Sommaire

  1. Overview p. 1
    Product introduction and general description of the Unistat 405w unit
  2. Spec pp. 1-2
    Technical data and performance specifications according to DIN 12876
  3. p. 2 table, 19 rows
  4. p. 1 table, 28 rows

Technical data according to DIN 12876

ParameterValue
Cooling water connectionG1/2 male
Consumption at water 15°C, flow 0°C72 l/h
Consumption at water 15°C, flow -20°C54 l/h
Consumption at water 25°C, flow 0°C186 l/h
Verbrauch b. Wasser 25°C, Vorlauf -20°C186 l/h
min. cooling water differential pressure0,5 bar
max. cooling water pressure6 bar
min. filling capacity1,5 l
Filling capacity expansion tank2,8 l
Overall dimensions WxDxH **426x327x631 mm
Net weight55 kg
Power supply requirement208V 2~ 60Hz
max. current18,5 A
Fuse20 A
Pressure equipment categoryArt. 4.3 PED
Degree of ProtectionIP20
min. ambient temperature5 °C
max. ambient temperature40 °C
from Serial-No.:S585019

** Please respect space requirements. See operating conditions at www.huber-online.com

Technical data according to DIN 12876

ParameterValue
Operating temperature range-45...250 °C
Temperature stability at -10°C0,01 K
temperature set point / display5,7" colour Touchscreen
Resolution of display0,01 K
Internal temperature sensorPt100
Sensor external connectionPt100
Interface digitalEthernet, USB (Host u. Device), RS232
digital inputECS ONE
digital outputPOKO ONE
Alarm messageoptic, acoustic, relay
Safety classificationIII / FL
Heating power3 kW
Cooling power with Thermooil at 250°C1,3 kW
Cooling power with Thermooil at 200°C1,3 kW
Cooling power with Thermooil at 100°C1,3 kW
Cooling power with Ethanol at 0°C1,3 kW
Cooling power with Ethanol at -20°C0,7 kW
Cooling power with Ethanol at -40°C0,15 kW
Refrigeration machinewater-cooled, natural refrigerant
Refrigerant (ASHRAE, GHS)R-1270 (A3, H220)
Global Warming Potential (GWP)0
Refrigerant quantity0,09 kg
CO2 equivalent0,0 t
UN-numberUN 3358
Circulation pump: max. delivery55 l/min
Circulation pump: max. delivery pressure0,9 bar
Pump connectionM24x1,5 male
max. permissible kin. viscosity50 mm²/s

Order-No.: 1002.0062.01; Unistat 405w

Engineering Facts

14 facts · every claim linked to a page
Installation

The unit requires cooling water supply with a minimum differential pressure of 0.5 bar and must not exceed 6 bar.

Ensures the water-cooled refrigeration circuit receives adequate flow without overpressurizing the condenser.

Source
“min. cooling water differential pressure 0,5 bar”
Installation

Cooling water inlet temperature must be kept between 5°C and 20°C for rated performance.

Defines the site cooling water requirements and warns that higher inlet temperatures reduce cooling capacity.

Source
“This temperature control unit is designed to operate with cooling water intake temperature between 5°C and 20°C.”
Installation

If cooling water temperature can fall below the dew point, the cooling water pipes inside the unit must be insulated.

Prevents condensation and corrosion inside the unit when cold water is supplied.

Source
“When cooling water temperatures fall below the condensation point, it is essential to insulate the cooling water pipes in the unit.”
Installation

The electrical supply must be 208 V two-phase 60 Hz with a 20 A fuse and maximum current of 18.5 A.

Determines the site power infrastructure and breaker sizing needed before installation.

Source
“Power supply requirement 208V 2~ 60Hz”
Installation

Voltage and frequency tolerances are ±5% and ±2% respectively, but the combination of -5% voltage with +2% frequency is not allowed.

Prevents motor and compressor damage from unfavorable supply conditions at the installation site.

Source
“-5% voltage and +2% frequency -> not allowed!”
Installation

The pump connection is M24x1.5 male and requires mating fittings per DIN 3863/3870.

Ensures correct hose or pipe adapters are ordered for the external circuit.

Source
“Pump connection M24x1,5 male”
Operation

The expansion tank is thermally passive, so no heat transfer fluid vapor or moisture absorption occurs, and for external open baths the tank is blocked off.

Allows the thermostat to be placed below or above the application level and avoids fluid degradation.

Source
“No HTF vapour and no moisture absorption because the expansion tank is thermally passive.”
Operation

The maximum permissible kinematic viscosity of the heat transfer fluid is 50 mm²/s.

Guides selection of a suitable heat transfer fluid that the pump can circulate at low temperatures.

Source
“max. permissible kin. viscosity 50 mm²/s”
Performance

Cooling power drops significantly at low temperatures with ethanol, from 1.3 kW at 0°C to 0.15 kW at -40°C.

Sets realistic expectations for heat removal capacity at the low end of the operating range.

Source
“at -40°C 0,15 kW”
Performance

Cooling water consumption increases with warmer supply water, e.g. 72 l/h at 15°C versus 186 l/h at 25°C for 0°C flow.

Affects sizing of the facility cooling water supply and drain capacity.

Source
“Consumption at water 25°C, flow 0°C 186 l/h”
Safety

The refrigerant is R-1270 (A3, H220) with GWP 0 and a charge of only 0.09 kg.

A3 flammability classification requires awareness of ventilation and ignition source control at the installation site.

Source
“Refrigerant (ASHRAE, GHS) R-1270 (A3, H220)”
Safety

The unit is classified as safety class III / FL according to DIN 12876.

Indicates suitability for flammable liquids and defines the safety category for the application.

Source
“Safety classification III / FL”
Conformité

The equipment is classified as EN55011 Class A, Group 1 for electromagnetic disturbance.

Class A equipment may require additional EMC measures in residential or light-industrial environments.

Source
“Classification (disturbance) to EN55011: Class A, Group 1”
Maintenance

The cooling water circuit uses copper, stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM, so suitable water quality is required.

Guides water treatment to avoid corrosion and fouling of these materials.

Source
“Materiels used in the cooling water circuit include; copper, Stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM.”