Unichiller GL 400w

Sommaire

  1. Overview p. 1
    Product introduction and general description of the chiller
  2. Spec p. 2
    Technical data according to DIN 12876
  3. p. 1 table, 27 rows

Unichiller GL 400w — Technical data according to DIN 12876

ParameterValue
Order-No.5511.0001.01
Operating temperature range-20...40 °C
temperature set point / display5,7" colour Touchscreen
Internal temperature sensorPt100
Sensor external connectionPt100
Temperature stability at -10°C0,2 K
Interface digitalEthernet, USB (Host u. Device), RS232
Safety classificationI / NFL
Cooling power at 20°C40 kW
Refrigeration machinewater-cooled, natural refrigerant
Refrigerant (ASHRAE, GHS)R-744 (A1, H280)
Global Warming Potential (GWP)1
UN-numberUN 2857
max. delivery100 l/min
max. delivery pressure3,5 bar
Pump connectionG1 1/4 male
Cooling water connectionG1 1/4 male
min. cooling water differential pressure1 bar
max. cooling water pressure6 bar
Volume of expansion30 l
Overall dimensions WxDxH **730x804x1738 mm
Power supply (3 Phase)400V 3~ 50Hz
Pressure equipment categoryII
Degree of ProtectionIP20
min. ambient temperature5 °C
max. ambient temperature40 °C
from Serial-No.1.0/26

** Technical details and dimensions are subject to change. No liability is accepted for errors or omissions. Illustrations can deviate from the original.

Engineering Facts

11 facts · every claim linked to a page
Installation

The chiller requires cooling water with a differential pressure of at least 1 bar and a maximum of 6 bar at the connections.

Ensures the cooling water supply can meet the minimum flow requirement without exceeding pressure limits.

Source
“min. cooling water differential pressure 1 bar”
Installation

The unit is water-cooled and must be connected to a cooling water circuit with G1 1/4 male connections.

Determines the type of fittings and piping needed for installation.

Source
“Cooling water connection G1 1/4 male”
Operation

The cooling water inlet temperature must be between 5°C and 20°C; below the condensation point, cooling water pipes must be insulated.

Operating outside this range can reduce cooling capacity or cause condensation and requires insulation.

Source
“This temperature control unit is designed to operate with cooling water intake temperature between 5°C and 20°C.”
Performance

Cooling power drops as cooling water temperature increases, and the required cooling water flow rate may increase.

Helps in sizing the cooling water supply and predicting performance under varying conditions.

Source
“As the cooling water temperature increases, drop in the cooling power should be expected, and also an increased cooling water flow rate possible.”
Installation

The cooling water circuit materials include copper, stainless steel 1.4401, MS, PA, PPE, PTFE, and EPDM, so only suitable cooling water should be used.

Prevents corrosion and compatibility issues by guiding water quality and material selection.

Source
“Materiels used in the cooling water circuit include; copper, Stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM.”
Conformité

Voltage tolerance is ±5% with simultaneous frequency tolerance of ±2%, but combinations like -5% voltage and +2% frequency are not allowed.

Ensures the power supply quality meets the unit's requirements and avoids invalid tolerance combinations.

Source
“Voltage + / - 5% with a simultaneous frequency tolerance of + / - 2%”
Installation

Three-phase devices with current consumption less than 63A are supplied with cable but without a plug.

Requires the installer to provide a suitable plug and connection, affecting electrical installation planning.

Source
“Three-phase devices with current consumption less than 63A --> with cable, without plug”
Safety

The refrigerant is R-744 (CO2) with safety group A1 and GHS hazard H280, and the unit has pressure equipment category II.

Informs about refrigerant safety classification and pressure equipment requirements for handling and installation.

Source
“R-744 (A1, H280)”
Operation

The ambient temperature must be between 5°C and 40°C for proper operation.

Ensures the installation location maintains suitable ambient conditions to avoid performance issues.

Source
“min. ambient temperature 5 °C”
Installation

The unit has a degree of protection IP20, so it must be installed in a dry environment and protected from water ingress.

Guides environmental protection requirements for the installation site.

Source
“Degree of Protection IP20”
Conformité

The equipment complies with US-SNAP and all applicable EU laws, with US-SNAP end-use for industrial process refrigeration.

Confirms regulatory compliance for industrial process refrigeration applications.

Source
“This equipment is compliant to US-SNAP and all applicable EU laws.”