Unichiller P015w Datasheet

Sommaire

  1. Overview p. 1
    Introduces the Unichiller P015w chiller unit
  2. Spec p. 2
    Technical data according to DIN 12876 standard
  3. p. 2 table, 11 rows
  4. p. 1 table, 32 rows

Technical data according to DIN 12876

ParameterValue
Volume of expansion1,7 l
Overall dimensions WxDxH **350x496x622 mm
sound pressure level +/- 4 dB(A)54 dB(A)
Power supply requirement220-240V 1~/2~ 50Hz
max. current5,5 A
min. Fuse10A
max. Fuse16A
Degree of ProtectionIP20
min. ambient temperature5 °C
max. ambient temperature40 °C
from Serial-No.:1.1/24

** Please respect space requirements. See operating conditions at www.huber-online.com

Unichiller P015w — Technical data according to DIN 12876

ParameterValue
Order-No.3051.0025.01
Operating temperature range-20...40 °C
temperature set point / display5,7" colour Touchscreen
Internal temperature sensorPt100
Sensor external connectionPt100
Temperature stability at -10°C0,5 K
Interface digitalEthernet, USB (Host u. Device), RS232
Safety classificationI / NFL
Cooling power at ambient temperature +20°C
at 15°C1,5 kW
at 0°C1 kW
at -10°C0,7 kW
at -20°C0,3 kW
Refrigeration machinewater-cooled, CFC- and HCFC-free
Refrigerant (ASHRAE, GHS)R-449A (A1, H280)
Global Warming Potential (GWP)1397
Circulation pumpB
max. delivery25 l/min
max. delivery pressure2,5 bar
Delivery at 0,5 bar21 l/min
Delivery at 1,0 bar17 l/min
Delivery at 1,5 bar11 l/min
Delivery at 2,0 bar6 l/min
Pump connectionG3/4 male
Cooling water connectionG1/2 male
Consumption at water 15°C, flow 15°C68 l/h
Consumption at water 15°C, flow 0°C60 l/h
Consumption at water 15°C, flow -10°C48 l/h
Consumption at water 15°C, flow -20°C36 l/h
min. cooling water differential pressure1 bar
max. cooling water pressure6 bar
min. filling capacity3,8 l

Cooling power values at 15°C, 0°C, -10°C, -20°C are sub-rows of 'Cooling power at ambient temperature +20°C'.

Engineering Facts

13 facts · every claim linked to a page
Installation

The chiller requires cooling water with a differential pressure of at least 1 bar and no more than 6 bar, so the site water supply must be verified against these limits before installation.

Insufficient or excessive cooling water pressure will prevent proper operation or damage the water circuit.

Source
“min. cooling water differential pressure 1 bar max. cooling water pressure 6 bar”
Installation

The unit is designed for cooling water intake between 5°C and 20°C, and if the cooling water temperature falls below the condensation point the cooling water pipes in the unit must be insulated.

Operating outside the intake temperature range or omitting insulation causes condensation and possible damage.

Source
“This temperature control unit is designed to operate with cooling water intake temperature between 5°C and 20°C. When cooling water temperatures fall below the condensation point, it is essential to insulate the cooling water pipes in the unit.”
Performance

Cooling power drops as cooling water temperature rises, and the required cooling water flow rate may increase, so the water supply must be sized for the actual site conditions.

Sizing the cooling water supply only for nominal conditions can lead to insufficient cooling capacity in practice.

Source
“As the cooling water temperature increases, drop in the cooling power should be expected, and also an increased cooling water flow rate possible.”
Operation

The plastic pump is not resistant to acetone or polyglycols, so these media must not be used; for freeze protection a water mixture with glysantine or ethylene glycol is recommended.

Using incompatible media will damage the pump and void reliable operation.

Source
“Special Case: Acetone and Polyglycol: The plastic pump is not resistant against acetone and polyglycols (depending on the manufacturer). It is recommended that water is mixed with either glysantine or ethylene glycol for freeze protection.”
Installation

The cooling water circuit contains copper, stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM, so only suitable cooling water may be used to avoid corrosion.

Aggressive or untreated cooling water can corrode the mixed-material water circuit.

Source
“Materiels used in the cooling water circuit include; copper, Stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM. Please use suitable cooling water.”
Conformité

Voltage tolerance is ±5% only with a simultaneous frequency tolerance of ±2%, and the combination of -5% voltage with +2% frequency is explicitly not allowed.

The site power supply must be checked against this combined tolerance to avoid invalid operating conditions.

Source
“Voltage + / - 5% with a simultaneous frequency tolerance of + / - 2% Example -5% voltage and +2% frequency -> not allowed!”
Installation

Single/two-phase devices are delivered with a power cable and country-specific plug that must be specified at ordering, while three-phase devices below 63A come with cable but without plug.

The plug configuration must be specified at order time and the site must provide the matching receptacle or a plug for three-phase units.

Source
“1. Single / two-phase devices (100V to 240V) --> with power cable and country-specific plug (please specify when ordering) 2. Three-phase devices with current consumption less than 63A --> with cable, without plug”
Installation

The unit has IP20 protection and requires ambient temperature between 5°C and 40°C, so it must be installed in a dry indoor location within this temperature range.

IP20 offers no water protection and the ambient limits define the permissible installation environment.

Source
“Degree of Protection IP20 min. ambient temperature 5 °C max. ambient temperature 40 °C”
Selection

The refrigeration machine is water-cooled and uses R-449A refrigerant with GWP 1397, which matters for refrigerant regulations and environmental compliance at the site.

The refrigerant type and GWP affect regulatory obligations and environmental reporting for the installation.

Source
“Refrigeration machine water-cooled, CFC- and HCFC-free Refrigerant (ASHRAE, GHS) R-449A (A1, H280) Global Warming Potential (GWP) 1397”
Selection

The pump delivers 25 l/min maximum and only 6 l/min at 2.0 bar, so applications with high pressure drop must be checked against the pump curve.

High-pressure-drop circuits may receive insufficient flow, so the pump curve must be matched to the application.

Source
“max. delivery 25 l/min max. delivery pressure 2,5 bar Delivery at 2,0 bar 6 l/min”
Installation

The minimum filling capacity is 3.8 l and the expansion vessel volume is 1.7 l, which determines the minimum heat transfer fluid charge and expansion allowance.

Underfilling below the minimum capacity prevents proper operation and expansion control.

Source
“min. filling capacity 3,8 l”
Operation

The unit is factory-set for 220-240V 1~/2~ 50Hz with maximum current 5.5 A and requires a fuse between 10A and 16A.

The electrical supply and protective device must be sized within these limits for safe connection.

Source
“Power supply requirement 220-240V 1~/2~ 50Hz max. current 5,5 A min. Fuse 10A max. Fuse 16A”
Conformité

The equipment is classified for electromagnetic disturbance as EN55011 Class A Group 1, which restricts its use to industrial environments.

Class A equipment may cause interference in residential areas and is intended for industrial use only.

Source
“Classification (disturbance) to EN55011: Class A, Group 1”