Unichiller 200Tw-H Datasheet

Sommaire

  1. Overview p. 1
    Product introduction and main features of the chiller
  2. Spec pp. 1-2
    Technical data, performance and dimensions per DIN 12876
  3. p. 2 table, 14 rows
  4. p. 1 table, 33 rows

Technical data according to DIN 12876

ParameterValue
Consumption at water 15°C, flow -20°C445 l/h
min. cooling water differential pressure1 bar
max. cooling water pressure6 bar
min. filling capacity15 l
Volume of expansion22 l
Overall dimensions WxDxH **760x800x1615 mm
Net weight374 kg
Power supply (3 Phase)400V 3~ 50Hz
max. current (3 Phase)17,5 A
Fuse (3 phase)3x20A
Degree of ProtectionIP20
min. ambient temperature5 °C
max. ambient temperature40 °C
from Serial-No.:S361439

** Please respect space requirements. See operating conditions at www.huber-online.com

Technical data according to DIN 12876 — Unichiller 200Tw-H (Order-No.: 3028.0113.01)

ParameterValue
Operating temperature range-20...100 °C
temperature set point / display5,7" colour Touchscreen
Internal temperature sensorPt100
Sensor external connectionPt100
Temperature stability at -10°C0,2 K
Interface digitalEthernet, USB (Host u. Device), RS232
Safety classificationIII / FL
Heating power2 kW
Cooling power at 15°C20 kW
Cooling power at 0°C11 kW
Cooling power at -10°C5 kW
Cooling power at -20°C3 kW
Refrigeration machinewater-cooled, CFC- and HCFC-free
Refrigerant (ASHRAE, GHS)R-449A (A1, H280)
Global Warming Potential (GWP)1397
Refrigerant quantity2,35 kg
CO2 equivalent3,3 t
Circulation pumpD3
max. delivery210 l/min
max. delivery pressure4,7 bar
Delivery at 1,1 bar202 l/min
Delivery at 1,5 bar193 l/min
Delivery at 2,0 bar179 l/min
Delivery at 2,5 bar159 l/min
Delivery at 3,0 bar140 l/min
Delivery at 3,5 bar115 l/min
Delivery at 4,0 bar77 l/min
Delivery at 4,5 bar31 l/min
Pump connectionG1 1/4 male
Cooling water connectionG3/4 male
Consumption at water 15°C, flow 15°C725 l/h
Consumption at water 15°C, flow 0°C860 l/h
Consumption at water 15°C, flow -10°C610 l/h

Engineering Facts

13 facts · every claim linked to a page
Installation

The chiller is water-cooled and requires a cooling water supply with a differential pressure between 1 and 6 bar, so the site must provide a suitable cooling water circuit.

Determines whether the facility's cooling water supply is adequate and what pressure regulation may be needed.

Source
“min. cooling water differential pressure 1 bar max. cooling water pressure 6 bar”
Installation

The unit is designed for cooling water intake between 5°C and 20°C, and if the cooling water falls below the condensation point the cooling water pipes inside the unit must be insulated.

Defines the allowable cooling water temperature window and a mandatory insulation measure to prevent condensation.

Source
“This temperature control unit is designed to operate with cooling water intake temperature between 5°C and 20°C.”
Performance

Cooling power drops as cooling water temperature rises, and the required cooling water flow rate may increase, so performance must be derated for warmer cooling water.

Prevents oversizing assumptions and ensures the cooling water supply can handle increased flow at higher temperatures.

Source
“As the cooling water temperature increases, drop in the cooling power should be expected, and also an increased cooling water flow rate possible.”
Installation

The cooling water circuit contains copper, stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM, so only compatible cooling water may be used.

Guides water quality and material compatibility to avoid corrosion or damage to the cooling circuit.

Source
“Materiels used in the cooling water circuit include; copper, Stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM.”
Conformité

Voltage and frequency tolerances are ±5% and ±2% respectively, but the combination of -5% voltage with +2% frequency is not allowed.

Ensures the electrical supply stays within the permitted combined tolerance envelope to avoid malfunction.

Source
“-5% voltage and +2% frequency -> not allowed!”
Installation

Three-phase devices drawing less than 63 A are supplied with a cable but without a plug, so a suitable plug must be provided on site.

Affects electrical installation planning and the need to procure and fit a compatible plug.

Source
“Three-phase devices with current consumption less than 63A --> with cable, without plug”
Safety

The refrigerant R-449A is classified A1 with GHS hazard H280, and the unit contains 2.35 kg corresponding to 3.3 t CO2 equivalent.

Informs refrigerant safety classification, leak reporting obligations and environmental impact assessment.

Source
“R-449A (A1, H280)”
Installation

The unit requires a minimum filling capacity of 15 l and has an expansion volume of 22 l, which must be accounted for when preparing the external circuit.

Determines the minimum heat transfer fluid volume and expansion tank sizing for the external loop.

Source
“min. filling capacity 15 l Volume of expansion 22 l”
Installation

The unit has IP20 protection and must be operated between 5°C and 40°C ambient temperature, so it needs a dry, climate-controlled location.

Constrains the installation environment and prevents ingress of water or dust that IP20 does not protect against.

Source
“Degree of Protection IP20 min. ambient temperature 5 °C max. ambient temperature 40 °C”
Installation

The unit is 760x800x1615 mm and weighs 374 kg net, and space requirements must be respected per the operating conditions.

Needed for floor loading, transport route and clearance planning during installation.

Source
“** Please respect space requirements. See operating conditions at www.huber-online.com”
Operation

The pump delivers 210 l/min maximum at 4.7 bar, but flow falls sharply at higher pressure (e.g. 31 l/min at 4.5 bar), so system pressure drop must be matched to the pump curve.

Ensures the external circuit pressure drop does not starve flow and that the pump operating point is acceptable.

Source
“max. delivery 210 l/min max. delivery pressure 4,7 bar”
Safety

The unit has adjustable overtemperature protection according to DIN 12876, which must be set for the application.

Provides a safety cut-out that must be configured to protect the process and the unit.

Source
“With adjustable overtemperature protection according to DIN 12876.”
Operation

The unit is intended for externally closed applications, so it must not be used with an open external bath.

Prevents misapplication that could cause fluid loss, evaporation or unsafe operation.

Source
“For externally closed applications.”