Product introduction and general description of the Unistat 525
Specp. 2
Technical data according to DIN 12876
p. 2table, 14 rows
p. 1table, 34 rows
Technical data according to DIN 12876
Parameter
Value
max. permissible kin. viscosity
50 mm²/s
min. filling capacity
5,1 l
Filling capacity expansion tank
18 l
Overall dimensions WxDxH **
1290x795x1377 mm
Net weight
417 kg
sound pressure level +/- 4 dB(A)
56 dB(A)
Power supply factory configured (3 Phase)
400V 3~ 50Hz
max. current (3 Phase)
16 A
Fuse (3 phase)
3x20A
Pressure equipment category
Art. 4.3 PED
Degree of Protection
IP20
min. ambient temperature
5 °C
max. ambient temperature
40 °C
from Serial-No.:
S565613
** Please respect space requirements. See operating conditions at www.huber-online.com
Unistat 525 — Technical data according to DIN 12876
Parameter
Value
Order-No.
1051.0010.01
Operating temperature range
-55...250 °C
Temperature stability at -10°C
0,01 K
temperature set point / display
5,7" colour Touchscreen
Resolution of display
0,01 K
Internal temperature sensor
Pt100
Sensor external connection
Pt100
Interface digital
Ethernet, USB (Host u. Device), RS232
digital input
ECS ONE
digital output
POKO ONE
Alarm message
optic, acoustic, relay
Safety classification
III / FL
Heating power
6 kW
Cooling power with Thermooil at 250°C
10 kW
Cooling power with Thermooil at 200°C
10 kW
Cooling power with Thermooil at 100°C
10 kW
Cooling power with Thermooil at 20°C
10 kW
Cooling power with Ethanol at 0°C
7 kW
Cooling power with Ethanol at -20°C
4,2 kW
Cooling power with Ethanol at -40°C
1,5 kW
Cooling power with Ethanol at -50°C
0,65 kW
Refrigeration machine
air-cooled, CFC- and HCFC-free
Refrigerant (ASHRAE, GHS)
R-452A (A1, H280)
Global Warming Potential (GWP)
2141
Refrigerant quantity
3,9 kg
CO2-Äquvivalent
8,3 t
Circulation pump:
MK pump
max. delivery
79 l/min
max. delivery pressure
1,5 bar
Delivery at 0,5 bar
63 l/min
Delivery at 0,75 bar
54 l/min
Delivery at 1,0 bar
43 l/min
Delivery at 1,25 bar
29 l/min
Pump connection
M30x1,5 male
Значения приведены дословно по тексту страницы; таблица транспонирована в формат «характеристика — значение».
Engineering Facts
14 facts · every claim linked to a page
Installation
The unit requires a 400 V three-phase supply with a 16 A maximum current and 3x20 A fusing, so the site electrical infrastructure must be sized accordingly.
Determines the required three-phase feed, breaker rating and cabling before installation.
Three-phase devices drawing less than 63 A are delivered with a cable but without a plug, so the customer must provide and fit the site plug.
Affects electrical installation scope and procurement of the mating plug.
Source
“Three-phase devices with current consumption less than 63A --> with cable, without plug”
Safety
Leakage current exceeds 3.5 mA, which requires appropriate protective earthing and RCD selection on site.
Impacts earth connection and residual-current device choice for safe operation.
Source
“Attention: leakage current > 3,5mA”
Installation
The unit is rated IP20, so it must be installed in a dry indoor location protected from water ingress.
Restricts installation environment and enclosure requirements.
Source
“Degree of Protection IP20”
Installation
Ambient temperature must stay between 5 °C and 40 °C, and cooling capacity may drop above 20 °C room temperature.
Sets the allowable plant-room temperature and warns that performance derates with rising ambient.
Source
“Output data valid for: Room temperature 20°C. If the ambient temperature rises, the cooling capacity may drop.”
Installation
Voltage and frequency tolerances are ±5% and ±2% respectively, but the combination of -5% voltage with +2% frequency is not allowed.
Defines the acceptable supply quality window for reliable operation.
Source
“-5% voltage and +2% frequency -> not allowed!”
Installation
The pump connection is M30x1.5 male with Y-bore and ball-socket fittings per DIN 3863/3870, so matching hoses and fittings must be specified.
Ensures correct mating of process hoses and avoids leaks at the circulation connections.
Source
“Pump connection M30x1,5 male”
Operation
The pump delivers 79 l/min maximum and 1.5 bar maximum pressure, with flow falling to 29 l/min at 1.25 bar, so external circuit pressure drop must be checked.
Allows verification that the external loop resistance is compatible with the pump curve.
Source
“max. delivery 79 l/min”
Operation
Minimum filling capacity is 5.1 l and the expansion tank holds 18 l, which sets the minimum heat-transfer fluid charge and expansion allowance.
Determines fluid volume needed and expansion volume for safe thermal cycling.
Source
“min. filling capacity 5,1 l”
Operation
Maximum permissible kinematic viscosity is 50 mm²/s, so the heat-transfer fluid must be selected to stay below this limit at the lowest operating temperature.
Prevents pump overload and poor circulation when choosing fluid for low-temperature operation.
Source
“max. permissible kin. viscosity 50 mm²/s”
Safety
The refrigerant is R-452A (A1, H280) with GWP 2141 and 3.9 kg charge, giving 8.3 t CO2 equivalent, which affects refrigerant handling and environmental reporting.
Needed for refrigerant safety classification, leak response and CO2-equivalent reporting.
Source
“R-452A (A1, H280)”
Conformité
The unit is EMC Class A Group 1, so it is intended for industrial environments and may cause interference in residential areas.
Informs site EMC assessment and suitability for non-residential installation.
Source
“Classification (disturbance) to EN55011: Class A, Group 1”
Selection
Cooling capacity with ethanol falls from 7 kW at 0 °C to 0.65 kW at -50 °C, so low-temperature duty must be checked against the actual fluid and temperature.
Prevents oversizing assumptions at low temperature where available cooling is much lower.
Source
“at -50°C 0,65 kW”
Maintenance
The unit has an optical level indicator and an optional float switch in the sight glass for extended security, supporting fluid-level monitoring.
Guides selection of level monitoring to protect the pump and heater from low fluid.
Source
“float switch in sight glass for extended security”