TC100E Immersion Circulator

Sommaire

  1. Overview p. 1
    Product identification and cover page for TC100E circulator
  2. Spec p. 2
    Specifications per DIN 12876 including temperature and performance data
  3. p. 1 table, 36 rows

Technical data according to DIN 12876

ParameterValue
Operating temperature range-100...40 °C
Temperature stability at -10°C0,5 K
temperature set point / displaydigital
Sensor external connectionPt100
Cooling power
at 0°C0,16 kW
at -20°C0,15 kW
at -30°C0,14 kW
at -50°C0,13 kW
at -60°C0,12 kW
at -80°C0,12 kW
at -90°C0,07 kW
at -100°C0,01 kW
Safety classificationI / NFL
Refrigeration machineair-cooled, CFC- and HCFC-free
Refrigerant (ASHRAE, GHS)R-452A (A1, H280)
Global Warming Potential (GWP)2141
Refrigerant quantity0,26 kg
CO2 equivalent0,6 t
Refrigerant 2nd stage (ASHRAE, GHS)R-1150 (A3, H220)
Global Warming Potential (GWP)4
Refrigerant quantity 2nd stage0,056 kg
CO2-Äquivalent 2.Stufe0,0 t
Nominal diameter probe48 mm
Length of probe155 mm
Length flexible connection1150 mm
Overall dimensions WxDxH **295x500x570 mm
Net weight60 kg
sound pressure level +/- 4 dB(A)55 dB(A)
Power supply requirement110-120V 1~ 60Hz
max. current14 A
Fuse16 A
Pressure equipment categoryArt. 4.3 PED
min. ambient temperature5 °C
max. ambient temperature40 °C
from Serial-No.S528889

** Overall dimensions WxDxH

Engineering Facts

12 facts · every claim linked to a page
Selection

The unit's operating temperature range is -100 to 40 °C, so it must be matched to the process temperature window.

Defines whether the cooler can cover the required process temperatures before selection.

Source
“Operating temperature range -100...40 °C”
Performance

Cooling capacity drops sharply at the cold end (0.01 kW at -100 °C), so low-temperature duty needs careful capacity sizing.

Prevents oversizing assumptions and ensures the process heat load is removable at the coldest set point.

Source
“at -100°C 0,01 kW”
Installation

The unit requires a 110-120 V 1~ 60 Hz supply with 14 A max current and a 16 A fuse.

Determines site power provisioning, breaker/fuse rating and cable selection.

Source
“Power supply requirement 110-120V 1~ 60Hz max. current 14 A Fuse 16 A”
Installation

Ambient temperature must stay between 5 °C and 40 °C, and cooling capacity may fall if ambient rises.

Room conditions must be verified at the installation site to avoid derating or tripping.

Source
“min. ambient temperature 5 °C max. ambient temperature 40 °C”
Installation

Voltage and frequency tolerances are coupled: ±5% voltage only with ±2% frequency, and -5% voltage with +2% frequency is not allowed.

Prevents installation on unstable or mismatched mains that could damage the compressor.

Source
“-5% voltage and +2% frequency -> not allowed!”
Safety

The refrigerant is R-452A (A1, H280) with GWP 2141 and 0.26 kg charge, giving 0.6 t CO2 equivalent.

Affects refrigerant handling, leak response and F-gas/CO2-equivalent reporting obligations.

Source
“Refrigerant (ASHRAE, GHS) R-452A (A1, H280) Global Warming Potential (GWP) 2141 Refrigerant quantity 0,26 kg CO2 equivalent 0,6 t”
Safety

The second refrigerant stage is R-1150 (A3, H220), a flammable refrigerant, so ignition-source control is required.

A3 classification imposes flammable-refrigerant installation and ventilation precautions.

Source
“Refrigerant 2nd stage (ASHRAE, GHS) R-1150 (A3, H220)”
Operation

The refrigeration unit runs continuously and temperature is controlled by a solenoid valve in the cooling circuit.

Explains control behaviour and why the compressor is not cycled on/off during normal operation.

Source
“The refrigeration unit works continuously. The temperature control actuates a solenoid valve in the cooling circuit.”
Installation

The probe is 48 mm in diameter and 155 mm long with a 1150 mm flexible connection, which constrains vessel port and routing.

Needed to confirm the immersion port size and the reach/routing of the cooling line.

Source
“Nominal diameter probe 48 mm Length of probe 155 mm Length flexible connection 1150 mm”
Conformité

The equipment is EMC Class A, Group 1, so it is intended for industrial, not residential, environments.

Class A equipment may cause interference in residential areas and must be sited accordingly.

Source
“Classification (disturbance) to EN55011: Class A, Group 1”
Installation

Single/two-phase devices are supplied with a power cable and country-specific plug that must be specified at order.

Avoids a missing or wrong plug on delivery and delays at commissioning.

Source
“Single / two-phase devices (100V to 240V) --> with power cable and country-specific plug (please specify when ordering)”
Conformité

The unit is compliant with US-SNAP for industrial process refrigeration and with applicable EU laws.

Confirms refrigerant acceptability for the intended end-use in regulated markets.

Source
“This equipment is compliant to US-SNAP and all applicable EU laws. The US-SNAP end-use for this equipment is the industrial process refrigeration.”