Product introduction and key features of the Unichiller P022w
Specpp. 1-2
Technical data and performance specifications according to DIN 12876
p. 2table, 11 rows
p. 1table, 31 rows
Technical data according to DIN 12876
Parameter
Value
Volume of expansion
1,7 l
Overall dimensions WxDxH **
420x487x579 mm
Net weight
67 kg
sound pressure level +/- 4 dB(A)
60 dB(A)
Power supply requirement
208-230V 1~/2~ 60Hz
max. current
12,5 A
Fuse
2x16 A
Degree of Protection
IP20
min. ambient temperature
5 °C
max. ambient temperature
40 °C
from Serial-No.:
1.1/24
** Please respect space requirements. See operating conditions at www.huber-online.com
Technical data according to DIN 12876
Parameter
Unichiller P022w
Order-No.
3010.0137.01
Operating temperature range
-10...40 °C
temperature set point / display
5,7" colour Touchscreen
Internal temperature sensor
Pt100
Sensor external connection
Pt100
Temperature stability at -10°C
0,5 K
Interface digital
Ethernet, USB (Host u. Device), RS232
Safety classification
I / NFL
Cooling power at ambient temperature +20°C at 15°C
2,2 kW
Cooling power at ambient temperature +20°C at 0°C
1,6 kW
Cooling power at ambient temperature +20°C at -10°C
1 kW
Refrigeration machine
water-cooled, CFC- and HCFC-free
Refrigerant (ASHRAE, GHS)
R-449A (A1, H280)
Global Warming Potential (GWP)
1397
Refrigerant quantity
0,45 kg
CO2 equivalent
0,6 t
Circulation pump
B
max. delivery
25 l/min
max. delivery pressure
2,5 bar
Delivery at 0,5 bar
21 l/min
Delivery at 1,0 bar
17 l/min
Delivery at 1,5 bar
11 l/min
Delivery at 2,0 bar
6 l/min
Pump connection
G3/4 male
Cooling water connection
G1/2 male
Consumption at water 15°C, flow 15°C
95 l/h
Consumption at water 15°C, flow 0°C
96 l/h
Consumption at water 15°C, flow -10°C
77 l/h
min. cooling water differential pressure
1 bar
max. cooling water pressure
6 bar
min. filling capacity
3,8 l
Engineering Facts
14 facts · every claim linked to a page
Installation
The chiller requires cooling water with a minimum differential pressure of 1 bar and must not exceed 6 bar, so the site water supply must be verified before installation.
Insufficient or excessive cooling water pressure will prevent correct operation or damage the unit.
Source
“min. cooling water differential pressure 1 bar”
Installation
The unit is designed for cooling water intake between 5°C and 20°C, and pipes must be insulated if the water temperature can fall below the condensation point.
Operating outside this range reduces cooling power and risks condensation, so the site cooling water temperature must be checked.
Source
“This temperature control unit is designed to operate with cooling water intake temperature between 5°C and 20°C.”
Performance
Cooling power is rated at 2.2 kW at 15°C, 1.6 kW at 0°C and 1 kW at -10°C, so capacity must be selected against the actual process temperature.
Cooling capacity drops significantly at lower temperatures, affecting process sizing.
Source
“Cooling power at ambient temperature +20°C”
Installation
The power supply must be 208-230V single or two-phase at 60 Hz with a maximum current of 12.5 A and 2x16 A fusing.
Electrical supply and protection must be provisioned accordingly before commissioning.
Source
“208-230V 1~/2~ 60Hz”
Conformité
Voltage and frequency tolerances are ±5% and ±2% respectively, but they may not be combined in opposite directions.
Supply quality must be verified to avoid invalid operating conditions and equipment stress.
Source
“Voltage + / - 5% with a simultaneous frequency tolerance of + / - 2%”
Safety
The plastic pump is not resistant to acetone or polyglycols, so these media must be avoided or a more resistant pump ordered.
Using incompatible media will damage the pump and void reliable operation.
Source
“The plastic pump is not resistant against acetone and polyglycols (depending on the manufacturer).”
Operation
For freeze protection, water should be mixed with glysantine or ethylene glycol rather than relying on pure water.
Prevents freezing and damage in the cooling circuit at low temperatures.
Source
“It is recommended that water is mixed with either glysantine or ethylene glycol for freeze protection.”
Installation
The refrigerant is R-449A with GWP 1397 and a charge of 0.45 kg, giving 0.6 t CO2 equivalent, which matters for leak reporting and environmental compliance.
Refrigerant type and charge determine environmental obligations and service requirements.
Source
“Refrigerant quantity 0,45 kg”
Installation
The unit has IP20 protection and must be installed in an ambient temperature between 5°C and 40°C.
IP20 offers no water protection, and ambient limits constrain the installation location.
Source
“min. ambient temperature 5 °C”
Installation
The pump connections are G3/4 male and cooling water connections are G1/2 male, so matching hose couplings must be prepared.
Correct fittings are needed for leak-free hydraulic connection.
Source
“Pump connection G3/4 male”
Performance
The circulation pump delivers 25 l/min maximum and 2.5 bar maximum pressure, with flow falling to 6 l/min at 2.0 bar.
Pump curve must be checked against system pressure drop to ensure adequate flow.
Source
“max. delivery pressure 2,5 bar”
Installation
The minimum filling capacity is 3.8 l and expansion volume is 1.7 l, which sets the heat transfer fluid charge and expansion allowance.
Determines fluid volume needed and expansion vessel sizing.
Source
“min. filling capacity 3,8 l”
Conformité
The equipment is classified Class A, Group 1 for electromagnetic disturbance to EN55011.
Class A equipment may require EMC measures in residential or sensitive environments.
Source
“Classification (disturbance) to EN55011: Class A, Group 1”
Installation
Three-phase devices below 63 A are supplied with cable but without a plug, so a suitable plug must be provided on site.
Affects electrical installation planning and commissioning.
Source
“Three-phase devices with current consumption less than 63A --> with cable, without plug”