Induztr
Résumé technique Visionneuse de documents Modèles associés

Unichiller P050-H Datasheet

Sommaire

  1. Overview p. 1
    Introduces the Unichiller P050-H chiller unit and its main features
  2. Spec p. 2
    Technical data according to DIN 12876 including performance and dimensions
  3. p. 2 table, 13 rows
  4. p. 1 table, 32 rows

Technical data according to DIN 12876

ParameterValue
min. filling capacity18 l
Volume of expansion48 l
Overall dimensions WxDxH **740x1160x1165 mm
Net weight300 kg
sound pressure level +/- 4 dB(A)62 dB(A)
Power supply (3 Phase)400V 3~ 50Hz
max. current (3 Phase)12 A
Fuse (3 phase)3x16A
Pressure equipment categoryArt. 4.3 PED
Degree of ProtectionIP20
min. ambient temperature5 °C
max. ambient temperature40 °C
from Serial-No.:S315756

** Please respect space requirements. See operating conditions at www.huber-online.com

Technical data according to DIN 12876

ParameterUnichiller P050-H
Order-No.3038.0005.01
Operating temperature range-20...100 °C
temperature set point / display5,7" colour Touchscreen
Internal temperature sensorPt100
Sensor external connectionPt100
Temperature stability at -10°C0,2 K
Interface digitalEthernet, USB (Host u. Device), RS232
Safety classificationIII / FL
Heating power2 kW
Cooling power at ambient temperature +20°C
at 100°C5 kW
at 20°C5 kW
at 0°C3,4 kW
at -10°C2,1 kW
at -20°C0,7 kW
Refrigeration machineair-cooled, CFC- and HCFC-free
Refrigerant (ASHRAE, GHS)R-449A (A1, H280)
Global Warming Potential (GWP)1397
Refrigerant quantity1,03 kg
CO2 equivalent1,4 t
Circulation pump:E2
max. delivery130 l/min
max. delivery pressure5,7 bar
Delivery at 2,4 bar116 l/min
Delivery at 2,5 bar115 l/min
Delivery at 3,0 bar108 l/min
Delivery at 3,5 bar100 l/min
Delivery at 4,0 bar92 l/min
Delivery at 4,5 bar80 l/min
Delivery at 5,0 bar59 l/min
Delivery at 5,5 bar20 l/min
Pump connectionG1 1/4 male

Engineering Facts

14 facts · every claim linked to a page
Selection

The chiller is intended only for externally closed applications, so an open-loop or vented process circuit is not permitted.

Determines whether the unit can be applied to the process loop at all.

Source
“For externally closed applications.”
Selection

Cooling capacity falls sharply at low temperature, from 5 kW at 20 °C to 0.7 kW at -20 °C, so the load must be checked at the actual set point.

Sizing on nominal capacity would undersize the unit at low-temperature duty.

Source
“at -20°C 0,7 kW”
Performance

Cooling capacity is rated at 20 °C room temperature and drops if ambient rises, so the installed ambient must be considered.

Prevents overestimating available cooling in a warm plant room.

Source
“If the ambient temperature rises, the cooling capacity may drop.”
Installation

The unit requires a 400 V three-phase supply with 12 A max current and 3x16 A fusing.

Defines the electrical supply and protection needed at the installation site.

Source
“max. current (3 Phase) 12 A”
Installation

Three-phase devices below 63 A are delivered with cable but without a plug, so a plug must be sourced separately.

Affects electrical installation planning and bill of materials.

Source
“Three-phase devices with current consumption less than 63A --> with cable, without plug”
Installation

Voltage and frequency tolerances are linked: -5% voltage with +2% frequency is not allowed.

Site power quality must be checked against the combined tolerance rule.

Source
“-5% voltage and +2% frequency -> not allowed!”
Installation

The unit needs 740x1160x1165 mm of space and weighs 300 kg, so floor loading and access must be verified.

Determines rigging, floor loading and room layout.

Source
“740x1160x1165 mm”
Installation

Ambient temperature must stay between 5 °C and 40 °C for correct operation.

Sets the allowable plant-room conditions for installation.

Source
“max. ambient temperature 40 °C”
Safety

The refrigerant R-449A is classified A1 but carries GHS hazard H280 (pressurised gas), with 1.4 t CO2 equivalent.

Affects refrigerant handling, labelling and leak-response procedures.

Source
“R-449A (A1, H280)”
Safety

The unit has adjustable overtemperature protection per DIN 12876 and safety classification III/FL.

Confirms the safety cut-out required for unattended operation.

Source
“With adjustable overtemperature protection according to DIN 12876.”
Operation

The pump delivers 130 l/min max at 5.7 bar, but flow collapses to 20 l/min at 5.5 bar, so system pressure drop must be matched.

High-pressure-drop circuits may starve the pump despite the high nominal head.

Source
“max. delivery pressure 5,7 bar”
Operation

The expansion tank is atmospheric and open, with 48 l expansion volume and 18 l minimum filling.

Sets the minimum fill and expansion allowance for the loop.

Source
“Volume of expansion 48 l”
Conformité

The unit is EMC Class A Group 1, which is industrial-grade and may not suit residential environments.

Determines EMC suitability of the installation location.

Source
“Classification (disturbance) to EN55011: Class A, Group 1”
Conformité

The pressure equipment falls under PED Article 4.3, so no CE pressure marking applies but documentation is still required.

Affects compliance documentation and acceptance at handover.

Source
“Pressure equipment category Art. 4.3 PED”