Unichiller P010

Sommaire

  1. Overview p. 1
    Product introduction and main features of the chiller
  2. Spec pp. 1-2
    Technical data according to DIN 12876
  3. p. 2 table, 4 rows
  4. p. 1 table, 32 rows

Technical data according to DIN 12876

ParameterValue
Degree of ProtectionIP20
min. ambient temperature5 °C
max. ambient temperature40 °C
from Serial-No.:1.0/20

Unichiller P010 — Technical data according to DIN 12876

ParameterValue
Order-No.3050.0024.01
Operating temperature range-20...40 °C
temperature set point / display5,7" colour Touchscreen
Internal temperature sensorPt100
Sensor external connectionPt100
Temperature stability at -10°C0,5 K
Interface digitalEthernet, USB (Host u. Device), RS232
Safety classificationI / NFL
Cooling power at ambient temperature +20°C
at 15°C1 kW
at 0°C0,8 kW
at -10°C0,5 kW
at -20°C0,15 kW
Refrigeration machineair-cooled, CFC- and HCFC-free
Refrigerant (ASHRAE, GHS)R-449A (A1, H280)
Global Warming Potential (GWP)1397
Circulation pump:B
max. delivery25 l/min
max. delivery pressure2,5 bar
Delivery at 0,5 bar21 l/min
Delivery at 1,0 bar17 l/min
Delivery at 1,5 bar11 l/min
Delivery at 2,0 bar6 l/min
Pump connectionG3/4 male
min. filling capacity3,8 l
Volume of expansion1,7 l
Overall dimensions WxDxH **350x496x622 mm
Net weight56 kg
sound pressure level +/- 4 dB(A)57 dB(A)
Power supply requirement208V 2~ 60Hz
max. current6,5 A
Fuse10 A

** Overall dimensions WxDxH

Engineering Facts

14 facts · every claim linked to a page
Selection

The unit is rated for an operating temperature range of -20 to 40 °C, so it cannot be used outside this span.

Defines the process temperature window the chiller can cover.

Source
“Operating temperature range -20...40 °C”
Performance

Cooling capacity drops sharply at low temperatures, from 1 kW at +15 °C to only 0.15 kW at -20 °C, so the load must be sized against the actual set point.

Prevents undersizing the chiller for low-temperature duty.

Source
“at 15°C 1 kW at 0°C 0,8 kW at -10°C 0,5 kW at -20°C 0,15 kW”
Installation

The unit requires a 208 V two-phase 60 Hz supply with a 10 A fuse, so the site power must match this configuration.

Determines electrical supply and protection needed at the installation site.

Source
“Power supply requirement 208V 2~ 60Hz max. current 6,5 A Fuse 10 A”
Installation

The pump connection is G3/4 male, so mating hoses and couplings must be G3/4 female.

Ensures correct plumbing connections are ordered.

Source
“Pump connection G3/4 male”
Operation

The pump delivers 25 l/min max and 2.5 bar max, with flow falling to 6 l/min at 2.0 bar, so high-pressure-drop loops must be checked against the pump curve.

Allows verification that the pump can overcome the external circuit pressure drop.

Source
“max. delivery 25 l/min max. delivery pressure 2,5 bar Delivery at 2,0 bar 6 l/min”
Safety

The refrigerant R-449A is classified A1 with GHS hazard H280, so refrigerant handling and ventilation precautions apply.

Informs safety assessment and refrigerant handling requirements.

Source
“R-449A (A1, H280)”
Conformité

The refrigerant has a GWP of 1397, which matters for F-gas reporting and environmental compliance.

Needed for refrigerant charge reporting and environmental permitting.

Source
“Global Warming Potential (GWP) 1397”
Installation

The unit is IP20 and must be installed where ambient temperature stays between 5 and 40 °C.

Constrains the installation location and room conditions.

Source
“Degree of Protection IP20 min. ambient temperature 5 °C max. ambient temperature 40 °C”
Performance

Cooling capacity data are valid at 20 °C room temperature and may drop if ambient rises, so derating is needed in warm rooms.

Prevents overestimating capacity in non-20 °C environments.

Source
“Output data valid for: Room temperature 20°C. If the ambient temperature rises, the cooling capacity may drop.”
Operation

The plastic pump is not resistant to acetone or polyglycols, so these fluids must be avoided or a more resistant pump specified.

Prevents chemical attack and pump failure from incompatible heat transfer fluids.

Source
“The plastic pump is not resistant against acetone and polyglycols (depending on the manufacturer).”
Operation

For freeze protection, water should be mixed with glysantine or ethylene glycol rather than relying on pure water.

Guides correct heat transfer fluid preparation for sub-zero operation.

Source
“It is recommended that water is mixed with either glysantine or ethylene glycol for freeze protection.”
Installation

Three-phase devices below 63 A are supplied with cable but without plug, so a plug must be provided on site.

Avoids installation delays from missing mains plug.

Source
“Three-phase devices with current consumption less than 63A --> with cable, without plug”
Conformité

Voltage and frequency tolerances are ±5% and ±2% respectively, but combined worst-case deviations in the same direction are not allowed.

Defines acceptable supply quality for reliable operation.

Source
“-5% voltage and +2% frequency -> not allowed!”
Conformité

The equipment is EMC Class A Group 1, which may require additional measures in residential or sensitive environments.

Informs EMC planning and site suitability.

Source
“Classification (disturbance) to EN55011: Class A, Group 1”