Unimotive GL 07w-XT Technical Data

Sommaire

  1. Overview p. 1
    Product introduction and general description of the unit
  2. Spec pp. 1-2
    Technical data according to DIN 12876 and performance values
  3. p. 2 table, 14 rows
  4. p. 1 table, 26 rows

Technical data according to DIN 12876

ParameterValue
Cooling water connectionG3/4 male
min. cooling water differential pressure1 bar
max. cooling water differential pressure4 bar
max. cooling water pressure6 bar
max. System pressure12 bar
min. filling capacity8,9 l
Filling capacity expansion tank19,9 l
Overall dimensions WxDxH **730x804x1738 mm
Power supply requirement400V 3~ 50Hz
Pressure equipment categoryI
Degree of ProtectionIP20
min. ambient temperature5 °C
max. ambient temperature40 °C
from Serial-No.:1.0/24

** Please respect space requirements. See operating conditions at www.huber-online.com

Technical data according to DIN 12876

ParameterValue
Operating temperature range-45...150 °C
Temperature stability at -10°C0,05 K
temperature set point / display5,7" colour Touchscreen
Resolution of display0,01 K
Internal temperature sensorPt100
Sensor external connectionPt100
Interface digitalEthernet, USB (Host u. Device), RS232
digital inputECS ONE
digital outputPOKO ONE
Interface analogoptional
Alarm messageoptic, acoustic, relay
Safety classificationI / NFL
Heating power12 kW
Cooling power with Water/ Glysantin 40/60 at 20°C12 kW
Cooling power with Water/ Glysantin 40/60 at 0°C10 kW
at -20°C7 kW
at -30°C4,5 kW
at -40°C2,5 kW
Refrigeration machinewater-cooled, natural refrigerant
Refrigerant (ASHRAE, GHS)R-744 (A1, H280)
Global Warming Potential (GWP)1
Circulation pump:MK pump
max. delivery150 l/min
max. delivery pressure5,3 bar
Pump connectionM38x1,5 male
max. permissible kin. viscosity200 mm²/s

Engineering Facts

14 facts · every claim linked to a page
Selection

The unit is designed for water-ethylene glycol mixtures with corrosion protection down to -45°C, so it suits automotive test applications but not plain water at low temperatures.

Determines the acceptable heat transfer fluid and the achievable low-temperature limit when selecting the unit.

Source
“The temperature control systems are constructed for operation with Water-ethylene glycol mixture incl. corrosion protection (for example Glysantin®) down to -45°C.”
Installation

Cooling water must be supplied at 1–4 bar differential pressure and no more than 6 bar static pressure, so the site water supply must be checked before installation.

Incorrect cooling water pressure can prevent rated cooling capacity or damage the water-cooled refrigeration circuit.

Source
“min. cooling water differential pressure 1 bar”
Installation

The unit requires a 400 V three-phase 50 Hz supply, so the electrical infrastructure must be verified before siting.

Three-phase power is mandatory; a single-phase supply cannot run this 12 kW heating/cooling unit.

Source
“Power supply requirement 400V 3~ 50Hz”
Installation

Three-phase devices below 63 A are delivered with cable but without a plug, so the customer must provide and fit the appropriate plug.

Affects electrical installation planning and the scope of supply at commissioning.

Source
“Three-phase devices with current consumption less than 63A --> with cable, without plug”
Installation

The cooling water circuit uses copper, stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM, so the cooling water must be compatible with these materials.

Incompatible cooling water chemistry can cause corrosion and clogging of the water-cooled condenser.

Source
“Materiels used in the cooling water circuit include; copper, Stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM.”
Installation

If cooling water temperature drops below the condensation point, the cooling water pipes inside the unit must be insulated.

Prevents condensation and corrosion inside the unit when using cold cooling water.

Source
“When cooling water temperatures fall below the condensation point, it is essential to insulate the cooling water pipes in the unit.”
Performance

Cooling capacity is rated at 20°C room temperature, 15°C cooling water inlet and 1 bar differential pressure, so actual capacity will differ if site conditions deviate.

Needed to correctly interpret the published cooling capacity curves during selection.

Source
“Output data valid for: Room temperature 20°C, cooling water inlet 15°C and 1 bar differential pressure between cooling water inlet and -outlet.”
Performance

Cooling capacity falls as cooling water temperature rises, so higher site water temperatures reduce available cooling power.

Affects sizing margin and cooling water supply design.

Source
“As the cooling water temperature increases, drop in the cooling power should be expected, and also an increased cooling water flow rate possible.”
Operation

The unit is intended for externally closed applications, so it should not be used with open baths.

Defines the permissible hydraulic circuit configuration for safe operation.

Source
“For externally closed applications.”
Operation

The maximum permissible kinematic viscosity of the circulating fluid is 200 mm²/s, which limits fluid choice at low temperatures.

Prevents pump overload and ensures adequate circulation when selecting heat transfer fluids.

Source
“max. permissible kin. viscosity 200 mm²/s”
Safety

Leakage current exceeds 3.5 mA, so protective earth and RCD selection must account for this.

Affects electrical safety design and choice of residual current protective device.

Source
“Attention: leakage current > 3,5mA”
Safety

The refrigerant is R-744 (CO2) with GHS hazard H280, so ventilation and pressure safety measures for the refrigerant must be considered.

Relevant for site risk assessment and handling of the natural refrigerant.

Source
“R-744 (A1, H280)”
Conformité

Voltage tolerance is ±5% only with simultaneous frequency tolerance of ±2%, and certain combinations are not allowed.

Ensures the supply quality at the installation site meets the unit's operating envelope.

Source
“-5% voltage and +2% frequency -> not allowed!”
Conformité

The equipment is EMC Class A Group 1, so it may cause radio interference in residential environments.

Affects installation location and EMC planning in sensitive environments.

Source
“Classification (disturbance) to EN55011: Class A, Group 1”