Table of contents, function description, normal operation and shutdown overview
Overviewpp. 11-21
General introduction, safety notes and device overview
Procedurepp. 22-32
Step-by-step commissioning procedures and initial setup
Overviewpp. 33-43
Functional description of components and operating modes
Procedurepp. 44-61
Setup mode procedures and parameter configuration
Procedurepp. 62-63
Operating procedures during normal use
Procedurepp. 64-70
Interface connections and software update procedures
Procedurepp. 71-78
Maintenance schedules, service tasks and troubleshooting
Procedurepp. 79-82
Shutdown procedures and decommissioning steps
pp. 83-84
Technical annex and temperature range data
p. 9table, 40 rows
p. 42table, 21 rows
p. 36table, 17 rows
p. 34table, 10 rows
p. 26table, 6 rows
p. 35table, 4 rows
p. 70table, 6 rows
p. 8table, 40 rows
Содержание (оглавление) страницы 9
Раздел
Название
Страница
4.2.5
Monitoring the Pt100 temperature sensors
49
4.2.6
Optimum control parameters for optimum temperature control
50
4.2.7
Sub-category: “Select auto/expert mode”.
50
4.2.8
Sub-category: “Configuration auto”
50
4.2.8.1
Sub-category: “Find parameters”
50
4.2.8.2
Sub-category: “Control Dynamics”
52
4.2.8.3
Sub-category: “Fluid Properties”
53
4.2.8.4
Sub-category: “Display parameters”
54
4.2.9
Sub-category: “Configuration manual”
55
4.2.9.1
Sub-category: “Change parameters”
55
4.2.9.2
Sub-category: “Display parameters”
56
4.2.9.3
Sub-category: “Control structure”
56
4.2.10
Sub-category: “Reset parameters”
56
4.2.11
Sub-category: “Display parameters”
57
4.2.12
Setting the setpoint thresholds
57
4.2.13
Setting the setpoint
58
4.3
Filling, venting, degassing and draining
58
4.3.1
Externally open application
58
4.3.1.1
Filling and venting externally open application
58
4.3.1.2
Degassing externally open applications
59
4.3.1.3
Draining externally open applications
60
5
Normal operation
62
5.1
Automatic operation
62
5.1.1
Temperature control
62
5.1.1.1
Starting the temperature control process
62
5.1.1.2
Ending the temperature control process
62
5.1.2
Temperature control via a created temperature control program
63
5.1.2.1
Starting the temperature control program
63
5.1.2.2
Ending/cancelling the temperature control program
63
6
Interfaces and software update
64
6.1
Interfaces at the “Pilot ONE®” controller
64
6.1.1
10/100 Mbps Ethernet for RJ45 network sockets
64
6.1.2
USB-2.0 interface
64
6.1.2.1
USB-2.0 interface, host
65
6.1.2.2
USB-2.0 interface, device
65
6.2
Interfaces on the temperature control unit
65
6.2.1
Interfaces at the back
65
6.2.1.1
Connector POKO (floating contact) alarm
65
6.2.1.2
Jack ECS (External Control Signal) standby
66
6.2.1.3
Service interface
67
На странице приведено оглавление (содержание), а не матрица конфигураций; данные приведены как напечатано.
Restoring the factory settings
Description
(A)
(B)
(C)
(D)
(E)
(F)
Acoustics
X
X
X
–
–
–
Display functions Warnings; Messages; Inactive Menu Items; Resolution Temperature
X
X
X
–
–
–
Display modes
X
X
X
–
–
–
AutoStart response
X
X
X
–
–
–
Limitations DeltaT limit; max. heating variable; max. cooling variable; max. permissible current consumption (for 230 V temperature control units)
X
X
X
–
–
–
Com.G@te (only if Com.G@te is connected) Analog Interface (configuration input) Meaning of AIF input; response on cable break; measurement range setting
X
X
X
–
–
X
Calibration
–
–
–
–
–
X
In case of analog error; current / voltage switching; filter constant
X
X
X
–
–
X
Configuration Output AIF output variable; Output range setting
Pump settings Pump speed setpoint; Pump pressure setpoint; Control Mode
X
X
X
–
–
–
Fluid properties Thermal fluid; liter indicator
X
X
X
–
–
–
