Sommaire

  1. Overview p. 1
    Product introduction and general description of the chiller
  2. Spec pp. 1-2
    Technical data and performance specifications per DIN 12876
  3. p. 2 table, 3 rows
  4. p. 1 table, 38 rows

Technical data according to DIN 12876

ParameterValue
min. ambient temperature5 °C
max. ambient temperature40 °C
from Serial-No.:S458965

Technical data according to DIN 12876

ParameterValue
Operating temperature range-10...100 °C
temperature set point / displaydigital
Internal temperature sensorPt100
Resolution of display0,1 K
Interface digitalUSB (Device), RS232
Interface
Temperature stability at -10°C0,2 K
Alarm messageoptic, acoustic
Safety classificationI / NFL
Heating power at 230V2 kW
Heating power at 220V1,8 kW
Heating power at 208V1,6 kW
Cooling power at 15°C2,2 kW
Cooling power at 0°C1,6 kW
Cooling power at -10°C1 kW
Refrigeration machineair-cooled, CFC- and HCFC-free
Refrigerant (ASHRAE, GHS)R-449A (A1, H280)
Global Warming Potential (GWP)1397
Refrigerant quantity0,56 kg
CO2 equivalent0,8 t
Circulation pumpB
at 0,5 bar21 l/min
at 1,0 bar17 l/min
at 1,5 bar11 l/min
at 2,0 bar6 l/min
max. delivery25 l/min
max. delivery pressure2,5 bar
Pump connectionG3/4 male
min. filling capacity3,8 l
expansion tank1,7 l
Overall dimensions WxDxH **460x590x743 mm
Net weight83 kg
sound pressure level +/- 4 dB(A)65 dB(A)
Power supply requirement208-230V 1~/2~ 60Hz
max. current15 A
Fuse16 A
Pressure equipment categoryArt. 4.3 PED
Degree of ProtectionIP20

** Overall dimensions WxDxH

Engineering Facts

14 facts · every claim linked to a page
Selection

The chiller is rated for a -10 to 100 °C operating range, so it suits both sub-ambient cooling and moderate heating duties.

Defines whether the unit can cover the required process temperature window before specifying it.

Source
“Operating temperature range -10...100 °C”
Performance

Cooling capacity drops with lower temperature (2.2 kW at 15 °C, 1.6 kW at 0 °C, 1 kW at -10 °C), so size the load against the actual set point.

Prevents undersizing when the application runs cold rather than near ambient.

Source
“at 15°C 2,2 kW at 0°C 1,6 kW at -10°C 1 kW”
Installation

The unit requires a 208-230 V single/two-phase 60 Hz supply with 15 A max current and a 16 A fuse.

Determines electrical feed, breaker sizing and whether a 60 Hz site is compatible.

Source
“Power supply requirement 208-230V 1~/2~ 60Hz max. current 15 A Fuse 16 A”
Installation

Voltage and frequency tolerances must be met simultaneously (±5% V with ±2% f), and a -5% voltage with +2% frequency combination is not allowed.

Guides site power quality checks to avoid invalid supply conditions and warranty issues.

Source
“-5% voltage and +2% frequency -> not allowed!”
Installation

Ambient temperature must stay between 5 and 40 °C, and rising ambient temperature may reduce cooling capacity.

Affects room ventilation/placement planning and realistic capacity expectations.

Source
“If the ambient temperature rises, the cooling capacity may drop.”
Safety

The refrigerant is R-449A (A1, H280) with GWP 1397 and 0.56 kg charge, giving 0.8 t CO2 equivalent.

Needed for refrigerant handling, leak reporting and environmental compliance planning.

Source
“R-449A (A1, H280)”
Operation

The plastic pump is not resistant to acetone or polyglycols, so these media must be avoided or a more resistant pump specified.

Prevents chemical attack on the pump when selecting the heat transfer fluid.

Source
“The plastic pump is not resistant against acetone and polyglycols (depending on the manufacturer).”
Operation

For freeze protection, water should be mixed with glysantine or ethylene glycol rather than used pure.

Ensures correct coolant preparation for sub-zero operation.

Source
“It is recommended that water is mixed with either glysantine or ethylene glycol for freeze protection.”
Installation

Single/two-phase devices are delivered with a power cable and country-specific plug, which must be specified at ordering.

Avoids a missing or wrong plug on delivery and rework at commissioning.

Source
“with power cable and country-specific plug (please specify when ordering)”
Conformité

The unit is EMC classified as Class A, Group 1 to EN55011, which has implications for use in residential environments.

Class A equipment may not be suitable for domestic/low-voltage residential supply without mitigation.

Source
“Classification (disturbance) to EN55011: Class A, Group 1”
Installation

The unit has IP20 protection, so it must be installed in a dry, protected location away from water ingress.

Restricts placement options and enclosure requirements at the installation site.

Source
“Degree of Protection IP20”
Maintenance

The expansion vessel has a 1.7 l volume and a minimum filling capacity of 3.8 l applies to the system.

Sets the minimum coolant charge and expansion headroom for loop design.

Source
“min. filling capacity 3,8 l expansion tank 1,7 l”
Performance

The pump delivers 21 l/min at 0.5 bar down to 6 l/min at 2.0 bar, with 25 l/min max delivery and 2.5 bar max pressure.

Allows checking whether the pump curve covers the pressure drop of the external circuit.

Source
“max. delivery 25 l/min max. delivery pressure 2,5 bar”
Safety

The unit includes adjustable overtemperature protection according to DIN 12876 and safety classification I/NFL.

Confirms built-in thermal safety for unattended or sensitive processes.

Source
“With adjustable overtemperature protection according to DIN 12876.”