Unichiller 100w-H Datasheet

Sommaire

  1. Overview p. 1
    Introduces the Unichiller 100w-H chiller and its main features
  2. Spec pp. 1-2
    Performance, temperature, and dimensional data per DIN 12876
  3. p. 2 table, 16 rows
  4. p. 1 table, 33 rows

Technical data according to DIN 12876

ParameterValue
Consumption at water 15°C, flow -20°C240 l/h
min. cooling water differential pressure0,5 bar
max. cooling water pressure6 bar
min. filling capacity18 l
Volume of expansion48 l
Overall dimensions WxDxH **740x1160x1050 mm
Net weight307 kg
sound pressure level +/- 4 dB(A)58 dB(A)
Power supply (3 Phase)400V 3~ 50Hz
max. current (3 Phase)14 A
Fuse (3 phase)3x16A
Pressure equipment categoryArt. 4.3 PED
Degree of ProtectionIP20
min. ambient temperature5 °C
max. ambient temperature40 °C
from Serial-No.:S534719

** Please respect space requirements. See operating conditions at www.huber-online.com

Technical data according to DIN 12876 — Unichiller 100w-H (Order-No. 3059.0010.01)

ParameterValue
Operating temperature range-20...100 °C
temperature set point / display5,7" colour Touchscreen
Internal temperature sensorPt100
Sensor external connectionPt100
Temperature stability at -10°C0,2 K
Interface digitalEthernet, USB (Host u. Device), RS232
Safety classificationIII / FL
Heating power2 kW
Cooling power at ambient temperature +20°C at 100°C10 kW
Cooling power at ambient temperature +20°C at 20°C10 kW
Cooling power at ambient temperature +20°C at 0°C8,6 kW
Cooling power at ambient temperature +20°C at -10°C6 kW
Cooling power at ambient temperature +20°C at -20°C3,9 kW
Refrigeration machinewater-cooled, CFC- and HCFC-free
Refrigerant (ASHRAE, GHS)R-449A (A1, H280)
Global Warming Potential (GWP)1397
Refrigerant quantity1,03 kg
CO2 equivalent1,4 t
Circulation pumpE1
max. delivery48 l/min
max. delivery pressure3,4 bar
Delivery at 0,3 bar46 l/min
Delivery at 0,5 bar44 l/min
Delivery at 1,0 bar40 l/min
Delivery at 1,5 bar33 l/min
Delivery at 2,0 bar27 l/min
Delivery at 2,5 bar20 l/min
Delivery at 3,0 bar11 l/min
Pump connectionG1 1/4 male
Cooling water connectionG1/2 male
Consumption at water 15°C, flow 20°C348 l/h
Consumption at water 15°C, flow 0°C330 l/h
Consumption at water 15°C, flow -10°C270 l/h

Engineering Facts

14 facts · every claim linked to a page
Installation

The chiller requires a three-phase 400 V supply with a 3x16 A fuse, so the electrical installation must be sized accordingly.

Determines the site power supply, cable and protection device rating before installation.

Source
“Power supply (3 Phase) 400V 3~ 50Hz”
Installation

The unit is water-cooled and needs a cooling water supply with differential pressure between 0.5 and 6 bar.

Cooling water pressure must be verified on site to ensure adequate flow and avoid overpressure damage.

Source
“min. cooling water differential pressure 0,5 bar”
Installation

Cooling water inlet temperature must be kept between 5 °C and 20 °C for the published performance data to apply.

Defines the permissible cooling water supply conditions and warns that higher temperatures reduce cooling capacity.

Source
“This temperature control unit is designed to operate with cooling water intake temperature between 5°C and 20°C.”
Installation

If cooling water temperature can fall below the dew point, the cooling water pipes inside the unit must be insulated.

Prevents condensation and corrosion inside the unit when cold cooling water is used.

Source
“When cooling water temperatures fall below the condensation point, it is essential to insulate the cooling water pipes in the unit.”
Installation

The cooling water circuit contains copper, stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM, so only compatible cooling water may be used.

Guides water quality and material compatibility to avoid corrosion or fouling of the cooling circuit.

Source
“Materiels used in the cooling water circuit include; copper, Stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM.”
Installation

Three-phase devices below 63 A are delivered with a cable but without a plug, so a suitable plug must be provided on site.

Affects electrical installation planning and the need to procure a matching plug.

Source
“Three-phase devices with current consumption less than 63A --> with cable, without plug”
Operation

The unit is rated for ambient temperatures from 5 °C to 40 °C, which limits where it can be installed.

Room conditions must be checked to keep the chiller within its operating envelope.

Source
“min. ambient temperature 5 °C”
Operation

The refrigerant is R-449A with GWP 1397 and a charge of 1.03 kg, giving a CO2 equivalent of 1.4 t.

Relevant for refrigerant handling, leak reporting and environmental compliance.

Source
“Refrigerant (ASHRAE, GHS) R-449A (A1, H280)”
Performance

Cooling capacity drops from 10 kW at 20 °C to 3.9 kW at -20 °C, so the required load must be checked at the lowest operating temperature.

Ensures the chiller can remove the process heat at the coldest set point, not just at ambient.

Source
“at -20°C 3,9 kW”
Performance

The circulation pump delivers 48 l/min at zero head but only 11 l/min at 3.0 bar, so system pressure drop must be matched to the pump curve.

Prevents undersized flow through the external circuit at high pressure drop.

Source
“Delivery at 3,0 bar 11 l/min”
Operation

The minimum filling capacity is 18 l and the expansion volume is 48 l, which sets the minimum heat transfer fluid charge.

Determines the required volume of heat transfer fluid and expansion tank sizing for safe operation.

Source
“min. filling capacity 18 l”
Conformité

Voltage tolerance is ±5% only with a simultaneous frequency tolerance of ±2%, and certain combinations are not allowed.

Site power quality must be checked to avoid invalid supply combinations that could damage the unit.

Source
“Voltage + / - 5% with a simultaneous frequency tolerance of + / - 2%”
Safety

The unit has safety classification III / FL and adjustable overtemperature protection according to DIN 12876.

Confirms the safety concept and the need to set the overtemperature limit for the application.

Source
“With adjustable overtemperature protection according to DIN 12876.”
Installation

The unit has IP20 protection, so it must be installed in a dry indoor location protected from water ingress.

Restricts installation environment and prevents moisture or dust ingress hazards.

Source
“Degree of Protection IP20”