Introduces the Unistat P510 temperature control system
Specp. 2
Technical data according to DIN 12876 standard
p. 2table, 18 rows
p. 1table, 30 rows
Technical data according to DIN 12876
Parameter
Value
Delivery at 1,5 bar
74 l/min
Delivery at 2,0 bar
55 l/min
Delivery at 2,5 bar
33 l/min
Pump connection
M30x1,5 male
max. permissible kin. viscosity
50 mm²/s
min. filling capacity
4,1 l
Filling capacity expansion tank
5,5 l
Overall dimensions WxDxH **
560x754x1457 mm
Net weight
232 kg
sound pressure level +/- 4 dB(A)
60 dB(A)
Power supply factory configured (3 Phase)
400V 3~ 50Hz
max. current (3 Phase)
14 A
Fuse (3 phase)
3x16A
Pressure equipment category
Art. 4.3 PED
Degree of Protection
IP20
min. ambient temperature
5 °C
max. ambient temperature
40 °C
from Serial-No.:
S426134
** Please respect space requirements. See operating conditions at www.huber-online.com
Unistat P510 Technical data according to DIN 12876
Characteristic
Value
Order-No.
1070.0010.01
Operating temperature range
-50...250 °C
Temperature stability at -10°C
0,01 K
temperature set point / display
5,7" colour Touchscreen
Resolution of display
0,01 K
Internal temperature sensor
Pt100
Sensor external connection
Pt100
Interface digital
Ethernet, USB (Host u. Device), RS232
digital input
ECS ONE
digital output
POKO ONE
Alarm message
optic, acoustic, relay
Safety classification
III / FL
Heating power
6 kW
Cooling power with Thermooil at 250°C
5,3 kW
Cooling power with Thermooil at 200°C
5,3 kW
Cooling power with Thermooil at 100°C
5,3 kW
Cooling power with Thermooil at 20°C
5,3 kW
Cooling power with Ethanol at 0°C
5,3 kW
Cooling power with Ethanol at -20°C
2,8 kW
Cooling power with Ethanol at -40°C
0,9 kW
Cooling power with Ethanol at -50°C
0,25 kW
Refrigeration machine
air-cooled, CFC- and HCFC-free
Refrigerant (ASHRAE, GHS)
R-452A (A1, H280)
Global Warming Potential (GWP)
2141
Refrigerant quantity
2,8 kg
CO2-Äquvivalent
6,0 t
Circulation pump: MK pump max. delivery
119 l/min
max. delivery pressure
3 bar
Delivery at 0,5 bar
106 l/min
Delivery at 1,0 bar
91 l/min
Engineering Facts
14 facts · every claim linked to a page
Selection
The unit is a refrigerated heating circulator with a -50 to 250 °C range, so it can cover both deep cooling and high-temperature process tasks from a single device.
Defines whether the circulator can meet the process temperature window without a separate chiller/heater.
Source
“Operating temperature range -50...250 °C”
Selection
Cooling capacity collapses at low temperature (0.25 kW at -50 °C versus 5.3 kW at 20 °C), so low-temperature duty must be sized against the actual set point, not the nominal 5.3 kW.
Prevents oversizing assumptions and ensures enough cooling at the coldest operating point.
Source
“at -50°C 0,25 kW”
Installation
The unit requires a 400 V three-phase 50 Hz supply with 14 A max current and 3x16 A fusing, so the electrical installation must be planned for three-phase power.
Determines site power availability, cable sizing and breaker selection before delivery.
Three-phase devices below 63 A are supplied with cable but without a plug, so the customer must provide and fit the site plug.
Avoids a commissioning delay caused by a missing mains plug.
Source
“Three-phase devices with current consumption less than 63A --> with cable, without plug”
Installation
The pump connection is M30x1.5 male with DIN 3863 ball-socket and DIN 3870 sleeve-nut mating, so hoses and fittings must match this standard.
Ensures correct mating of process hoses and avoids leaks or adapters.
Source
“Pump connection M30x1,5 male”
Installation
The unit needs 560x754x1457 mm of space and the manufacturer explicitly requires respecting space requirements, so the installation bay must be planned accordingly.
Prevents installation in an undersized bay and ensures access for operation and service.
Source
“Please respect space requirements.”
Operation
Cooling capacity is rated at 20 °C room temperature and drops if ambient rises, so the machine room must be kept cool to preserve performance.
Explains why performance may fall short in a hot plant room and supports HVAC planning.
Source
“If the ambient temperature rises, the cooling capacity may drop.”
Operation
The permitted ambient range is 5 to 40 °C, so operation outside this window is not allowed.
Sets the environmental envelope for reliable operation and warranty compliance.
Source
“max. ambient temperature 40 °C”
Safety
Leakage current exceeds 3.5 mA, so protective earthing and RCD selection must account for high leakage current.
Affects earth connection and residual-current device choice to avoid nuisance tripping.
Source
“Attention: leakage current > 3,5mA”
Safety
The refrigerant is R-452A with GHS hazard H280 and GWP 2141, so refrigerant handling and leak response must follow pressurized-gas rules.
Informs site risk assessment, ventilation and refrigerant leak procedures.
Source
“R-452A (A1, H280)”
Conformité
Voltage and frequency tolerances are ±5% and ±2% but not simultaneously in the unfavorable combination, so supply quality must be checked against this rule.
Prevents operation on a supply that is out of the permitted combined tolerance.
Source
“-5% voltage and +2% frequency -> not allowed!”
Conformité
The equipment is EMC Class A Group 1, so it is intended for industrial environments and may cause interference in residential areas.
Guides EMC planning and restricts use in sensitive or residential locations.
Source
“Classification (disturbance) to EN55011: Class A, Group 1”
Maintenance
The refrigeration machine is air-cooled, so adequate air flow and heat rejection space around the unit are required.
Determines ventilation and clearance requirements in the installation room.
Source
“air-cooled, CFC- and HCFC-free”
Performance
The circulation pump delivers 119 l/min at zero head but only 33 l/min at 2.5 bar, so external pressure drop must be checked against the pump curve.
Ensures the pump can overcome the process loop resistance at the required flow.