Unichiller 230w-H Datasheet

Sommaire

  1. Overview p. 1
    Introduces the Unichiller 230w-H chiller unit
  2. Spec pp. 1-2
    Performance, temperature, and dimensional data per DIN 12876
  3. p. 2 table, 10 rows
  4. p. 1 table, 34 rows

Technical data according to DIN 12876

ParameterValue
max. cooling water pressure6 bar
min. filling capacity18 l
Volume of expansion48 l
Overall dimensions WxDxH **940x1290x1130 mm
Power supply (3 Phase)460V 3~ 60Hz
Pressure equipment categoryI
Degree of ProtectionIP20
min. ambient temperature5 °C
max. ambient temperature40 °C
from Serial-No.:1.0/21

** Please respect space requirements. See operating conditions at www.huber-online.com

Unichiller 230w-H — Technical data according to DIN 12876

ParameterValue
Order-No.3039.0038.01
Operating temperature range-20...100 °C
temperature set point / display5,7" colour Touchscreen
Internal temperature sensorPt100
Sensor external connectionPt100
Temperature stability at -10°C0,2 K
Interface digitalEthernet, USB (Host u. Device), RS232
Safety classificationIII / FL
Heating power2 kW
Cooling power at ambient temperature +20°C at 100°C23 kW
Cooling power at ambient temperature +20°C at 20°C23 kW
Cooling power at ambient temperature +20°C at 0°C13,5 kW
Cooling power at ambient temperature +20°C at -10°C9 kW
Cooling power at ambient temperature +20°C at -20°C5,5 kW
Refrigeration machinewater-cooled, CFC- and HCFC-free
Refrigerant (ASHRAE, GHS)R-449A (A1, H280)
Global Warming Potential (GWP)1397
Circulation pumpE1
max. delivery56 l/min
max. delivery pressure3,8 bar
Förderleistung bei 0,17 bar53 l/min
Delivery at 0,5 bar51 l/min
Delivery at 1,0 bar46 l/min
Delivery at 1,5 bar40 l/min
Delivery at 2,0 bar34 l/min
Delivery at 2,5 bar27 l/min
Delivery at 3,0 bar18 l/min
Pump connectionG1 1/4 male
Cooling water connectionG3/4 male
Consumption at water 15°C, flow 20°C744 l/h
Consumption at water 15°C, flow 0°C540 l/h
Consumption at water 15°C, flow -10°C414 l/h
Consumption at water 15°C, flow -20°C324 l/h
min. cooling water differential pressure1 bar

Engineering Facts

14 facts · every claim linked to a page
Installation

The chiller requires a three-phase 460 V 60 Hz supply, so the site must provide this exact power configuration.

Determines electrical infrastructure and whether a transformer or different model is needed.

Source
“Power supply (3 Phase) 460V 3~ 60Hz”
Installation

External branch circuit protection per UL 489 must be provided by the installer.

Affects panel design and compliance with local electrical codes.

Source
“External branch circuit protection according to UL 489 required.”
Installation

Three-phase devices drawing less than 63 A are supplied with cable but no plug, so a suitable plug must be sourced separately.

Prevents delays at commissioning due to missing plug and clarifies scope of delivery.

Source
“Three-phase devices with current consumption less than 63A --> with cable, without plug”
Installation

Cooling water inlet temperature must be between 5 °C and 20 °C, and pipes must be insulated if below the condensation point.

Defines cooling water supply requirements and condensation risk mitigation.

Source
“This temperature control unit is designed to operate with cooling water intake temperature between 5°C and 20°C.”
Installation

A minimum cooling water differential pressure of 1 bar is required for proper operation.

Ensures adequate flow through the water-cooled refrigeration circuit.

Source
“min. cooling water differential pressure 1 bar”
Installation

Cooling water pressure must not exceed 6 bar.

Protects the cooling water circuit from overpressure damage.

Source
“max. cooling water pressure 6 bar”
Installation

The unit requires a minimum filling capacity of 18 l and has an expansion volume of 48 l.

Determines heat transfer fluid volume needed and expansion tank sizing.

Source
“min. filling capacity 18 l”
Installation

The unit must be installed in an environment between 5 °C and 40 °C ambient temperature.

Affects room selection and ventilation/cooling requirements.

Source
“min. ambient temperature 5 °C”
Installation

The unit has IP20 protection, so it must be kept in a dry, indoor location free from water ingress.

Restricts installation to protected indoor environments.

Source
“Degree of Protection IP20”
Operation

Cooling capacity drops as cooling water temperature rises, and flow rate may need to increase.

Affects performance expectations and cooling water system sizing.

Source
“As the cooling water temperature increases, drop in the cooling power should be expected, and also an increased cooling water flow rate possible.”
Operation

The cooling water circuit contains copper, stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM, so only compatible water quality should be used.

Guides selection of cooling water treatment to avoid corrosion and fouling.

Source
“Materiels used in the cooling water circuit include; copper, Stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM.”
Conformité

Voltage tolerance is ±5% only with simultaneous frequency tolerance of ±2%, and certain combinations are not allowed.

Ensures the site power quality is within acceptable limits for warranty and safe operation.

Source
“Voltage + / - 5% with a simultaneous frequency tolerance of + / - 2%”
Conformité

The equipment is classified as Class A, Group 1 for electromagnetic disturbance per EN55011.

Informs EMC planning and potential interference with sensitive equipment nearby.

Source
“Classification (disturbance) to EN55011: Class A, Group 1”
Safety

The refrigerant R-449A has a GHS hazard H280 (pressurized gas) and a GWP of 1397.

Requires awareness of pressurized refrigerant hazards and environmental impact for selection and service.

Source
“R-449A (A1, H280)”