Configuration manual parameters Protective function High Limit Alarm Internal; Low Limit Alarm Internal; High Limit Alarm Process; Low Limit Alarm Process; Hydrostatic Correction; Warning Time Limit (CC-E only); Min. Level (for temperature control units with analogue level sensor); Max. Level (for temperature controllers with analogue level sensor); Freeze Protection (optional)
X
X
X
–
–
–
Setpoint
X
X
X
–
–
–
Setpoint limitation Min. Setpoint; Max. Setpoint
X
X
X
–
–
–
Language
X
X
X
–
–
–
“X” = Value is reset, “–” = Value is not reset. (A) = All together; (B) = Device parameters without OT; (C) = Device Parameter; (D) = Program Controller; (E) = Menu; (F) = Com.G@te
Overview of E-grade functions
Function
E-grade Basic
E-grade Exclusive
E-grade Professional
Temperature control
Controller parameterization: predefined1 / TAC2
X/–
–/X
–/X
Find parameters: Fast Identification / With Preliminary Test / Estimate Control Parameters
–/–/X
X/X/–
X/X/–
Sensor calibration for control sensor3: x -point
2
5
5
Monitoring: Low level and overtemperature4
X
X
X
Adjustable alarm thresholds
X
X
X
VPC5 (Variable Pressure Control)
X
X
X
Ventilation program
X
X
X
Automated compressor
X
X
X
Setpoint limitation
X
X
X
Program controller: x programs / max x steps
–/–
3/15
10/100
Ramp function: linear / non-linear
–/–
X/–
X/X
Temperature control mode: Internal and process
–
X
X
Adjustable max. heating and cooling capacity
–
X
X
Display & operation
Temperature display: 5.7" Touchscreen
X
X
X
Display mode: graphically / numerically large / Explore
–/X/–
–/X/–
X/X/–
X = Standard equipment, O = optional, – = not possible. 1 TAC function available as a 30 day evaluation version. 2 True Adaptive Control. 3 Internal Pt100 and External Pt100. 4 For temperature control units with integrated overtemperature protection. 5 For units with variable-speed pump or external bypass.
Table
Designation
Specification
Calcium carbonate per liter
≤ 1.5 mmol/l; corresponds to a water hardness of: ≤ 8.4 °dH (soft)
pH value
between 6.0 and 8.5
Ultrapure water, distillates
Add 0.1 g of sodium carbonate (Na2CO3) per liter
Non-approved water
Distilled, deionized, demineralized, chloric, ferruginous, ammoniacal, contaminated or untreated river water or sea water
Volume circulated (at least)
3 l/min. (not valid for cooling baths)
Thermal fluid: Water without ethylene glycol
Use
excluded
Thermal fluid: Water-ethylene glycol mixture
Use
-35 °C (with anti-free possible up to -38 °C) to +95 °C
Thermal fluid composition
The mixture’s temperature must be 10 K below the permissible min. temperature. For the permissible temperature range, refer to the datasheet.→ From page 83, section »Annex«.
Side / Distance in cm (for operation in a tub)
Side
Air cooling
Water cooling
[A1] Top
Air outlet on top of unit: free standing
–
[A2] Top
can be located under a bench
can be located under a bench
[B] Left
min. 20
min. 20
[C] Right
min. 20
min. 20
[D] Front
min. 20
min. 20
[E] Rear
min. 20
min. 20
Overview of the E-grade variants
Temperature control units / E-grade
E-grade Basic
E-grade Exclusive
E-grade Profes-sional
Brewing thermostats
–
X
O
Unistat temperature control units
–
–
X
UniCAL
–
–
X
Other temperature control units
X
O
O
Pin assignment
Pin
Description
Signal
1
Current output, T external
0 a)/4 - 20 mA or 0 - 10 V If using 0 - 10 V, integrate a 500 Ω resistor.
2
Current output, setpoint
0 a)/4 - 20 mA or 0 - 10 V
3
GND for analog outputs
GND
4
Analog input (programmable)
0 a)/4 - 20 mA or 0 - 10 V Power input: 200 Ω working resistance Voltage input: 100 kΩ input resistance
5
Current output, freely programmable
0 a)/4 - 20 mA or 0 - 10 V If using 0 - 10 V, integrate a 500 Ω resistor.
6
GND for analog input
GND
a) Consult with our Customer Support.
