Unimotive 10w Immersion Circulator

Sommaire

  1. Overview p. 1
    Product introduction and main features of the circulator
  2. Spec pp. 1-2
    Technical data, performance figures and operating limits
  3. Spec p. 2
    Detailed parameters according to DIN 12876 standard
  4. p. 2 table, 25 rows
  5. p. 1 table, 27 rows

Technical data according to DIN 12876

ParameterValue
Delivery at 1,0 bar179 l/min
Delivery at 2,0 bar154 l/min
Delivery at 3,0 bar124 l/min
Delivery at 4,0 bar87 l/min
Delivery at 5,0 bar42 l/min
Pump connectionM38x1,5 male
max. permissible kin. viscosity200 mm²/s
Cooling water connectionG3/4 male
Consumption at water 15°C, flow 0°C620 l/h
min. cooling water differential pressure1 bar
max. cooling water differential pressure4 bar
max. cooling water pressure6 bar
max. System pressure6 bar
min. filling capacity8,2 l
Filling capacity expansion tank19,9 l
Overall dimensions WxDxH **730x804x1738 mm
Net weight440 kg
Power supply requirement400V 3~ 50Hz
max. current31,5 A
Fuse3x32A
Pressure equipment categoryI
Degree of ProtectionIP20
min. ambient temperature5 °C
max. ambient temperature40 °C
from Serial-No.:S588653

** Please respect space requirements. See operating conditions at www.huber-online.com

Technical data according to DIN 12876

ParameterValue
Order-No.5004.0001.01
Operating temperature range-45...95 °C
Temperature stability at -10°C0,05 K
temperature set point / display5,7" colour Touchscreen
Resolution of display0,01 K
Internal temperature sensorPt100
Sensor external connectionPt100
Interface digitalEthernet, USB (Host u. Device), RS232
digital inputECS ONE
digital outputPOKO ONE
Interface analogoptional
Alarm messageoptic, acoustic, relay
Safety classificationIII / FL
Heating power12 kW
Cooling power with Water/ Glysantin 40/60 at 20°C14 kW
Cooling power with Water/ Glysantin 40/60 at 0°C10 kW
at -20°C5 kW
at -30°C2,1 kW
at -40°C0,8 kW
Refrigeration machinewater-cooled, CFC- and HCFC-free
Refrigerant (ASHRAE, GHS)R-452A (A1, H280)
Global Warming Potential (GWP)2141
Refrigerant quantity2,9 kg
CO2 equivalent6,2 t
Circulation pump:MK pump
max. delivery201 l/min
max. delivery pressure5,3 bar

Engineering Facts

14 facts · every claim linked to a page
Selection

The unit is purpose-built for automotive applications using water-glycol mixtures down to -45°C, so it should be selected only for that fluid class.

Confirms the intended heat transfer fluid and low-temperature limit before specifying the unit.

Source
“The temperature control systems are constructed for operation with Water-ethylene glycol mixture incl. corrosion protection (for example Glysantin®) down to -45°C.”
Installation

The cooling water supply must be kept between 5°C and 20°C, and pipes must be insulated if condensation can occur.

Defines the required facility cooling water conditions and condensation protection at the installation site.

Source
“This temperature control unit is designed to operate with cooling water intake temperature between 5°C and 20°C.”
Installation

Cooling water differential pressure must stay within 1 to 4 bar and static pressure must not exceed 6 bar.

Sets the allowable cooling water circuit pressure window for safe and correct operation.

Source
“min. cooling water differential pressure 1 bar”
Installation

The unit requires a 400 V three-phase 50 Hz supply with 31.5 A max current and 3x32 A fusing.

Needed to size the electrical feed, protection and connection before installation.

Source
“Power supply requirement 400V 3~ 50Hz”
Installation

Three-phase devices below 63 A are delivered with cable but without a plug, so a suitable plug must be provided on site.

Prevents surprises during commissioning regarding the missing mains plug.

Source
“Three-phase devices with current consumption less than 63A --> with cable, without plug”
Installation

Voltage and frequency tolerances are ±5% and ±2% but must not combine in the unfavorable direction.

Guides assessment of site power quality before connecting the unit.

Source
“-5% voltage and +2% frequency -> not allowed!”
Safety

Leakage current exceeds 3.5 mA, which affects residual current device selection and earthing requirements.

Important for electrical safety design and RCD compatibility at installation.

Source
“Attention: leakage current > 3,5mA”
Safety

The refrigerant R-452A is classified A1 with GHS hazard H280, so refrigerant handling and ventilation must be planned.

Informs safety measures and labeling for the refrigerant circuit.

Source
“R-452A (A1, H280)”
Conformité

The unit is EMC Class A Group 1, so it is intended for industrial environments and may cause interference in residential areas.

Determines EMC environment suitability and possible filtering needs.

Source
“Classification (disturbance) to EN55011: Class A, Group 1”
Operation

Ambient temperature must remain between 5°C and 40°C for correct operation.

Defines the allowable room conditions where the unit can be operated.

Source
“min. ambient temperature 5 °C”
Operation

Cooling power drops as temperature falls, from 14 kW at 20°C to 0.8 kW at -40°C, so low-temperature duty must be checked.

Critical for verifying sufficient cooling capacity at the intended process temperature.

Source
“at -40°C 0,8 kW”
Operation

The pump delivers 201 l/min max with 5.3 bar max pressure, and flow falls with rising pressure.

Used to match the external circuit pressure drop and required flow rate.

Source
“max. delivery pressure 5,3 bar”
Maintenance

The cooling water circuit uses copper, stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM, so only compatible water quality should be used.

Guides water treatment and material compatibility to avoid corrosion and fouling.

Source
“Materiels used in the cooling water circuit include; copper, Stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM.”
Installation

Pump connections are M38x1.5 male and cooling water connections are G3/4 male, so matching hoses and fittings are required.

Needed to procure correct adapters and hoses for piping.

Source
“Pump connection M38x1,5 male”