Installation
The unit requires a 400 V three-phase supply with a 3x16 A fuse, so the electrical installation must be sized accordingly.
Determines the site power supply, cable and breaker rating needed before installation.
Source
“Power supply requirement 400V 3~ 50Hz”
Installation
The cooling water circuit must be supplied at 5–20 °C inlet with at least 1 bar differential pressure and no more than 6 bar.
Cooling water conditions directly affect achievable cooling power and must be verified on site.
Source
“This temperature control unit is designed to operate with cooling water intake temperature between 5°C and 20°C.”
Installation
Cooling water pipes must be insulated if the water temperature can drop below the dew point to prevent condensation.
Prevents condensation damage and electrical/safety issues in the installation.
Source
“When cooling water temperatures fall below the condensation point, it is essential to insulate the cooling water pipes in the unit.”
Installation
The cooling water circuit uses copper, stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM, so only compatible water quality should be used.
Guides selection of water treatment and avoids corrosion or material incompatibility.
Source
“Materiels used in the cooling water circuit include; copper, Stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM.”
Conformité
Voltage and frequency tolerances are ±5% and ±2% but must not be combined in the unfavorable direction.
Defines acceptable supply quality for reliable operation and warranty compliance.
Source
“Voltage + / - 5% with a simultaneous frequency tolerance of + / - 2%”
Safety
The unit has a leakage current above 3.5 mA, which requires appropriate protective earthing and RCD selection.
Affects electrical protection design and residual current device selection.
Source
“Attention: leakage current > 3,5mA”
Installation
Three-phase devices below 63 A are delivered with cable but without a plug, so a suitable plug must be provided on site.
Prevents delays at commissioning because the customer must supply the plug.
Source
“Three-phase devices with current consumption less than 63A --> with cable, without plug”
Operation
The unit is rated for ambient temperatures from 5 °C to 40 °C, which limits where it can be installed.
Room conditions must be checked to avoid derating or malfunction.
Source
“min. ambient temperature 5 °C”
Performance
Cooling power drops significantly at low temperatures, from 5.3 kW at 0 °C with ethanol to 0.25 kW at -50 °C.
Critical for selecting the unit for low-temperature process cooling duties.
Source
“at -50°C 0,25 kW”
Selection
The pump delivers 112 l/min maximum and 54 l/min at 1.0 bar, which must match the external circuit pressure drop.
Ensures the circulation pump can meet the flow and pressure needs of the application.
Source
“max. delivery 112 l/min”
Safety
The refrigerant R-452A is classified A1 with GHS H280, and the charge corresponds to 3.1 t CO2 equivalent.
Relevant for refrigerant handling, safety classification and environmental reporting.
Source
“R-452A (A1, H280)”
Installation
The pump connection is M30x1.5 male and the cooling water connection is G1/2 male, so matching hoses and adapters are needed.
Determines the fittings and hoses required for commissioning.
Source
“Pump connection M30x1,5 male”
Operation
The unit has IP20 protection, so it must be kept in a dry indoor environment free from water ingress.
Restricts installation location and protects against moisture damage.
Source
“Degree of Protection IP20”