Introduces the Unichiller 210Tw-H chiller and its main features
Specpp. 1-2
Performance, temperature, and dimensional data per DIN 12876
p. 2table, 10 rows
p. 1table, 33 rows
Technical data according to DIN 12876
Parameter
Value
Volume of expansion
22 l
Overall dimensions WxDxH **
760x800x1615 mm
Power supply (3 Phase)
400V 3~ 50Hz
max. current (3 Phase)
24 A
Fuse (3 phase)
3x25A
Pressure equipment category
I
Degree of Protection
IP20
min. ambient temperature
5 °C
max. ambient temperature
40 °C
from Serial-No.:
S455148
** Please respect space requirements. See operating conditions at www.huber-online.com
Technical data according to DIN 12876 — Unichiller 210Tw-H (Order-No.: 3020.0047.01)
Parameter
Value
Operating temperature range
-20...100 °C
temperature set point / display
5,7" colour Touchscreen
Internal temperature sensor
Pt100
Sensor external connection
Pt100
Temperature stability at -10°C
0,2 K
Interface digital
Ethernet, USB (Host u. Device), RS232
Safety classification
III / FL
Heating power
2 kW
Cooling power at 15°C
21 kW
Cooling power at 0°C
21 kW
Cooling power at -10°C
15,5 kW
Cooling power at -20°C
9,5 kW
Refrigeration machine
water-cooled, CFC- and HCFC-free
Refrigerant (ASHRAE, GHS)
R-449A (A1, H280)
Global Warming Potential (GWP)
1397
Refrigerant quantity
2,7 kg
CO2 equivalent
3,8 t
Circulation pump
D3
max. delivery
210 l/min
max. delivery pressure
4,7 bar
Delivery at 1,1 bar
202 l/min
Delivery at 1,5 bar
193 l/min
Delivery at 2,0 bar
179 l/min
Delivery at 2,5 bar
159 l/min
Delivery at 3,0 bar
140 l/min
Delivery at 3,5 bar
115 l/min
Delivery at 4,0 bar
77 l/min
Delivery at 4,5 bar
31 l/min
Pump connection
G1 1/4 male
Cooling water connection
G3/4 male
min. cooling water differential pressure
1 bar
max. cooling water pressure
6 bar
min. filling capacity
15 l
Engineering Facts
12 facts · every claim linked to a page
Installation
The unit requires a three-phase 400 V supply with a 3x25 A fuse, so the electrical installation must be sized accordingly.
Determines the site power supply, cable and breaker requirements before installation.
Source
“Power supply (3 Phase) 400V 3~ 50Hz”
Installation
The chiller is water-cooled and needs a cooling water supply with at least 1 bar differential pressure and no more than 6 bar.
Cooling water pressure must be verified on site to ensure adequate heat rejection and avoid overpressure.
Source
“min. cooling water differential pressure 1 bar”
Installation
Cooling water inlet temperature must be kept between 5 °C and 20 °C, and pipes must be insulated if it drops below the condensation point.
Out-of-range cooling water temperature reduces cooling capacity and can cause condensation and corrosion.
Source
“This temperature control unit is designed to operate with cooling water intake temperature between 5°C and 20°C.”
Performance
Cooling power drops from 21 kW at 0 °C to 9.5 kW at -20 °C, so the cooling capacity must be checked at the actual operating temperature.
Selecting the chiller based only on nominal capacity can lead to insufficient cooling at low temperatures.
Source
“at -20°C 9,5 kW”
Performance
The circulation pump delivers 210 l/min at zero head but only 31 l/min at 4.5 bar, so system pressure drop must be matched to the pump curve.
Ensures the pump can overcome the external circuit pressure drop at the required flow rate.
Source
“max. delivery 210 l/min”
Installation
The unit needs a minimum filling capacity of 15 l and has a 22 l expansion volume, which must be considered when sizing the external loop.
Prevents underfilling and ensures the expansion tank can accommodate thermal volume changes.
Source
“min. filling capacity 15 l”
Safety
The refrigerant R-449A is classified A1/H280 with GWP 1397 and a CO2 equivalent of 3.8 t, which affects leak detection and environmental compliance.
Requires appropriate refrigerant handling, leak monitoring and F-gas reporting obligations.
Source
“R-449A (A1, H280)”
Installation
The unit is IP20 and must be installed where ambient temperature stays between 5 °C and 40 °C.
IP20 offers no water protection, and ambient limits affect reliable operation and cooling performance.
Source
“Degree of Protection IP20”
Installation
Three-phase devices below 63 A are supplied with a cable but without a plug, so a suitable plug must be provided on site.
Avoids delays at commissioning because the customer must supply and fit the plug.
Source
“Three-phase devices with current consumption less than 63A --> with cable, without plug”
Conformité
Voltage and frequency tolerances are ±5% and ±2% respectively, but certain combinations such as -5% voltage with +2% frequency are not allowed.
Ensures the site power quality is within the permitted envelope before connecting the chiller.
Source
“-5% voltage and +2% frequency -> not allowed!”
Operation
The unit is classified as pressure equipment category I and uses an atmospheric open expansion tank, so it is intended for externally closed applications.
Prevents applying the chiller to open or unsuitable circuits that could cause overflow or contamination.
Source
“For externally closed applications.”
Maintenance
The cooling water circuit contains copper, stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM, so only compatible cooling water may be used.
Guides water quality and treatment to avoid corrosion and fouling of the cooling circuit.
Source
“Materiels used in the cooling water circuit include; copper, Stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM.”