Unichiller GL 075w Technical Data

Sommaire

  1. Overview p. 1
    Product name and general introduction to the chiller unit
  2. Spec p. 2
    Technical data according to DIN 12876 standard
  3. p. 1 table, 27 rows

Unichiller GL 075w — Technical data according to DIN 12876

ParameterValue
Order-No.5506.0005.01
Operating temperature range-20...40 °C
temperature set point / display5,7" colour Touchscreen
Internal temperature sensorPt100
Sensor external connectionPt100
Temperature stability at -10°C0,2 K
Interface digitalEthernet, USB (Host u. Device), RS232
Safety classificationI / NFL
Cooling power at 20°C7,5 kW
Refrigeration machinewater-cooled, natural refrigerant
Refrigerant (ASHRAE, GHS)R-744 (A1, H280)
Global Warming Potential (GWP)1
UN-numberUN 2857
max. delivery50 l/min
max. delivery pressure3,5 bar
Pump connectionG1 1/4 male
Cooling water connectionG1/2 male
min. cooling water differential pressure1 bar
max. cooling water pressure6 bar
Volume of expansion9,5 l
Overall dimensions WxDxH **550x760x1455 mm
Power supply (3 Phase)460V 3~ 60Hz
Pressure equipment categoryI
Degree of ProtectionIP20
min. ambient temperature5 °C
max. ambient temperature40 °C
from Serial-No.1.0/26

** Technical details and dimensions are subject to change. No liability is accepted for errors or omissions. Illustrations can deviate from the original.

Engineering Facts

12 facts · every claim linked to a page
Installation

The chiller requires cooling water with a differential pressure of at least 1 bar and a maximum supply pressure of 6 bar.

Ensures the cooling water circuit can provide adequate flow and pressure without exceeding component limits.

Source
“min. cooling water differential pressure 1 bar”
Installation

The unit is water-cooled and must be connected to a cooling water supply with G1/2 male connections.

Determines the piping and fittings needed for installation.

Source
“Cooling water connection G1/2 male”
Operation

Cooling water inlet temperature must be between 5°C and 20°C; below the condensation point, cooling water pipes inside the unit must be insulated.

Operating outside this range can cause condensation, reduced cooling capacity, or increased water flow requirements.

Source
“This temperature control unit is designed to operate with cooling water intake temperature between 5°C and 20°C.”
Performance

Cooling power is rated at 7.5 kW at 20°C, but higher cooling water temperatures reduce capacity and may require increased flow.

Affects sizing and expected performance under actual site conditions.

Source
“As the cooling water temperature increases, drop in the cooling power should be expected, and also an increased cooling water flow rate possible.”
Installation

The cooling water circuit materials include copper, stainless steel 1.4401, MS, PA, PPE, PTFE, and EPDM, so only compatible cooling water should be used.

Prevents corrosion or damage from incompatible water chemistry.

Source
“Materiels used in the cooling water circuit include; copper, Stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM.”
Conformité

Voltage tolerance is ±5% with simultaneous frequency tolerance of ±2%, but combinations like -5% voltage and +2% frequency are not allowed.

Ensures the power supply stays within safe operating limits for the compressor and electronics.

Source
“Voltage + / - 5% with a simultaneous frequency tolerance of + / - 2%”
Installation

External branch circuit protection according to UL 489 is required.

Mandatory for electrical safety and code compliance in the installation.

Source
“External branch circuit protection according to UL 489 required.”
Installation

Three-phase devices with current consumption less than 63A are supplied with cable but without a plug.

Installer must provide a suitable plug or hardwired connection.

Source
“Three-phase devices with current consumption less than 63A --> with cable, without plug”
Safety

The unit uses R-744 refrigerant, which is classified A1 but has a GHS hazard statement H280 (contains gas under pressure).

Requires awareness of pressure hazard and proper handling during service.

Source
“R-744 (A1, H280)”
Operation

The ambient temperature must be between 5°C and 40°C for proper operation.

Operating outside this range may affect performance or reliability.

Source
“min. ambient temperature 5 °C”
Selection

The chiller has a maximum delivery of 50 l/min and maximum delivery pressure of 3.5 bar.

Needed to match the external circuit's flow and pressure requirements.

Source
“max. delivery 50 l/min”
Installation

The pump connection is G1 1/4 male.

Determines the fittings required for the process loop.

Source
“Pump connection G1 1/4 male”