Product introduction and general description of the chiller unit
Specpp. 1-2
Technical data, performance figures and operating parameters
Specp. 2
Detailed specifications according to DIN 12876 standard
p. 2table, 16 rows
p. 1table, 34 rows
Technical data according to DIN 12876
Parameter
Value
Consumption at water 15°C, flow -20°C
137 l/h
min. cooling water differential pressure
0,4 bar
max. cooling water pressure
6 bar
min. filling capacity
18 l
Volume of expansion
48 l
Overall dimensions WxDxH **
740x1160x1050 mm
Net weight
290 kg
sound pressure level +/- 4 dB(A)
62 dB(A)
Power supply (3 Phase)
460V 3~ 60Hz
max. current (3 Phase)
10,5 A
Fuse (3 phase)
3x16A
Pressure equipment category
Art. 4.3 PED
Degree of Protection
IP20
min. ambient temperature
5 °C
max. ambient temperature
40 °C
from Serial-No.:
S474338
** Please respect space requirements. See operating conditions at www.huber-online.com
Technical data according to DIN 12876 — Unichiller 050w-H (Order-No.: 3038.0061.01)
Characteristic
Value
Operating temperature range
-20...100 °C
temperature set point / display
5,7" colour Touchscreen
Internal temperature sensor
Pt100
Sensor external connection
Pt100
Temperature stability at -10°C
0,2 K
Interface digital
Ethernet, USB (Host u. Device), RS232
Safety classification
III / FL
Heating power
2 kW
Cooling power at ambient temperature +20°C
at 100°C
5 kW
at 20°C
5 kW
at 0°C
4,6 kW
at -10°C
3,4 kW
at -20°C
2 kW
Refrigeration machine
water-cooled, CFC- and HCFC-free
Refrigerant (ASHRAE, GHS)
R-449A (A1, H280)
Global Warming Potential (GWP)
1397
Refrigerant quantity
0,85 kg
CO2 equivalent
1,2 t
Circulation pump
E1
max. delivery
56 l/min
max. delivery pressure
3,8 bar
Delivery at 0,3 bar
53 l/min
Delivery at 0,5 bar
51 l/min
Delivery at 1,0 bar
46 l/min
Delivery at 1,5 bar
40 l/min
Delivery at 2,0 bar
34 l/min
Delivery at 2,5 bar
27 l/min
Delivery at 3,0 bar
18 l/min
Pump connection
G1 1/4 male
Cooling water connection
G1/2 male
Consumption at water 15°C, flow 20°C
240 l/h
Consumption at water 15°C, flow 0°C
216 l/h
Consumption at water 15°C, flow -10°C
167 l/h
Строка «Cooling power at ambient temperature +20°C» является заголовком группы для последующих строк at 100°C … at -20°C; значение в ней пустое.
Engineering Facts
14 facts · every claim linked to a page
Installation
The chiller requires a three-phase 460 V 60 Hz supply with a 3x16 A fuse, so the site must provide a suitable branch circuit.
Determines electrical infrastructure and protection sizing before installation.
Source
“Power supply (3 Phase) 460V 3~ 60Hz”
Installation
External branch circuit protection according to UL 489 is mandatory, so a listed breaker must be included in the installation.
Affects electrical panel design and compliance for North American installations.
Source
“External branch circuit protection according to UL 489 required.”
Installation
Three-phase devices drawing less than 63 A are supplied with cable but without a plug, so the customer must provide the final connection.
Clarifies what is in the scope of delivery and what the installer must supply.
Source
“Three-phase devices with current consumption less than 63A --> with cable, without plug”
Installation
Cooling water inlet temperature must stay between 5 °C and 20 °C, and pipes must be insulated if condensation conditions occur.
Defines the cooling water supply requirements and condensation risk at the site.
Source
“This temperature control unit is designed to operate with cooling water intake temperature between 5°C and 20°C.”
Installation
The cooling water circuit must provide at least 0.4 bar differential pressure and must not exceed 6 bar.
Sets the acceptable cooling water supply pressure window for correct operation.
Source
“min. cooling water differential pressure 0,4 bar”
Installation
The unit needs at least 18 l filling capacity and has 48 l expansion volume, which matters for loop design and fluid charge.
Informs the required external loop volume and expansion tank sizing.
Source
“min. filling capacity 18 l”
Installation
The chiller is water-cooled and requires a cooling water flow of 240 l/h at 15 °C water and 20 °C flow temperature.
Determines cooling water supply capacity needed at the installation site.
Source
“Consumption at water 15°C, flow 20°C 240 l/h”
Installation
The cooling water circuit materials include copper, stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM, so only compatible water quality should be used.
Guides water quality and material compatibility to avoid corrosion or fouling.
Source
“Materiels used in the cooling water circuit include; copper, Stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM.”
Operation
The unit is rated for ambient temperatures from 5 °C to 40 °C, so the room must stay within this range.
Ensures the installation location meets the environmental operating envelope.
Source
“min. ambient temperature 5 °C”
Operation
Cooling power drops as cooling water temperature rises and flow rate may need to increase, so performance must be derated accordingly.
Prevents overestimating cooling capacity when site water is warmer than 15 °C.
Source
“As the cooling water temperature increases, drop in the cooling power should be expected, and also an increased cooling water flow rate possible.”
Operation
Voltage and frequency tolerances are ±5% and ±2% respectively, but not simultaneously in opposite directions.
Defines acceptable supply quality limits for reliable operation.
Source
“Voltage + / - 5% with a simultaneous frequency tolerance of + / - 2%”
Safety
The refrigerant R-449A is classified A1 with GHS H280, and the unit contains 0.85 kg with 1.2 t CO2 equivalent.
Informs refrigerant safety classification and environmental reporting obligations.
Source
“Refrigerant (ASHRAE, GHS) R-449A (A1, H280)”
Safety
The unit has safety classification III / FL according to DIN 12876 and adjustable overtemperature protection.
Confirms the safety category relevant for flammable liquids and protection design.
Source
“Safety classification III / FL”
Conformité
The equipment is classified Class A, Group 1 for electromagnetic disturbance to EN55011, which affects EMC planning.
Determines EMC environment suitability and possible interference measures.
Source
“Classification (disturbance) to EN55011: Class A, Group 1”