Содержание (оглавление)
Раздел
Название
Страница
2.10
Connecting to the power supply
31
2.10.1
Connection using socket with protective earth (PE)
31
2.10.2
Connection via hard wiring
32
3
Function description
33
3.1
Function description of the temperature control unit
33
3.1.1
General functions
33
3.1.2
Other functions
33
3.2
Information on the thermal fluids
33
3.3
To be noted when planning the test
34
3.4
“Pilot ONE®” controller
35
3.4.1
Functional overview of “Pilot ONE®”
35
3.5
Clock/event function
38
3.5.1
Rechargeable accumulator
38
3.5.2
Programmable event function
38
3.5.2.1
Event function “Alarm clock event”
38
3.5.2.2
Event function “Program event”
38
3.6
Operation via the touch screen
38
3.7
Display instruments
38
3.7.1
The touchscreen [88]
39
3.8
Control instruments
39
3.8.1
The touchbuttons
39
3.8.2
The categories
39
3.8.3
The sub-categories
39
3.8.4
The dialogs
40
3.9
Function examples
40
3.9.1
Display of software version
40
3.9.2
Start & Stop
40
3.9.3
Copying the settings to a data carrier
41
3.9.3.1
Saving to a USB flash drive
41
3.9.3.2
Loading from a USB flash drive
41
3.9.4
Restore factory settings
41
3.9.4.1
Restore to factory settings without overtemperature protection
43
3.9.4.2
Restore to factory settings including overtemperature protection
43
4
Setup mode
44
4.1
Setup mode
44
4.1.1
Turning on the temperature control unit
44
4.1.2
Turning off the temperature control unit
44
4.1.3
Setting the overtemperature protection
45
4.1.3.1
General information on the overtemperature protection
45
4.1.3.2
Setting “OT limit: heating”
45
Engineering Facts
13 facts · every claim linked to a page
Safety
The unit must not be installed or started in an ATEX zone without an Ex px cabinet.
Prevents explosion risk and defines the permissible hazardous-area installation scope.
Source
“Do NOT install or start up the temperature control unit within an ATEX zone.”
Installation
For natural-refrigerant units, if the charge exceeds 8 g per m³ of room air, a calibrated gas warning sensor must be installed and maintained every 6–12 months.
Determines room-size limits and mandatory gas detection for NR units.
Source
“Over 8 g refrigerant per m³ room air: A gas warning sensor must be fitted and functioning.”
Installation
Units with more than 1 kg of refrigerant must not be installed below ground level.
Constrains basement or pit installation for larger refrigerant charges.
Source
“Temperature control units with more than 1 kg refrigerant must not be installed below ground level (GL).”
Installation
The power cable must not exceed 3 m so the unit can be disconnected from mains at any time.
Affects electrical installation layout and safe disconnection.
Source
“Do not use a power cable that is longer than 3 m to be able to disconnect the temperature control unit at any time from the mains.”
Installation
For externally open applications, circulation flow must go to the lower connection point and circulation return to the higher point.
Correct piping orientation is required to vent air and operate the application properly.
Source
“the >Circulation flow< [1] connection from the temperature control unit is attached to the lower connection point [B] of the application and the >Circulation return< [2] connection is attached to the higher connection point [A] of the application.”
Installation
Recommended and maximum torques are specified per connection size, e.g. 20 Nm recommended and 24 Nm maximum for M16x1.
Prevents leaks and thread damage when tightening pump connections.
Source
“Recommended torques in Nm”
Operation
The thermal fluid viscosity must not exceed 50 mm²/s at the lowest working temperature and density must not exceed 1 kg/dm³.
Defines fluid selection limits for pump and heat transfer performance.
Source
“The maximum viscosity of the thermal fluid must not exceed 50 mm²/s at the lowest working temperature!”
Operation
Different thermal fluid types must not be mixed; the circuit must be rinsed when changing fluid type.
Prevents property damage and maintains fluid compatibility.
Source
“Do not mix different types of thermofluid (such as mineral oil, silicone oil, synthetic oil, water, etc.) in a thermofluid circuit.”
Operation
Cooling water supply should be shut off even for short downtimes such as overnight.
Reduces risk of condensation and water damage during idle periods.
Source
“Shut off the cooling water supply to the temperature control unit even for shorter downtimes (e.g. overnight).”
Safety
Electrical lines directly below a water-cooled unit must be protected against liquid ingress because condensation can form below the dew point.
Prevents electrical shock hazard from condensation at cooling water connections.
Source
“Electrical lines directly below the temperature control unit must be protected against liquid ingress.”
Maintenance
The air filter mat behind the air inlet connection must be checked and replaced at regular intervals to avoid reduced airflow.
Maintains cooling performance and prevents overheating due to blocked airflow.
Source
“This air filter mat must be checked / replaced at regular intervals so that the air flowing into the temperature control unit is not reduced.”
Conformité
The responsible body must keep F-gas records for at least 5 years and present them to authorities on request.
Defines documentation obligations for units with fluorinated refrigerants.
Source
“The responsible body must keep the records for at least 5 years after their creation and present it to the responsible authority upon request.”
Selection
The unit is classified to DIN 12876 with overheat protection for non-combustible media (NFL) or adjustable overheat/overtemperature plus low-level protection for combustible media (FL).
Determines the safety class and suitable thermal fluid for the application.
Source
“Adjustable overtemperature protection and additional low-level protection”