  1. Overview p. 1
    Product introduction and key features of the Unichiller 130T
  2. Spec p. 2
    Technical data according to DIN 12876 for the chiller
  3. p. 2 table, 9 rows
  4. p. 1 table, 35 rows

Technical data according to DIN 12876

ParameterValue
sound pressure level +/- 4 dB(A)74 dB(A)
Power supply (3 Phase)400V 3~ 50Hz
max. current (3 Phase)12 A
Fuse (3 phase)3x16A
Pressure equipment categoryI
Degree of ProtectionIP20
min. ambient temperature5 °C
max. ambient temperature40 °C
from Serial-No.:S477709

Unichiller 130T — Technical data according to DIN 12876

ParameterValue
Order-No.3018.0016.01
Operating temperature range-10...40 °C
temperature set point / display5,7" colour Touchscreen
Internal temperature sensorPt100
Sensor external connectionPt100
Interface digitalEthernet, USB (Host u. Device), RS232
Temperature stability at -10°C0,5 K
Safety classificationI / NFL
Cooling power at 15°C13 kW
Cooling power at 0°C7 kW
Cooling power at -10°C4,5 kW
Refrigeration machineair-cooled, CFC- and HCFC-free
Refrigerant (ASHRAE, GHS)R-449A (A1, H280)
Global Warming Potential (GWP)1397
Refrigerant quantity4,4 kg
CO2 equivalent6,1 t
Circulation pump:C3
max. delivery90 l/min
max. delivery pressure5,6 bar
Delivery at 0,33 bar94 l/min
Delivery at 0,5 bar92 l/min
Delivery at 1,0 bar87 l/min
Delivery at 1,5 bar82 l/min
Delivery at 2,0 bar75,5 l/min
Delivery at 2,5 bar69 l/min
Delivery at 3,0 bar62 l/min
Delivery at 3,5 bar54,5 l/min
Delivery at 4,0 bar47 l/min
Delivery at 4,5 bar38 l/min
Delivery at 5,0 bar26 l/min
Pump connectionG1 1/4 male
min. filling capacity14 l
Volume of expansion39 l
Overall dimensions WxDxH **904x1582x1895 mm
Net weight494 kg

** см. сноску в оригинале (на странице не приведена)

Engineering Facts

14 facts · every claim linked to a page
Installation

The chiller requires a three-phase 400 V supply with a 3x16 A fuse, so the electrical installation must be sized accordingly.

Determines the mains connection, cable and protection device needed at the installation site.

Source
“Power supply (3 Phase) 400V 3~ 50Hz”
Installation

Three-phase devices below 63 A are delivered with a cable but without a plug, so the customer must provide and fit the plug.

Affects site preparation and the scope of electrical works before commissioning.

Source
“Three-phase devices with current consumption less than 63A --> with cable, without plug”
Installation

The unit must be installed in an environment between 5 °C and 40 °C ambient temperature.

Room conditions must be checked during selection and site planning.

Source
“min. ambient temperature 5 °C”
Performance

Cooling capacity is rated at 20 °C room temperature and may drop if the ambient temperature rises.

The specified 13 kW at 15 °C is not guaranteed at higher ambient temperatures, so selection must account for the actual room temperature.

Source
“Output data valid for: Room temperature 20°C. If the ambient temperature rises, the cooling capacity may drop.”
Installation

The enclosure is IP20, so it must be protected from water and dust ingress at the installation location.

Restricts where the chiller can be installed and what environmental protection is needed.

Source
“Degree of Protection IP20”
Safety

The refrigerant R-449A is classified A1 with GHS hazard H280, so refrigerant handling and ventilation requirements apply.

Affects safety planning, leak detection and personnel qualification for service.

Source
“R-449A (A1, H280)”
Conformité

The CO2 equivalent of the refrigerant charge is 6.1 t, which triggers F-gas record-keeping and leak-check obligations in many jurisdictions.

Determines regulatory obligations for operation and maintenance of the refrigeration circuit.

Source
“CO2 equivalent 6,1 t”
Installation

The pump connection is G1 1/4 male, so the external circuit piping must match this thread.

Needed to order the correct hose couplings and plan the hydraulic connection.

Source
“Pump connection G1 1/4 male”
Operation

The minimum filling capacity is 14 l and the expansion volume is 39 l, which sets the minimum external loop volume and expansion allowance.

The external circuit must provide at least this volume and accommodate expansion to avoid operating problems.

Source
“min. filling capacity 14 l”
Installation

The unit is air-cooled and requires space around it, so the stated dimensions and space requirements must be respected.

Insufficient clearance reduces cooling performance and can cause overheating.

Source
“Please respect space requirements.”
Conformité

Voltage tolerance is ±5% only with a simultaneous frequency tolerance of ±2%, and certain combinations are not allowed.

The site power quality must be checked to avoid invalid supply conditions and equipment damage.

Source
“Voltage + / - 5% with a simultaneous frequency tolerance of + / - 2%”
Operation

The sound pressure level is 74 dB(A), which must be considered for personnel exposure and room acoustics.

Affects workplace noise assessment and installation location.

Source
“sound pressure level +/- 4 dB(A) 74 dB(A)”
Selection

The temperature stability at -10 °C is 0.5 K, which defines the achievable process control accuracy at low temperature.

Used to verify whether the chiller meets the process tolerance requirements.

Source
“Temperature stability at -10°C 0,5 K”
Maintenance

A 4-year warranty is available only with registration, so the unit must be registered after purchase.

Impacts lifecycle cost and the need to complete registration at commissioning.

Source
“4-year warranty - registration required.”