Unichiller 230w Datasheet

Sommaire

  1. Overview p. 1
    Product introduction and main features of the chiller
  2. Spec pp. 1-2
    Technical data and performance specifications per DIN 12876
  3. p. 2 table, 12 rows
  4. p. 1 table, 35 rows

Technical data according to DIN 12876

ParameterValue
min. filling capacity18 l
Volume of expansion48 l
Overall dimensions WxDxH **940x1290x1130 mm
Net weight339 kg
Power supply (3 Phase)400V 3~ 50Hz
max. current (3 Phase)14 A
Fuse (3 phase)3x16A
Pressure equipment categoryI
Degree of ProtectionIP20
min. ambient temperature5 °C
max. ambient temperature40 °C
from Serial-No.:S541401

** Please respect space requirements. See operating conditions at www.huber-online.com

Technical data according to DIN 12876 — Unichiller 230w (Order-No.: 3039.0033.01)

ParameterValue
Operating temperature range-20...40 °C
temperature set point / display5,7" colour Touchscreen
Internal temperature sensorPt100
Sensor external connectionPt100
Temperature stability at -10°C0,5 K
Interface digitalEthernet, USB (Host u. Device), RS232
Safety classificationI / NFL
Cooling power at ambient temperature +20°C
at 20°C23 kW
at 0°C13,5 kW
at -10°C9 kW
at -20°C5,5 kW
Refrigeration machinewater-cooled, CFC- and HCFC-free
Refrigerant (ASHRAE, GHS)R-449A (A1, H280)
Global Warming Potential (GWP)1397
Refrigerant quantity1,6 kg
CO2 equivalent2,2 t
Circulation pumpE1
max. delivery54 l/min
max. delivery pressure3,5 bar
Förderleistung bei 0,17 bar52 l/min
Delivery at 0,5 bar48 l/min
Delivery at 1,0 bar42 l/min
Delivery at 1,5 bar36 l/min
Delivery at 2,0 bar28 l/min
Delivery at 2,5 bar21 l/min
Delivery at 3,0 bar11 l/min
Pump connectionG1 1/4 male
Cooling water connectionG3/4 male
Consumption at water 15°C, flow 20°C744 l/h
Consumption at water 15°C, flow 0°C540 l/h
Consumption at water 15°C, flow -10°C414 l/h
Consumption at water 15°C, flow -20°C324 l/h
min. cooling water differential pressure1 bar
max. cooling water pressure6 bar

Engineering Facts

11 facts · every claim linked to a page
Installation

The chiller is water-cooled and requires a cooling water supply with a minimum differential pressure of 1 bar and a maximum pressure of 6 bar.

Determines the required cooling water pump and piping design for installation.

Source
“min. cooling water differential pressure 1 bar”
Installation

The unit is designed for cooling water intake temperatures between 5°C and 20°C, and pipes must be insulated if condensation occurs.

Ensures proper cooling water supply conditions and prevents condensation issues.

Source
“This temperature control unit is designed to operate with cooling water intake temperature between 5°C and 20°C.”
Performance

Cooling power drops as cooling water temperature rises, and flow rate may need to increase.

Affects sizing of cooling water supply and expected performance.

Source
“As the cooling water temperature increases, drop in the cooling power should be expected, and also an increased cooling water flow rate possible.”
Installation

The cooling water circuit materials include copper, stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM, so only suitable cooling water should be used.

Guides water quality requirements to avoid corrosion or damage.

Source
“Materiels used in the cooling water circuit include; copper, Stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM.”
Conformité

Voltage tolerance is ±5% with simultaneous frequency tolerance of ±2%, but certain combinations are not allowed.

Ensures power supply quality meets equipment requirements.

Source
“Voltage + / - 5% with a simultaneous frequency tolerance of + / - 2%”
Installation

Three-phase devices with current consumption less than 63A are supplied with cable but without plug.

Affects electrical installation planning and plug procurement.

Source
“Three-phase devices with current consumption less than 63A --> with cable, without plug”
Safety

The refrigerant R-449A has a GHS classification of H280 (pressurized gas) and a GWP of 1397.

Important for safety handling and environmental compliance.

Source
“R-449A (A1, H280)”
Operation

The unit has a minimum filling capacity of 18 liters and an expansion volume of 48 liters.

Needed to prepare heat transfer fluid volume for operation.

Source
“min. filling capacity 18 l”
Installation

The unit requires a 400V 3-phase 50Hz power supply with max current 14A and fuse 3x16A.

Determines electrical supply requirements for installation.

Source
“Power supply (3 Phase) 400V 3~ 50Hz”
Operation

The ambient temperature must be between 5°C and 40°C for proper operation.

Ensures the installation location meets environmental conditions.

Source
“min. ambient temperature 5 °C”
Conformité

The equipment is classified as Class A, Group 1 for electromagnetic compatibility according to EN55011.

Important for EMC compliance in industrial environments.

Source
“Classification (disturbance) to EN55011: Class A, Group 1”