Installation
The unit requires cooling water supply with differential pressure between 3 and 6 bar and a G1/2 male connection.
Ensures the facility's cooling water loop can meet the minimum pressure for proper operation.
Source
“min. cooling water differential pressure 3 bar”
Installation
The chiller is water-cooled and must be connected to a cooling water circuit; it is not air-cooled.
Determines whether the installation site needs a cooling water supply at all.
Source
“Chiller with water-cooled refrigerating unit and circulation pump.”
Operation
Cooling water inlet temperature must be kept between 5°C and 20°C, and pipes must be insulated if below the dew point.
Prevents condensation, corrosion, and loss of cooling performance in the water circuit.
Source
“This temperature control unit is designed to operate with cooling water intake temperature between 5°C and 20°C.”
Performance
Cooling power drops as cooling water temperature rises, and the required cooling water flow rate may increase.
Allows proper sizing of the cooling water supply and realistic performance expectations.
Source
“As the cooling water temperature increases, drop in the cooling power should be expected, and also an increased cooling water flow rate possible.”
Safety
The refrigerant is R-290 (propane, A3 flammability class), so ignition sources and ventilation must be considered.
Affects site safety assessment and compliance with flammable refrigerant regulations.
Source
“R-290 (A3, H220)”
Conformité
Voltage tolerance is ±5% only with simultaneous frequency tolerance of ±2%, and opposite-sign deviations are not allowed.
Ensures the power supply quality at the installation site meets the unit's requirements.
Source
“Voltage + / - 5% with a simultaneous frequency tolerance of + / - 2%”
Operation
The plastic pump is not resistant to acetone or polyglycols, so alternative fluids or a more resistant pump must be used.
Prevents chemical attack and premature failure of the circulation pump.
Source
“The plastic pump is not resistant against acetone and polyglycols (depending on the manufacturer).”
Operation
For freeze protection, water should be mixed with glysantine or ethylene glycol rather than relying on pure water.
Avoids freezing and damage in the cooling circuit at low temperatures.
Source
“It is recommended that water is mixed with either glysantine or ethylene glycol for freeze protection.”
Installation
The unit has IP20 protection and must be installed in an ambient temperature between 5°C and 40°C.
Defines the acceptable environmental conditions and protection level at the installation location.
Source
“min. ambient temperature 5 °C”
Installation
The return line temperature must not exceed 80°C even though the operating range is listed as -20...40(80)°C.
Prevents overheating of the unit when used in high-temperature return applications.
Source
“Permissible temperature in return line 80 °C”
Installation
The cooling water circuit contains copper, stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM, so only suitable water may be used.
Guides water quality selection to avoid corrosion and material incompatibility.
Source
“Materiels used in the cooling water circuit include; copper, Stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM.”
Conformité
The equipment emits Class A, Group 1 disturbances per EN55011, which may require EMC measures in residential areas.
Informs EMC planning and whether additional filtering or separation is needed.
Source
“Classification (disturbance) to EN55011: Class A, Group 1”
Installation
Single/two-phase devices are supplied with a country-specific plug that must be specified at ordering.
Ensures the correct power connection is delivered for the destination country.
Source
“with power cable and country-specific plug (please specify when ordering)”