Minichiller 1200w OLÉ

Sommaire

  1. Overview p. 1
    Product name and cover page of the datasheet
  2. Spec p. 2
    Technical data according to DIN 12876
  3. p. 2 table, 12 rows

Technical data according to DIN 12876

ParameterValue
expansion tank2,2 l
Overall dimensions WxDxH **280x490x424 mm
Net weight36 kg
Power supply requirement208-240V 1~/2~ 50/60Hz
max. current4 A
min. Fuse10A
max. Fuse16A
Pressure equipment categoryArt. 4.3 PED
Degree of ProtectionIP20
min. ambient temperature5 °C
max. ambient temperature40 °C
from Serial-No.:S611149

** Please respect space requirements. See operating conditions at www.huber-online.com

Engineering Facts

13 facts · every claim linked to a page
Installation

The unit requires cooling water with a differential pressure between 3 and 6 bar, so the facility water supply must be verified before installation.

Insufficient or excessive cooling water pressure will prevent proper operation or damage the unit.

Source
“min. cooling water differential pressure 3 bar”
Installation

Cooling water inlet temperature must be kept between 5°C and 20°C, and pipes must be insulated if condensation conditions occur.

Operating outside this range reduces cooling power and risks condensation inside the unit.

Source
“This temperature control unit is designed to operate with cooling water intake temperature between 5°C and 20°C.”
Performance

Cooling power drops significantly as the setpoint decreases, from 1.2 kW at 15°C to 0.35 kW at -20°C, which must be considered when sizing the load.

The required cooling capacity at the lowest operating temperature determines whether this chiller is suitable.

Source
“at 15°C 1,2 kW”
Safety

The refrigerant is R-290 (propane, A3 flammable), so the installation area must be assessed for flammable refrigerant handling.

Flammable refrigerant classification affects site safety requirements and ventilation.

Source
“R-290 (A3, H220)”
Operation

The plastic pump is not resistant to acetone or polyglycols, so these fluids must not be used as heat transfer media.

Using incompatible fluids will damage the pump and void reliable operation.

Source
“The plastic pump is not resistant against acetone and polyglycols (depending on the manufacturer).”
Operation

For freeze protection, water should be mixed with glysantine or ethylene glycol rather than relying on pure water.

Proper antifreeze selection prevents freezing and protects the cooling circuit.

Source
“It is recommended that water is mixed with either glysantine or ethylene glycol for freeze protection.”
Installation

The electrical supply must be 208-240 V single or two-phase at 50/60 Hz with a fuse rated between 10 A and 16 A.

Correct supply and protection sizing is required for safe electrical installation.

Source
“208-240V 1~/2~ 50/60Hz”
Conformité

Voltage and frequency tolerances must be applied simultaneously: ±5% voltage with ±2% frequency, and certain combinations are not allowed.

Non-compliant supply combinations can cause motor overheating or failure.

Source
“-5% voltage and +2% frequency -> not allowed!”
Installation

The unit has IP20 protection and must be installed in an ambient temperature between 5°C and 40°C.

IP20 offers no water protection, and ambient limits affect reliable operation.

Source
“min. ambient temperature 5 °C”
Operation

The return line temperature must not exceed 80°C even though the operating range is listed as -20...40(80)°C.

Exceeding the return line limit can damage components and is a key process constraint.

Source
“Permissible temperature in return line 80 °C”
Maintenance

The cooling water circuit contains copper, stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM, so only compatible cooling water should be used.

Water chemistry must be compatible with these materials to avoid corrosion and fouling.

Source
“Materiels used in the cooling water circuit include; copper, Stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM.”
Performance

Whisper mode reduces sound pressure to 51 dB(A) but also reduces pump delivery to 14 l/min and 0.2 bar.

Noise-sensitive installations may require whisper mode, which lowers hydraulic performance.

Source
“Delivery rate: 14 l/min”
Installation

The minimum filling capacity is 2.8 l and the expansion tank is 2.2 l, which determines the required heat transfer fluid volume.

Fluid volume and expansion capacity must be planned for correct commissioning.

Source
“min. filling capacity 2,8 l”