Unichiller 060T Technical Data

Sommaire

  1. Overview p. 1
    Introduces the Unichiller 060T model and its main features
  2. Spec pp. 1-2
    Lists performance, temperature, and dimensional data per DIN 12876
  3. p. 2 table, 6 rows
  4. p. 1 table, 35 rows

Technical data according to DIN 12876

ParameterValue
Fuse (3 phase)3x16A
Pressure equipment categoryArt. 4.3 PED
Degree of ProtectionIP20
min. ambient temperature5 °C
max. ambient temperature40 °C
from Serial-No.:S540365

Unichiller 060T — Technical data according to DIN 12876

ParameterValue
Order-No.3026.0111.01
Operating temperature range-20...40 °C
temperature set point / display5,7" colour Touchscreen
Internal temperature sensorPt100
Sensor external connectionPt100
Interface digitalEthernet, USB (Host u. Device), RS232
Temperature stability at -10°C0,5 K
Safety classificationI / NFL
Cooling power at 15°C6 kW
Cooling power at 0°C5 kW
Cooling power at -10°C2,8 kW
Cooling power at -20°C1,4 kW
Refrigeration machineair-cooled, CFC- and HCFC-free
Refrigerant (ASHRAE, GHS)R-449A (A1, H280)
Global Warming Potential (GWP)1397
Refrigerant quantity2,2 kg
CO2 equivalent3,1 t
Circulation pump:C3
max. delivery80 l/min
max. delivery pressure5,6 bar
Delivery at 0,35 bar79 l/min
Delivery at 0,5 bar76 l/min
Delivery at 1,0 bar71 l/min
Delivery at 1,5 bar66 l/min
Delivery at 2,0 bar61 l/min
Delivery at 3,0 bar49 l/min
Delivery at 4,0 bar35 l/min
Delivery at 5,0 bar16 l/min
Pump connectionG1 1/4 male
min. filling capacity5 l
Volume of expansion7,5 l
Overall dimensions WxDxH **600x692x1613 mm
Net weight228 kg
Power supply (3 Phase)400V 3~ 50Hz
max. current (3 Phase)10 A

** см. сноску в оригинале (Overall dimensions WxDxH **)

Engineering Facts

14 facts · every claim linked to a page
Selection

The chiller is rated for an operating temperature range of -20 to 40 °C, so the process must stay within these limits.

Defines the application window and excludes processes requiring lower or higher temperatures.

Source
“Operating temperature range -20...40 °C”
Performance

Cooling capacity drops from 6 kW at 15 °C to 1.4 kW at -20 °C, so capacity must be checked at the actual working temperature.

Sizing on nominal capacity alone risks insufficient cooling at low set points.

Source
“at 15°C 6 kW at 0°C 5 kW at -10°C 2,8 kW at -20°C 1,4 kW”
Installation

The unit requires a 400 V 3-phase 50 Hz supply with 10 A max current and 3x16 A fusing.

Electrical infrastructure and protection must be planned accordingly before installation.

Source
“Power supply (3 Phase) 400V 3~ 50Hz max. current (3 Phase) 10 A”
Installation

Three-phase devices below 63 A are delivered with cable but without a plug, so a suitable plug must be provided on site.

Prevents delays at commissioning due to missing power connection hardware.

Source
“Three-phase devices with current consumption less than 63A --> with cable, without plug”
Installation

Voltage and frequency tolerances are ±5% and ±2% but must not be combined in the unfavorable direction.

Site power quality must be verified to avoid invalid supply conditions and equipment damage.

Source
“Voltage + / - 5% with a simultaneous frequency tolerance of + / - 2%”
Installation

The chiller is air-cooled, so sufficient ambient air and space for heat rejection must be ensured.

Air-cooled units need adequate ventilation and clearance to maintain cooling performance.

Source
“Refrigeration machine air-cooled, CFC- and HCFC-free”
Operation

Cooling capacity is valid at 20 °C room temperature and may drop if ambient temperature rises.

Performance must be derated in warm rooms to avoid undersized cooling.

Source
“Output data valid for: Room temperature 20°C. If the ambient temperature rises, the cooling capacity may drop.”
Operation

The unit may only be operated between 5 °C and 40 °C ambient temperature.

Ambient conditions outside this range can impair operation and must be controlled.

Source
“min. ambient temperature 5 °C max. ambient temperature 40 °C”
Safety

The refrigerant R-449A is classified A1 with GHS hazard H280 and has a GWP of 1397.

Refrigerant handling, labeling and environmental obligations depend on this classification.

Source
“R-449A (A1, H280)”
Conformité

The CO2 equivalent of the refrigerant charge is 3.1 t, which triggers F-gas documentation and leak-check obligations.

Determines regulatory maintenance and reporting requirements for the refrigerant circuit.

Source
“CO2 equivalent 3,1 t”
Installation

The pump connection is G1 1/4 male, so mating hoses and fittings must match this thread.

Ensures correct hydraulic connection and avoids leaks or adapters.

Source
“Pump connection G1 1/4 male”
Operation

The circulation pump delivers 80 l/min max at 5.6 bar max, with flow falling as pressure rises.

Hydraulic design must respect the pump curve to avoid insufficient flow in the loop.

Source
“max. delivery 80 l/min max. delivery pressure 5,6 bar”
Installation

The unit has IP20 protection, so it must be installed in a dry, protected environment.

Prevents installation in wet or dusty locations that would violate protection rating.

Source
“Degree of Protection IP20”
Conformité

The equipment is EMC class A group 1, which may require additional measures in residential environments.

Class A equipment is intended for industrial use and may cause interference in residential areas.

Source
“Classification (disturbance) to EN55011: Class A, Group 1”