C0,55 kW
Cooling power at ambient temperature +20°C at -20°C0,35 kW
Refrigeration machinewater-cooled, natural refrigerant
Refrigerant (ASHRAE, GHS)R-290 (A3, H220)
Global Warming Potential (GWP)0,02
Refrigerant quantity0,07 kg
CO2 equivalent0,0 t
Circulation pumpPressure- and suction pump
max. delivery24 l/min
max. delivery pressure0,7 bar

Cooling power values are given at ambient temperature +20°C for the respective temperatures 15°C, 0°C, -10°C, -20°C.

Engineering Facts

12 facts · every claim linked to a page
Installation

The chiller requires cooling water with a differential pressure between 3 and 6 bar, so the facility water supply must be sized accordingly.

Insufficient or excessive cooling water pressure will prevent proper operation or damage the unit.

Source
“min. cooling water differential pressure 3 bar max. cooling water pressure 6 bar”
Installation

The unit is water-cooled and must be fed with cooling water between 5°C and 20°C; below the dew point the pipes must be insulated.

Cooling water temperature outside this range reduces cooling capacity or causes condensation.

Source
“This temperature control unit is designed to operate with cooling water intake temperature between 5°C and 20°C.”
Safety

The refrigerant is R-290 (propane, A3, H220), a flammable natural refrigerant, so ignition sources and ventilation must be considered.

Flammable refrigerant classification affects room ventilation and electrical installation requirements.

Source
“R-290 (A3, H220)”
Operation

The plastic pump is not resistant to acetone or polyglycols, so these media must not be used without a more resistant pump option.

Using incompatible media will damage the pump and void reliable operation.

Source
“The plastic pump is not resistant against acetone and polyglycols (depending on the manufacturer).”
Operation

For freeze protection, water should be mixed with glysantine or ethylene glycol rather than relying on pure water.

Prevents freezing and damage in the cooling circuit when operating below ambient freezing conditions.

Source
“It is recommended that water is mixed with either glysantine or ethylene glycol for freeze protection.”
Installation

The electrical supply must be 208–240 V single or two-phase at 50/60 Hz with a maximum current of 12 A and a 16 A fuse.

Determines the site power connection, cable sizing and breaker rating.

Source
“208-240V 1~/2~ 50/60Hz max. current 12 A Fuse 16 A”
Conformité

Voltage and frequency tolerances are ±5% and ±2% respectively, but the combination of -5% voltage with +2% frequency is not allowed.

Ensures the supply quality at the installation site meets the unit's permissible operating envelope.

Source
“-5% voltage and +2% frequency -> not allowed!”
Installation

The unit has IP20 protection, so it must be installed in a dry indoor location protected from water ingress.

IP20 is not water-protected; installation environment must avoid splashing or condensation.

Source
“Degree of Protection IP20”
Operation

Ambient temperature must stay between 5°C and 40°C for reliable operation.

Operating outside this range affects cooling performance and may trip protection.

Source
“min. ambient temperature 5 °C max. ambient temperature 40 °C”
Performance

Cooling power drops from 1 kW at +15°C to 0.35 kW at -20°C, so capacity must be selected for the lowest operating temperature.

Sizing on nominal capacity alone will underperform at low temperatures.

Source
“at 15°C 1 kW at 0°C 0,8 kW at -10°C 0,55 kW at -20°C 0,35 kW”
Installation

The pump connections are M16x1 male and the cooling water connections are G1/2 male, so matching fittings are required.

Ensures correct hose adapters and fittings are ordered before installation.

Source
“Pump connection M16x1 male Cooling water connection G1/2 male”
Maintenance

The cooling water circuit contains copper, stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM, so only compatible cooling water may be used.

Incompatible water chemistry can corrode or foul the wetted materials.

Source
“Materiels used in the cooling water circuit include; copper, Stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM.”