Product name and general description of the chiller unit
Specpp. 1-2
Technical data, performance values and operating parameters
Specp. 2
Detailed specifications according to DIN 12876 standard
p. 2table, 9 rows
p. 1table, 38 rows
Technical data according to DIN 12876
Parameter
Value
sound pressure level +/- 4 dB(A)
60 dB(A)
Power supply requirement
220-240V 1~ 50Hz
max. current
15,9 A
Fuse
16 A
Pressure equipment category
Art. 4.3 PED
Degree of Protection
IP20
min. ambient temperature
5 °C
max. ambient temperature
40 °C
from Serial-No.:
1.1/24
Technical data according to DIN 12876 — Unichiller P025w-H OLÉ (Order-No.: 3052.0002.98)
Parameter
Value
Operating temperature range
-10...100 °C
temperature set point / display
digital
Internal temperature sensor
Pt100
Resolution of display
0,1 K
Interface digital
USB (Device), RS232
Interface
Temperature stability at -10°C
0,2 K
Alarm message
optic, acoustic
Safety classification
III / FL
Heating power at 240V
2,1 kW
Heating power at 230V
2 kW
Heating power at 220V
1,8 kW
Cooling power at 15°C
2,5 kW
Cooling power at 0°C
2 kW
Cooling power at -10°C
1,2 kW
Refrigeration machine
water-cooled, CFC- and HCFC-free
Refrigerant (ASHRAE, GHS)
R-449A (A1, H280)
Global Warming Potential (GWP)
1397
Refrigerant quantity
0,4 kg
CO2 equivalent
0,6 t
Circulation pump
B
at 0,5 bar
21 l/min
at 1,0 bar
17 l/min
at 1,5 bar
11 l/min
at 2,0 bar
6 l/min
max. delivery
25 l/min
max. delivery pressure
2,5 bar
Pump connection
G3/4 male
Consumption at water 15°C, flow 15°C
84 l/h
Consumption at water 15°C, flow 0°C
84 l/h
Consumption at water 15°C, flow -10°C
72 l/h
Cooling water connection
G1/2 male
min. cooling water differential pressure
1 bar
max. cooling water pressure
6 bar
min. filling capacity
3,8 l
expansion tank
1,7 l
Overall dimensions WxDxH **
420x487x579 mm
Net weight
67 kg
** см. сноску в оригинале (Overall dimensions WxDxH **)
Engineering Facts
12 facts · every claim linked to a page
Installation
The chiller requires cooling water with a differential pressure of at least 1 bar and a maximum of 6 bar at the G1/2 male connection.
Ensures the water supply system can meet the minimum pressure for proper cooling and avoid overpressure damage.
Source
“min. cooling water differential pressure 1 bar”
Installation
The unit is water-cooled and requires a cooling water intake temperature between 5°C and 20°C; below the condensation point, cooling water pipes must be insulated.
Defines the allowable cooling water temperature range and insulation requirement to prevent condensation and ensure performance.
Source
“This temperature control unit is designed to operate with cooling water intake temperature between 5°C and 20°C.”
Operation
Cooling power drops as cooling water temperature increases, and the cooling water flow rate may need to be increased.
Helps predict performance changes and size the cooling water supply accordingly.
Source
“As the cooling water temperature increases, drop in the cooling power should be expected, and also an increased cooling water flow rate possible.”
Safety
The plastic pump is not resistant to acetone and polyglycols; for freeze protection, use water mixed with glysantine or ethylene glycol instead.
Prevents chemical attack on the pump and ensures compatibility of the heat transfer fluid.
Source
“The plastic pump is not resistant against acetone and polyglycols (depending on the manufacturer).”
Installation
The power supply must be 220-240V 1~ 50Hz with a maximum current of 15.9 A and a 16 A fuse.
Ensures the electrical supply and protection are correctly rated for safe operation.
Source
“Power supply requirement 220-240V 1~ 50Hz”
Installation
Voltage tolerance is ±5% with a simultaneous frequency tolerance of ±2%, but combinations like -5% voltage and +2% frequency are not allowed.
Guides the acceptable power quality and prevents invalid supply conditions that could damage the unit.
Source
“Voltage + / - 5% with a simultaneous frequency tolerance of + / - 2%”
Operation
The maximum ambient temperature for operation is 40°C and the minimum is 5°C.
Ensures the installation location maintains ambient conditions within the allowed range for reliable operation.
Source
“max. ambient temperature 40 °C”
Installation
The cooling water circuit materials include copper, stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM, so suitable cooling water must be used.
Helps assess water quality and corrosion compatibility with the wetted materials.
Source
“Materiels used in the cooling water circuit include; copper, Stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM.”
Selection
The chiller has a maximum delivery pressure of 2.5 bar and a maximum flow of 25 l/min, with flow decreasing at higher pressures.
Allows proper sizing of the external circuit and ensures the pump can overcome system pressure drops.
Source
“max. delivery pressure 2,5 bar”
Selection
The refrigerant is R-449A with a GWP of 1397 and a charge of 0.4 kg, corresponding to 0.6 t CO2 equivalent.
Relevant for environmental compliance, refrigerant handling, and leak reporting obligations.
Source
“Refrigerant (ASHRAE, GHS) R-449A (A1, H280)”
Installation
The unit has a protection degree of IP20, so it must be installed in a dry environment protected from water ingress.
Determines the required environmental protection at the installation site.
Source
“Degree of Protection IP20”
Operation
The OLÉ controller includes an autostart function for power failure, which can restart the unit automatically after power restoration.
Affects operational safety and process continuity; may require additional precautions depending on the application.