Product name and general introduction to the Unistat 705w unit
Specp. 2
Technical data per DIN 12876 including performance and dimensions
p. 2table, 25 rows
p. 1table, 28 rows
Technical data according to DIN 12876
Parameter
Value
CO2 equivalent 2nd stage
0,0 t
UN-number
UN 3358
Circulation pump:
max. delivery
55 l/min
max. delivery pressure
0,9 bar
Delivery at 0,1 bar
53 l/min
Delivery at 0,5 bar
40 l/min
Delivery at 0,8 bar
25 l/min
Pump connection
M24x1,5 male
max. permissible kin. viscosity
50 mm²/s
Cooling water connection
G1/2 male
min. cooling water differential pressure
0,5 bar
max. cooling water pressure
6 bar
min. filling capacity
1,5 l
Filling capacity expansion tank
2,8 l
Overall dimensions WxDxH **
426x399x719 mm
Net weight
90 kg
Power supply requirement
230V 1~ 50Hz
max. current
11 A
Fuse
16 A
Pressure equipment category
I
Degree of Protection
IP20
min. ambient temperature
5 °C
max. ambient temperature
40 °C
from Serial-No.:
S600758
** Please respect space requirements. See operating conditions at www.huber-online.com
Technical data according to DIN 12876
Parameter
Value
Operating temperature range
-75...250 °C
Temperature stability at -10°C
0,01 K
temperature set point / display
5,7" colour Touchscreen
Resolution of display
0,01 K
Internal temperature sensor
Pt100
Sensor external connection
Pt100
Interface digital
Ethernet, USB (Host u. Device), RS232
digital input
ECS ONE
digital output
POKO ONE
Alarm message
optic, acoustic, relay
Safety classification
III / FL
Heating power
1,5 kW
Cooling power with Thermooil at 250°C
0,6 kW
Cooling power with Thermooil at 200°C
0,6 kW
Cooling power with Thermooil at 100°C
0,6 kW
Cooling power with Ethanol at 0°C
0,65 kW
Cooling power with Ethanol at -20°C
0,6 kW
Cooling power with Ethanol at -40°C
0,6 kW
Cooling power with Ethanol at -60°C
0,3 kW
Cooling power with Ethanol at -75°C
0,06 kW
Refrigeration machine
water-cooled, natural refrigerant
Refrigerant (ASHRAE, GHS)
R-1270 (A3, H220)
Global Warming Potential (GWP)
0
Refrigerant quantity
0,15 kg
CO2 equivalent
0,0 t
Refrigerant 2nd stage (ASHRAE, GHS)
R-170 (A3, H220)
Global Warming Potential (GWP)
0,437
Refrigerant quantity 2nd stage
0,099 kg
Engineering Facts
14 facts · every claim linked to a page
Installation
The unit requires a water-cooled refrigeration machine, so a cooling water supply with adequate pressure and flow must be available at the installation site.
Determines site requirements for cooling water supply before purchase and installation.
Source
“Hydraulically sealed Refrigerated Heating Circulator with water-cooled refrigerating unit.”
Installation
The cooling water connection is G1/2 male and requires a minimum differential pressure of 0.5 bar and maximum pressure of 6 bar.
Needed to specify piping, fittings and supply pressure for the cooling water circuit.
Source
“Cooling water connection G1/2 male min. cooling water differential pressure 0,5 bar max. cooling water pressure 6 bar”
Operation
The unit is designed for cooling water intake between 5°C and 20°C, and cooling power drops as the cooling water temperature rises.
Defines the allowable cooling water temperature range and expected performance derating.
Source
“This temperature control unit is designed to operate with cooling water intake temperature between 5°C and 20°C.”
Installation
If cooling water temperature falls below the condensation point, the cooling water pipes in the unit must be insulated.
Prevents condensation and corrosion damage in the cooling water circuit.
Source
“When cooling water temperatures fall below the condensation point, it is essential to insulate the cooling water pipes in the unit.”
Installation
The cooling water circuit uses copper, stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM, so only suitable cooling water may be used.
Guides water quality and material compatibility to avoid corrosion or fouling.
Source
“Materiels used in the cooling water circuit include; copper, Stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM.”
Installation
The electrical supply must be 230V 1~ 50Hz with a maximum current of 11 A and a 16 A fuse.
Determines the electrical supply, breaker and wiring requirements for installation.
Source
“Power supply requirement 230V 1~ 50Hz max. current 11 A Fuse 16 A”
Installation
Voltage and frequency tolerances are ±5% and ±2% respectively, but a -5% voltage combined with +2% frequency is not allowed.
Ensures the supply quality is within the allowed envelope to avoid damage or malfunction.
Source
“Voltage + / - 5% with a simultaneous frequency tolerance of + / - 2%”
Installation
The pump connection is M24x1.5 male, and pipework or flexible hoses must use ball socket per DIN 3863 and sleeve nut per DIN 3870.
Needed to select compatible hoses and fittings for the external circuit.
Source
“Pump connection M24x1,5 male”
Operation
The maximum permissible kinematic viscosity of the heat transfer fluid is 50 mm²/s.
Limits the choice of heat transfer fluid to ensure proper pump operation.
Source
“max. permissible kin. viscosity 50 mm²/s”
Installation
The minimum filling capacity is 1.5 l and the expansion tank capacity is 2.8 l.
Determines the minimum fluid volume and expansion volume for safe operation.
Source
“min. filling capacity 1,5 l Filling capacity expansion tank 2,8 l”
Installation
The unit has degree of protection IP20 and must be operated between 5°C and 40°C ambient temperature.
Defines the environmental conditions and protection level for installation location.
Source
“Degree of Protection IP20 min. ambient temperature 5 °C max. ambient temperature 40 °C”
Safety
The refrigerant is R-1270 (A3, H220) with GWP 0 and quantity 0.15 kg, and the second stage is R-170 (A3, H220) with quantity 0.099 kg.
Flammable refrigerants require appropriate safety measures and ventilation.
Source
“Refrigerant (ASHRAE, GHS) R-1270 (A3, H220)”
Conformité
The equipment is classified as Class A, Group 1 for electromagnetic disturbance according to EN55011.
Class A equipment may require additional EMC measures in residential environments.
Source
“Classification (disturbance) to EN55011: Class A, Group 1”
Installation
For single/two-phase devices, a power cable with country-specific plug is supplied; for three-phase devices below 63A, cable without plug; above 63A, no cable or plug.
Clarifies what electrical accessories are included and what must be specified at order.
Source
“Single / two-phase devices (100V to 240V) --> with power cable and country-specific plug (please specify when ordering)”