Product name and cover page for the Minichiller unit
Specp. 2
Technical data table according to DIN 12876
p. 2table, 16 rows
Technical data according to DIN 12876
Parameter
Value
Consumption at water 15°C, flow -20°C
24 l/h
Cooling water connection
G1/2 male
min. cooling water differential pressure
3 bar
max. cooling water pressure
6 bar
min. filling capacity
2,8 l
expansion tank
2,2 l
Overall dimensions WxDxH **
280x490x424 mm
Net weight
37 kg
Power supply requirement
208-240V 1~/2~ 50/60Hz
max. current
12 A
Fuse
16 A
Pressure equipment category
Art. 4.3 PED
Degree of Protection
IP20
min. ambient temperature
5 °C
max. ambient temperature
40 °C
from Serial-No.:
S562918
** Please respect space requirements. See operating conditions at www.huber-online.com
Engineering Facts
13 facts · every claim linked to a page
Installation
The chiller requires cooling water with a minimum differential pressure of 3 bar and a maximum pressure of 6 bar, so the facility water supply must be sized accordingly.
Insufficient differential pressure will prevent the water-cooled refrigeration unit from operating correctly.
Source
“min. cooling water differential pressure 3 bar”
Installation
The unit is designed for cooling water intake between 5°C and 20°C, and pipes must be insulated if the water temperature can fall below the condensation point.
Out-of-range cooling water temperature affects cooling capacity and can cause condensation and corrosion.
Source
“This temperature control unit is designed to operate with cooling water intake temperature between 5°C and 20°C.”
Performance
Cooling power drops as cooling water temperature rises, and the required cooling water flow rate may increase, so the water supply must be sized with margin.
Sizing the cooling water circuit only for nominal conditions can lead to insufficient cooling in practice.
Source
“As the cooling water temperature increases, drop in the cooling power should be expected, and also an increased cooling water flow rate possible.”
Safety
The refrigerant is R-290 (propane, A3, H220), a flammable natural refrigerant, which affects installation location and ignition-source restrictions.