Introduces the Minichiller 800-H unit and its main features
Specpp. 1-2
Performance, temperature, and physical data per DIN 12876
p. 2table, 6 rows
p. 1table, 34 rows
Technical data according to DIN 12876
Parameter
Value
Fuse
16 A
Pressure equipment category
Art. 4.3 PED
Degree of Protection
IP20
min. ambient temperature
5 °C
max. ambient temperature
40 °C
from Serial-No.:
S623474
Minichiller 800-H — Technical data according to DIN 12876
Parameter
Value
Order-No.
3079.0010.01
Operating temperature range
-20...100 °C
temperature set point / display
5,7" colour Touchscreen
Internal temperature sensor
Pt100
Sensor external connection
Pt100
Temperature stability at -10°C
0,2 K
Interface digital
Ethernet, USB (Host u. Device), RS232
Safety classification
III / FL
Heating power at 240V
2,1 kW
Heating power at 230V
2 kW
Heating power at 220V
1,8 kW
Heating power at 208V
1,6 kW
Cooling power at ambient temperature +20°C
at 15°C
0,8 kW
at 0°C
0,6 kW
at -10°C
0,45 kW
at -20°C
0,3 kW
Refrigeration machine
air-cooled, natural refrigerant
Refrigerant (ASHRAE, GHS)
R-290 (A3, H220)
Global Warming Potential (GWP)
0,02
Refrigerant quantity
0,099 kg
CO2 equivalent
0,0 t
Circulation pump:
Pressure- and suction pump
max. delivery
24 l/min
max. delivery pressure
0,7 bar
max. delivery (suction)
18 l/min
max. delivery pressure (suction)
0,4 bar
Pump connection
M16x1 male
min. filling capacity
2,8 l
Volume of expansion
2,2 l
Overall dimensions WxDxH **
280x490x424 mm
Net weight
39 kg
Power supply requirement
208-240V 1~/2~ 50/60Hz
max. current
12 A
** Overall dimensions WxDxH — как напечатано в исходном тексте.
Engineering Facts
12 facts · every claim linked to a page
Selection
The unit uses R-290 (A3, H220) refrigerant, which is flammable, so the installation location must be assessed for A3 refrigerant safety requirements.
Flammable A3 refrigerant affects room ventilation, ignition-source control, and local safety compliance during selection and installation.
Source
“Refrigerant (ASHRAE, GHS) R-290 (A3, H220)”
Installation
The chiller is air-cooled, so sufficient clearance and airflow must be provided around the unit to avoid reduced cooling capacity.
Air-cooled condensers require unobstructed airflow; poor clearance will degrade performance and may cause overheating.
Source
“Chiller with air-cooled refrigerating unit and circulation pump.”
Installation
The power supply must be 208-240 V single-phase or two-phase at 50/60 Hz with a maximum current of 12 A and a 16 A fuse.
Electrical supply, breaker sizing, and cabling must match this requirement before installation.
Source
“Power supply requirement 208-240V 1~/2~ 50/60Hz”
Installation
Voltage and frequency tolerances are ±5% and ±2% respectively, but a -5% voltage combined with +2% frequency is not allowed.
Supply quality must be verified; certain combined tolerances can damage the compressor or electronics.
Source
“-5% voltage and +2% frequency -> not allowed!”
Operation
The ambient temperature must stay between 5 °C and 40 °C, and cooling capacity may drop if ambient temperature rises above 20 °C.
Installation room temperature directly affects available cooling power and must be considered in the thermal design.
Source
“If the ambient temperature rises, the cooling capacity may drop.”
Operation
The plastic pump is not resistant to acetone or polyglycols, so water mixed with glysantine or ethylene glycol should be used for freeze protection.
Fluid compatibility must be checked before commissioning to avoid pump damage and leaks.
Source
“The plastic pump is not resistant against acetone and polyglycols (depending on the manufacturer).”
Installation
Single- and two-phase devices are delivered with a power cable and country-specific plug, which must be specified at ordering.
The correct plug type must be ordered to avoid delays and rework during installation.
Source
“Single / two-phase devices (100V to 240V) --> with power cable and country-specific plug (please specify when ordering)”
Conformité
The equipment is classified as EN55011 Class A, Group 1, so it is intended for industrial electromagnetic environments and may cause interference in residential areas.
Class A equipment has EMC restrictions and may require additional filtering or separation from sensitive electronics.
Source
“Classification (disturbance) to EN55011: Class A, Group 1”
Safety
The unit has IP20 protection, so it must be installed in a dry indoor location protected from water and dust ingress.
IP20 is not suitable for wet or dusty environments; installation location must match the enclosure rating.
Source
“Degree of Protection IP20”
Selection
The minimum filling capacity is 2.8 l and the expansion volume is 2.2 l, which determines the minimum circuit volume and expansion vessel sizing.
External circuit volume and expansion vessel must be sized to avoid overflow or pump cavitation.
Source
“min. filling capacity 2,8 l”
Performance
Cooling capacity drops from 0.8 kW at +15 °C to 0.3 kW at -20 °C, so the required heat load must be checked at the actual operating temperature.
Selecting the chiller only by nominal capacity can lead to insufficient cooling at low temperatures.
Source
“at -20°C 0,3 kW”
Installation
The pump connection is M16x1 male, and hose connection NW12 and sleeve nuts M16x1 are included as accessories.
External piping and hoses must match the M16x1 thread to ensure leak-free connection.