Unichiller 015w-H Recirculating Chiller Datasheet

Sommaire

  1. Overview p. 1
    Introduces the Unichiller 015w-H chiller and its main features
  2. Spec pp. 1-2
    Lists performance, temperature, and dimensional data per DIN 12876
  3. p. 2 table, 11 rows
  4. p. 1 table, 33 rows

Technical data according to DIN 12876

ParameterValue
Volume of expansion1,7 l
Overall dimensions WxDxH **350x496x622 mm
Net weight57 kg
sound pressure level +/- 4 dB(A)54 dB(A)
Power supply requirement220-240V 1~ 50Hz
max. current13 A
Fuse16 A
Degree of ProtectionIP20
min. ambient temperature5 °C
max. ambient temperature40 °C
from Serial-No.:1.1/24

** Please respect space requirements. See operating conditions at www.huber-online.com

Technical data according to DIN 12876 — Unichiller 015w-H (Order-No.: 3051.0010.01)

CharacteristicValue
Operating temperature range-20...100 °C
temperature set point / display5,7" colour Touchscreen
Internal temperature sensorPt100
Sensor external connectionPt100
Temperature stability at -10°C0,2 K
Interface digitalEthernet, USB (Host u. Device), RS232
Safety classificationIII / FL
Heating power at 240V2,1 kW
Heating power at 230V2 kW
Heating power at 220V1,8 kW
Cooling power at ambient temperature +20°C at 15°C1,5 kW
Cooling power at ambient temperature +20°C at 0°C1 kW
Cooling power at ambient temperature +20°C at -10°C0,7 kW
Cooling power at ambient temperature +20°C at -20°C0,3 kW
Refrigeration machinewater-cooled, CFC- and HCFC-free
Refrigerant (ASHRAE, GHS)R-449A (A1, H280)
Global Warming Potential (GWP)1397
Refrigerant quantity0,25 kg
CO2 equivalent0,3 t
UN-numberUN 2857
Circulation pumpImmersion pump
max. delivery29 l/min
max. delivery pressure1 bar
Delivery at 0,5 bar21 l/min
Pump connectionG3/4 male
Cooling water connectionG1/2 male
Consumption at water 15°C, flow 15°C68 l/h
Consumption at water 15°C, flow 0°C60 l/h
Consumption at water 15°C, flow -10°C48 l/h
Consumption at water 15°C, flow -20°C36 l/h
min. cooling water differential pressure1 bar
max. cooling water pressure6 bar
min. filling capacity3,8 l

Cooling power values are given at ambient temperature +20°C for the respective flow temperatures (15°C, 0°C, -10°C, -20°C).

Engineering Facts

13 facts · every claim linked to a page
Installation

The chiller requires cooling water with a differential pressure of at least 1 bar and no more than 6 bar, so the facility water supply must be verified before installation.

Insufficient or excessive cooling water pressure will prevent proper operation or damage the unit.

Source
“min. cooling water differential pressure 1 bar max. cooling water pressure 6 bar”
Installation

The unit is designed for cooling water intake between 5°C and 20°C, and if the water temperature falls below the condensation point the cooling water pipes must be insulated.

Operating outside this range or without insulation risks condensation, corrosion, and reduced performance.

Source
“This temperature control unit is designed to operate with cooling water intake temperature between 5°C and 20°C. When cooling water temperatures fall below the condensation point, it is essential to insulate the cooling water pipes in the unit.”
Performance

Cooling capacity drops as cooling water temperature rises, and the required cooling water flow rate may increase, so sizing must account for actual site water temperature.

Rated cooling power is based on 15°C water; higher site water temperatures reduce available capacity.

Source
“As the cooling water temperature increases, drop in the cooling power should be expected, and also an increased cooling water flow rate possible.”
Selection

The plastic pump is not resistant to acetone or polyglycols, so these media must be avoided or a more resistant plastic pump ordered at extra cost.

Using incompatible media will damage the pump and void reliable operation.

Source
“Special Case: Acetone and Polyglycol: The plastic pump is not resistant against acetone and polyglycols (depending on the manufacturer).”
Operation

For freeze protection, water should be mixed with glysantine or ethylene glycol rather than relying on pure water.

Prevents freezing and damage when operating at low temperatures.

Source
“It is recommended that water is mixed with either glysantine or ethylene glycol for freeze protection.”
Installation

The cooling water circuit contains copper, stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM, so only suitable cooling water may be used.

Unsuitable water chemistry can corrode or foul these materials and impair heat transfer.

Source
“Materiels used in the cooling water circuit include; copper, Stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM. Please use suitable cooling water.”
Conformité

Voltage tolerance is ±5% only with simultaneous frequency tolerance of ±2%, and certain combinations such as -5% voltage with +2% frequency are not allowed.

Ensures the unit is not operated outside its specified supply tolerances, which could cause malfunction.

Source
“Voltage + / - 5% with a simultaneous frequency tolerance of + / - 2%”
Installation

The unit has IP20 protection and must be installed in an environment between 5°C and 40°C ambient temperature.

IP20 offers no water protection, and ambient limits affect cooling performance and safety.

Source
“Degree of Protection IP20 min. ambient temperature 5 °C max. ambient temperature 40 °C”
Safety

The refrigerant R-449A has a GWP of 1397 and a charge of 0.25 kg, corresponding to 0.3 t CO2 equivalent, which matters for leak reporting and environmental compliance.

Determines regulatory obligations for refrigerant containment and leak checks.

Source
“Global Warming Potential (GWP) 1397 Refrigerant quantity 0,25 kg CO2 equivalent 0,3 t”
Installation

The unit requires a 220-240V 1~ 50Hz supply with maximum current 13 A and a 16 A fuse.

Electrical supply and protection must be sized accordingly to avoid nuisance tripping or unsafe wiring.

Source
“Power supply requirement 220-240V 1~ 50Hz max. current 13 A Fuse 16 A”
Selection

The minimum filling capacity is 3.8 l and the expansion vessel volume is 1.7 l, which must be considered when selecting heat transfer fluid volume and expansion space.

Ensures sufficient fluid volume and proper expansion allowance for safe operation.

Source
“min. filling capacity 3,8 l”
Operation

The pump delivers 29 l/min maximum and 21 l/min at 0.5 bar, with a maximum delivery pressure of 1 bar, which constrains external circuit pressure drop.

External piping and equipment must be sized so the required flow is achievable within the pump curve.

Source
“max. delivery 29 l/min max. delivery pressure 1 bar Delivery at 0,5 bar 21 l/min”
Conformité

The equipment is classified for electromagnetic disturbance as EN55011 Class A, Group 1, which has implications for use in residential environments.

Class A equipment may cause radio interference and is not intended for residential use without mitigation.

Source
“Classification (disturbance) to EN55011: Class A, Group 1”