Installation
The chiller requires cooling water with a differential pressure between 3 and 6 bar, so the site water supply must be verified before installation.
Insufficient cooling water differential pressure will prevent the water-cooled refrigeration unit from operating correctly.
Source
“min. cooling water differential pressure 3 bar”
Installation
Cooling water inlet temperature must be kept between 5°C and 20°C, and pipes must be insulated if condensation conditions occur.
Operating outside this cooling water range reduces cooling capacity and risks condensation on uninsulated piping.
Source
“This temperature control unit is designed to operate with cooling water intake temperature between 5°C and 20°C.”
Safety
The refrigerant is R-290 (propane, A3, H220), a flammable natural refrigerant, which affects site safety classification and ventilation requirements.
A3 flammability classification requires appropriate handling, ventilation and ignition-source control during installation and service.
Source
“R-290 (A3, H220)”
Operation
The plastic circulation pump is not resistant to acetone or polyglycols, so these media must not be used in the bath circuit.
Using incompatible media will damage the pump; alternative resistant plastic is only available on request at extra cost.
Source
“The plastic pump is not resistant against acetone and polyglycols (depending on the manufacturer).”
Operation
For freeze protection, water should be mixed with glysantine or ethylene glycol rather than relying on pure water.
Proper antifreeze mixture prevents freezing and damage when operating near or below 0°C.
Source
“It is recommended that water is mixed with either glysantine or ethylene glycol for freeze protection.”
Installation
Voltage and frequency tolerances are ±5% and ±2% respectively, but the combination of -5% voltage with +2% frequency is not allowed.
The power supply must be checked against this combined tolerance rule to avoid operating the unit outside its specified envelope.
Source
“-5% voltage and +2% frequency -> not allowed!”
Installation
The unit is rated IP20, so it must be installed in a dry location protected from water ingress.
IP20 offers no water protection, so washdown or wet environments are unsuitable for this chiller.
Source
“Degree of Protection IP20”
Installation
Ambient temperature at the installation site must remain between 5°C and 40°C.
Ambient conditions outside this range affect performance and may void the specified operating envelope.
Source
“min. ambient temperature 5 °C”
Selection
Cooling power drops with lower temperatures, from 0.8 kW at +15°C to 0.3 kW at -20°C, which must be matched to the process heat load.
Sizing must account for reduced cooling capacity at low setpoints to avoid insufficient heat removal.
Source
“at -20°C 0,3 kW”
Performance
Whisper mode reduces pump delivery to 14 l/min and 0.2 bar, which may be insufficient for high-pressure-drop circuits.
If low noise is required, the reduced pump performance must still satisfy the external circuit pressure drop.
Source
“Delivery rate: 14 l/min”
Conformité
The unit is EMC Class A, Group 1, meaning it is intended for industrial environments and may cause interference in residential areas.
Class A equipment may not be suitable for residential or light-commercial installations without additional EMC measures.
Source
“Classification (disturbance) to EN55011: Class A, Group 1”
Installation
Three-phase devices below 63 A are supplied with cable but without a plug, so a suitable plug must be provided on site.
Planning the electrical connection requires ordering or providing the correct plug for the site supply.
Source
“Three-phase devices with current consumption less than 63A --> with cable, without plug”
Maintenance
The cooling water circuit contains copper, stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM, so only compatible cooling water may be used.
Aggressive or incompatible cooling water can corrode or degrade these materials and cause leaks or fouling.
Source
“Materiels used in the cooling water circuit include; copper, Stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM.”