Unistat 540w Technical Data

Sommaire

  1. Overview p. 1
    Introduces the Unistat 540w unit and its main features
  2. Spec pp. 1-2
    Performance, temperature, and dimensional data per DIN 12876
  3. p. 2 table, 20 rows
  4. p. 1 table, 33 rows

Technical data according to DIN 12876

ParameterValue
Pump connectionM38x1,5 male
max. permissible kin. viscosity100 mm²/s
Cooling water connectionG3/4 male
Consumption at water 15°C, flow 0°C1480 l/h
min. cooling water differential pressure1 bar
max. cooling water differential pressure4 bar
max. cooling water pressure6 bar
max. System pressure6 bar
min. filling capacity9,9 l
Filling capacity expansion tank19,9 l
Overall dimensions WxDxH **730x804x1738 mm
Net weight472 kg
Power supply requirement400V 3~ 50Hz
max. current55 A
Fuse3x63A
Pressure equipment categoryI
Degree of ProtectionIP20
min. ambient temperature5 °C
max. ambient temperature40 °C
from Serial-No.:S548344

** Please respect space requirements. See operating conditions at www.huber-online.com

Technical data according to DIN 12876

CharacteristicValue
Operating temperature range-55...250 °C
Temperature stability at -10°C0,01 K
temperature set point / display5,7" colour Touchscreen
Resolution of display0,01 K
Internal temperature sensorPt100
Sensor external connectionPt100
Interface digitalEthernet, USB (Host u. Device), RS232
digital inputECS ONE
digital outputPOKO ONE
Interface analogoptional
Alarm messageoptic, acoustic, relay
Safety classificationIII / FL
Heating power24 kW
Cooling power with Thermooil at 250°C30 kW
Cooling power with Thermooil at 200°C30 kW
Cooling power with Thermooil at 100°C30 kW
Cooling power with Thermooil at 20°C30 kW
Cooling power with Ethanol at 0°C30 kW
Cooling power with Ethanol at -20°C16 kW
Cooling power with Ethanol at -40°C4 kW
Cooling power with Ethanol at -50°C1,5 kW
Refrigeration machinewater-cooled, CFC- and HCFC-free
Refrigerant (ASHRAE, GHS)R-452A (A1, H280)
Global Warming Potential (GWP)2141
Refrigerant quantity3,6 kg
CO2 equivalent7,7 t
Circulation pump:MK pump
max. delivery196 l/min
max. delivery pressure2,5 bar
Delivery at 0,5 bar175 l/min
Delivery at 1,0 bar152 l/min
Delivery at 1,5 bar123 l/min
Delivery at 2,0 bar86 l/min

Engineering Facts

13 facts · every claim linked to a page
Installation

The unit requires a 400 V three-phase supply with a 3x63 A fuse and draws up to 55 A, so the electrical installation must be sized accordingly.

Determines the site power supply, cable cross-section and protection device needed before installation.

Source
“Power supply requirement 400V 3~ 50Hz max. current 55 A Fuse 3x63A”
Installation

The cooling water circuit must be supplied at 1–4 bar differential pressure and no more than 6 bar, with 1480 l/h consumption at 15 °C water and 0 °C flow.

Sizing of the cooling water supply, pump and piping depends on these pressure and flow limits.

Source
“Consumption at water 15°C, flow 0°C 1480 l/h min. cooling water differential pressure 1 bar max. cooling water differential pressure 4 bar max. cooling water pressure 6 bar”
Operation

The unit is designed for cooling water intake between 5 °C and 20 °C, and cooling power drops as the water temperature rises.

The available cooling capacity and reliable operation depend on keeping the cooling water within this range.

Source
“This temperature control unit is designed to operate with cooling water intake temperature between 5°C and 20°C.”
Installation

If cooling water temperature falls below the condensation point, the cooling water pipes in the unit must be insulated.

Prevents condensation and corrosion/electrical risk when cold cooling water is used.

Source
“When cooling water temperatures fall below the condensation point, it is essential to insulate the cooling water pipes in the unit.”
Installation

The cooling water circuit contains copper, stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM, so only compatible cooling water may be used.

Water quality must be matched to these materials to avoid corrosion and fouling.

Source
“Materiels used in the cooling water circuit include; copper, Stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM. Please use suitable cooling water.”
Conformité

Voltage tolerance is ±5% only with a simultaneous frequency tolerance of ±2%, and certain combinations such as -5% voltage with +2% frequency are not allowed.

The site power quality must be checked against these combined tolerances before commissioning.

Source
“Voltage + / - 5% with a simultaneous frequency tolerance of + / - 2%”
Safety

The equipment has a leakage current greater than 3.5 mA, which affects the choice of residual current protection.

RCD selection and earthing design must account for the high leakage current.

Source
“Attention: leakage current > 3,5mA”
Installation

Three-phase devices drawing less than 63 A are delivered with a cable but without a plug, so a suitable plug must be provided on site.

The electrical installer must supply and fit the appropriate plug for commissioning.

Source
“Three-phase devices with current consumption less than 63A --> with cable, without plug”
Installation

The pump connection is M38x1.5 male with bore shape Y (60°) per DIN 3863, and pipework must use ball sockets per DIN 3863 and sleeve nuts per DIN 3870.

Ensures the correct mating fittings and hoses are procured for leak-free connection.

Source
“Pump connections: Bore shape Y (60°) according to DIN 3863, pipework/flexible tempering hoses: Ball socket according to DIN 3863, sleeve nut according to DIN 3870.”
Selection

The unit uses refrigerant R-452A with GWP 2141 and a charge of 3.6 kg, giving a CO2 equivalent of 7.7 t.

Relevant for refrigerant regulations, leak monitoring and environmental reporting.

Source
“Refrigerant (ASHRAE, GHS) R-452A (A1, H280) Global Warming Potential (GWP) 2141 Refrigerant quantity 3,6 kg CO2 equivalent 7,7 t”
Operation

The maximum permissible kinematic viscosity of the heat transfer fluid is 100 mm²/s.

Limits the choice of heat transfer fluid and ensures adequate pump performance.

Source
“max. permissible kin. viscosity 100 mm²/s”
Installation

The unit has protection degree IP20 and must be operated between 5 °C and 40 °C ambient temperature.

Defines the permissible installation environment and ventilation requirements.

Source
“Degree of Protection IP20 min. ambient temperature 5 °C max. ambient temperature 40 °C”
Selection

The unit is water-cooled and uses a CFC- and HCFC-free refrigeration machine.

Confirms the cooling medium and refrigerant compliance for selection.

Source
“Refrigeration machine water-cooled, CFC- and HCFC-free”