Unichiller P230w-H Datasheet

Sommaire

  1. Overview p. 1
    Product introduction and key features of the chiller
  2. Spec pp. 1-2
    Technical data and performance specifications per DIN 12876
  3. p. 2 table, 12 rows
  4. p. 1 table, 31 rows

Technical data according to DIN 12876

ParameterValue
Consumption at water 15°C, flow -20°C324 l/h
min. cooling water differential pressure1 bar
max. cooling water pressure6 bar
min. filling capacity18 l
Volume of expansion48 l
Overall dimensions WxDxH **940x1290x1130 mm
Power supply (3 Phase)460V 3~ 60Hz
Pressure equipment categoryI
Degree of ProtectionIP20
min. ambient temperature5 °C
max. ambient temperature40 °C
from Serial-No.:1.0/21

** Please respect space requirements. See operating conditions at www.huber-online.com

Unichiller P230w-H — Technical data according to DIN 12876

ParameterValue
Order-No.3039.0040.01
Operating temperature range-20...100 °C
temperature set point / display5,7" colour Touchscreen
Internal temperature sensorPt100
Sensor external connectionPt100
Temperature stability at -10°C0,2 K
Interface digitalEthernet, USB (Host u. Device), RS232
Safety classificationIII / FL
Heating power2 kW
Cooling power at ambient temperature +20°C at 100°C23 kW
Cooling power at ambient temperature +20°C at 20°C23 kW
Cooling power at ambient temperature +20°C at 0°C13,5 kW
Cooling power at ambient temperature +20°C at -10°C9 kW
Cooling power at ambient temperature +20°C at -20°C5,5 kW
Refrigeration machinewater-cooled, CFC- and HCFC-free
Refrigerant (ASHRAE, GHS)R-449A (A1, H280)
Global Warming Potential (GWP)1397
Circulation pumpE2
max. delivery140 l/min
max. delivery pressure5,3 bar
Delivery at 2,8 bar126 l/min
Delivery at 3,0 bar124 l/min
Delivery at 3,5 bar115 l/min
Delivery at 4,0 bar103 l/min
Delivery at 4,5 bar83 l/min
Delivery at 5,0 bar43 l/min
Pump connectionG1 1/4 male
Cooling water connectionG3/4 male
Consumption at water 15°C, flow 20°C744 l/h
Consumption at water 15°C, flow 0°C540 l/h
Consumption at water 15°C, flow -10°C414 l/h

Engineering Facts

14 facts · every claim linked to a page
Installation

The chiller requires a three-phase 460 V 60 Hz supply, so the site must provide this power configuration.

Determines electrical infrastructure and whether a transformer or different model is needed.

Source
“Power supply (3 Phase) 460V 3~ 60Hz”
Installation

External branch circuit protection according to UL 489 is required, so a suitable breaker must be installed in the supply circuit.

Necessary for code-compliant and safe electrical installation.

Source
“External branch circuit protection according to UL 489 required.”
Installation

Three-phase devices drawing less than 63 A are supplied with cable but without a plug, so the installer must provide a suitable plug or hard-wired connection.

Affects electrical termination planning and bill of materials.

Source
“Three-phase devices with current consumption less than 63A --> with cable, without plug”
Installation

The unit is designed for cooling water intake between 5 °C and 20 °C, so the facility water supply must stay within this range.

Out-of-range cooling water can reduce cooling capacity or cause condensation issues.

Source
“This temperature control unit is designed to operate with cooling water intake temperature between 5°C and 20°C.”
Installation

If cooling water temperature falls below the condensation point, the cooling water pipes in the unit must be insulated.

Prevents condensation and corrosion damage inside the unit.

Source
“When cooling water temperatures fall below the condensation point, it is essential to insulate the cooling water pipes in the unit.”
Installation

The cooling water circuit contains copper, stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM, so the water quality must be compatible with these materials.

Guides selection of water treatment and avoids corrosion or fouling.

Source
“Materiels used in the cooling water circuit include; copper, Stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM.”
Installation

The cooling water differential pressure must be at least 1 bar and must not exceed 6 bar.

Ensures adequate flow for heat rejection without overpressurizing the circuit.

Source
“min. cooling water differential pressure 1 bar”
Installation

The maximum cooling water pressure is 6 bar, so the supply must be regulated or limited accordingly.

Protects the cooling water circuit from overpressure damage.

Source
“max. cooling water pressure 6 bar”
Installation

The unit requires a minimum filling capacity of 18 l and has an expansion volume of 48 l.

Determines heat transfer fluid volume needed and expansion tank sizing.

Source
“min. filling capacity 18 l”
Installation

The overall dimensions are 940x1290x1130 mm, so the installation space must accommodate this footprint and height.

Needed for floor space planning and access clearance.

Source
“Overall dimensions WxDxH ** 940x1290x1130 mm”
Operation

The ambient temperature must remain between 5 °C and 40 °C for proper operation.

Operating outside this range may affect performance and reliability.

Source
“min. ambient temperature 5 °C”
Operation

The maximum ambient temperature is 40 °C, so the room must not exceed this limit.

High ambient temperature reduces cooling capacity and can trip the unit.

Source
“max. ambient temperature 40 °C”
Performance

Cooling power drops as temperature decreases, from 23 kW at 20 °C to 5.5 kW at -20 °C, so capacity must be checked at the actual operating temperature.

Prevents undersizing when selecting the chiller for low-temperature duty.

Source
“at -20°C 5,5 kW”
Performance

The circulation pump delivers a maximum of 140 l/min at up to 5.3 bar, so system pressure drop must be checked against the pump curve.

Ensures adequate flow through the external circuit at its pressure drop.

Source
“max. delivery 140 l/min”