Unichiller 030Tw Technical Data

Sommaire

  1. Overview p. 1
    Introduces the Unichiller 030Tw chiller unit and its main features
  2. Spec pp. 1-2
    Lists performance, temperature, and physical specifications per DIN 12876
  3. p. 2 table, 11 rows
  4. p. 1 table, 34 rows

Technical data according to DIN 12876

ParameterValue
Overall dimensions WxDxH **400x440x1230 mm
Net weight137 kg
sound pressure level +/- 4 dB(A)61 dB(A)
Power supply (3 Phase)400V 3~ 50Hz
max. current (3 Phase)4,5 A
Fuse (3 phase)3x10A
Pressure equipment categoryArt. 4.3 PED
Degree of ProtectionIP20
min. ambient temperature5 °C
max. ambient temperature40 °C
from Serial-No.:S517916

** Please respect space requirements. See operating conditions at www.huber-online.com

Technical data according to DIN 12876 — Unichiller 030Tw (Order-No.: 3025.0056.01)

ParameterValue
Operating temperature range-20...40 °C
temperature set point / display5,7" colour Touchscreen
Internal temperature sensorPt100
Sensor external connectionPt100
Temperature stability at -10°C0,5 K
Interface digitalEthernet, USB (Host u. Device), RS232
Safety classificationI / NFL
Cooling power at 15°C3 kW
Cooling power at 0°C2,75 kW
Cooling power at -10°C2 kW
Cooling power at -20°C1 kW
Refrigeration machinewater-cooled, CFC- and HCFC-free
Refrigerant (ASHRAE, GHS)R-449A (A1, H280)
Global Warming Potential (GWP)1397
Refrigerant quantity0,7 kg
CO2 equivalent1,0 t
Circulation pumpB
max. delivery26 l/min
max. delivery pressure3 bar
Delivery at 0,2 bar22 l/min
Delivery at 0,5 bar20 l/min
Delivery at 1,0 bar16 l/min
Delivery at 1,5 bar12 l/min
Delivery at 2,0 bar8 l/min
Pump connectionG3/4 male
Cooling water connectionG1/2 male
Consumption at water 15°C, flow 15°C99 l/h
Consumption at water 15°C, flow 0°C99 l/h
Consumption at water 15°C, flow -10°C65 l/h
Consumption at water 15°C, flow -20°C25 l/h
min. cooling water differential pressure1 bar
max. cooling water pressure6 bar
min. filling capacity2,5 l
Volume of expansion4 l

Engineering Facts

12 facts · every claim linked to a page
Installation

The chiller is water-cooled and requires a cooling water differential pressure of at least 1 bar and no more than 6 bar at the connection.

Ensures the facility cooling water supply can meet the minimum pressure for proper operation without exceeding the maximum.

Source
“min. cooling water differential pressure 1 bar”
Installation

The unit is designed for cooling water intake between 5°C and 20°C, and pipes must be insulated if the water temperature can fall below the condensation point.

Defines the allowable cooling water supply temperature range and condensation protection requirement for installation.

Source
“This temperature control unit is designed to operate with cooling water intake temperature between 5°C and 20°C.”
Performance

Cooling power drops as cooling water temperature rises, and the required cooling water flow rate may increase.

Engineers must account for reduced capacity and higher water demand when the facility cooling water is warmer than the rated 15°C.

Source
“As the cooling water temperature increases, drop in the cooling power should be expected, and also an increased cooling water flow rate possible.”
Installation

The cooling water circuit contains copper, stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM, so only compatible cooling water may be used.

Material compatibility determines water quality requirements and prevents corrosion or damage in the cooling circuit.

Source
“Materiels used in the cooling water circuit include; copper, Stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM.”
Installation

Three-phase devices with current consumption below 63 A are supplied with a cable but without a plug.

The installer must provide and fit a suitable plug or hard-wire the supplied cable on site.

Source
“Three-phase devices with current consumption less than 63A --> with cable, without plug”
Conformité

Voltage tolerance is ±5% only with a simultaneous frequency tolerance of ±2%, and certain combinations such as -5% voltage with +2% frequency are not allowed.

Power supply quality must be verified against these combined tolerances to avoid invalid operating conditions.

Source
“-5% voltage and +2% frequency -> not allowed!”
Safety

The refrigerant R-449A is classified A1 with GHS hazard H280, and the charge is 0.7 kg corresponding to 1.0 t CO2 equivalent.

Affects refrigerant handling, leak reporting obligations and safety classification for the installation site.

Source
“R-449A (A1, H280)”
Installation

The unit has protection degree IP20 and must be operated within an ambient temperature of 5°C to 40°C.

IP20 limits installation to dry indoor locations and the ambient range constrains the room environment.

Source
“Degree of Protection IP20”
Selection

The chiller is intended for externally closed applications and has an atmospheric open expansion tank.

Confirms the hydraulic circuit type the unit is designed for, preventing misapplication with open baths or unsuitable loops.

Source
“For externally closed applications.”
Operation

The pump delivers 26 l/min maximum and 8 l/min at 2.0 bar, with G3/4 male pump connections.

Pump curve data is needed to match the external circuit pressure drop and required flow.

Source
“max. delivery 26 l/min”
Installation

The unit requires a 400 V 3~ 50 Hz supply with maximum current 4.5 A and 3x10 A fusing.

Electrical supply and protection must be provisioned accordingly before installation.

Source
“Power supply (3 Phase) 400V 3~ 50Hz”
Maintenance

The minimum filling capacity is 2.5 l and the expansion volume is 4 l.

Determines the minimum heat transfer fluid volume needed and expansion allowance for the external loop.

Source
“min. filling capacity 2,5 l”