Introduces the Unichiller P007 chiller unit and its main features
Specp. 2
Lists technical data according to DIN 12876 standard
p. 2table, 8 rows
p. 1table, 32 rows
Technical data according to DIN 12876
Parameter
Value
Power supply requirement
208-240V 1~/2~ 50/60Hz
max. current
5 A
Fuse
16 A
Pressure equipment category
Art. 4.3 PED
Degree of Protection
IP20
min. ambient temperature
5 °C
max. ambient temperature
40 °C
from Serial-No.:
S591725
Unichiller P007 — Technical data according to DIN 12876
Parameter
Value
Order-No.
3012.0317.01
Operating temperature range
-20...40 °C
temperature set point / display
5,7" colour Touchscreen
Internal temperature sensor
Pt100
Sensor external connection
Pt100
Temperature stability at -10°C
0,5 K
Interface digital
Ethernet, USB (Host u. Device), RS232
Safety classification
I / NFL
Cooling power at ambient temperature +20°C
at 15°C
0,7 kW
at 0°C
0,55 kW
at -10°C
0,4 kW
at -20°C
0,2 kW
Refrigeration machine
air-cooled, natural refrigerant
Refrigerant (ASHRAE, GHS)
R-290 (A3, H220)
Global Warming Potential (GWP)
0,02
Refrigerant quantity
0,11 kg
CO2 equivalent
0,0 t
UN-number
UN 3358
Circulation pump:
B
max. delivery
25 l/min
max. delivery pressure
2,5 bar
Delivery at 0,5 bar
21 l/min
Delivery at 1,0 bar
17 l/min
Delivery at 1,5 bar
11 l/min
Delivery at 2,0 bar
6 l/min
Pump connection
G3/4 male
min. filling capacity
3,8 l
Volume of expansion
1,7 l
Overall dimensions WxDxH **
350x496x622 mm
Net weight
56 kg
sound pressure level +/- 4 dB(A)
58 dB(A)
** — сноска в исходном тексте не раскрыта; значение Cooling power at ambient temperature +20°C в тексте отсутствует (пустая строка).
Engineering Facts
14 facts · every claim linked to a page
Selection
The chiller is rated for an operating temperature range of -20 to 40 °C, so process requirements outside this window cannot be met.
Defines the process temperature envelope the unit can support during selection.
Source
“Operating temperature range
-20...40 °C”
Selection
Cooling capacity drops with lower temperature, from 0.7 kW at 15 °C to 0.2 kW at -20 °C, so the load must be checked at the actual set point.
Prevents oversizing or undersizing based on a single nominal capacity figure.
Source
“at 15°C
0,7 kW
at 0°C
0,55 kW
at -10°C
0,4 kW
at -20°C
0,2 kW”
Safety
The refrigerant is R-290 (A3, H220), a flammable natural refrigerant, which affects room ventilation and ignition-source planning.
Flammable refrigerant classification drives installation safety and site requirements.
Source
“R-290 (A3, H220)”
Installation
The pump connection is G3/4 male, so mating hoses and fittings must match this thread.
Determines the hose coupling and adapter selection for the hydraulic circuit.
Source
“Pump connection
G3/4 male”
Operation
The pump delivers 25 l/min max and 2.5 bar max pressure, with flow falling to 6 l/min at 2.0 bar, so system pressure drop must be matched to the pump curve.
Allows checking that the pump can overcome the external circuit pressure drop.