Unistat P510 Temperature Control System Datasheet

Sommaire

  1. Overview p. 1
    Introduces the Unistat P510 temperature control system
  2. Spec p. 2
    Technical data according to DIN 12876 standard
  3. p. 2 table, 18 rows
  4. p. 1 table, 30 rows

Technical data according to DIN 12876

ParameterValue
Delivery at 1,5 bar74 l/min
Delivery at 2,0 bar55 l/min
Delivery at 2,5 bar33 l/min
Pump connectionM30x1,5 male
max. permissible kin. viscosity50 mm²/s
min. filling capacity4,1 l
Filling capacity expansion tank5,5 l
Overall dimensions WxDxH **560x754x1457 mm
Net weight232 kg
sound pressure level +/- 4 dB(A)60 dB(A)
Power supply factory configured (3 Phase)400V 3~ 50Hz
max. current (3 Phase)14 A
Fuse (3 phase)3x16A
Pressure equipment categoryArt. 4.3 PED
Degree of ProtectionIP20
min. ambient temperature5 °C
max. ambient temperature40 °C
from Serial-No.:S426134

** Please respect space requirements. See operating conditions at www.huber-online.com

Unistat P510 Technical data according to DIN 12876

CharacteristicValue
Order-No.1070.0010.01
Operating temperature range-50...250 °C
Temperature stability at -10°C0,01 K
temperature set point / display5,7" colour Touchscreen
Resolution of display0,01 K
Internal temperature sensorPt100
Sensor external connectionPt100
Interface digitalEthernet, USB (Host u. Device), RS232
digital inputECS ONE
digital outputPOKO ONE
Alarm messageoptic, acoustic, relay
Safety classificationIII / FL
Heating power6 kW
Cooling power with Thermooil at 250°C5,3 kW
Cooling power with Thermooil at 200°C5,3 kW
Cooling power with Thermooil at 100°C5,3 kW
Cooling power with Thermooil at 20°C5,3 kW
Cooling power with Ethanol at 0°C5,3 kW
Cooling power with Ethanol at -20°C2,8 kW
Cooling power with Ethanol at -40°C0,9 kW
Cooling power with Ethanol at -50°C0,25 kW
Refrigeration machineair-cooled, CFC- and HCFC-free
Refrigerant (ASHRAE, GHS)R-452A (A1, H280)
Global Warming Potential (GWP)2141
Refrigerant quantity2,8 kg
CO2-Äquvivalent6,0 t
Circulation pump: MK pump max. delivery119 l/min
max. delivery pressure3 bar
Delivery at 0,5 bar106 l/min
Delivery at 1,0 bar91 l/min

Engineering Facts

14 facts · every claim linked to a page
Selection

The unit is a refrigerated heating circulator with a -50 to 250 °C range, so it can cover both deep cooling and high-temperature process tasks from a single device.

Defines whether the circulator can meet the process temperature window without a separate chiller/heater.

Source
“Operating temperature range -50...250 °C”
Selection

Cooling capacity collapses at low temperature (0.25 kW at -50 °C versus 5.3 kW at 20 °C), so low-temperature duty must be sized against the actual set point, not the nominal 5.3 kW.

Prevents oversizing assumptions and ensures enough cooling at the coldest operating point.

Source
“at -50°C 0,25 kW”
Installation

The unit requires a 400 V three-phase 50 Hz supply with 14 A max current and 3x16 A fusing, so the electrical installation must be planned for three-phase power.

Determines site power availability, cable sizing and breaker selection before delivery.

Source
“Power supply factory configured (3 Phase) 400V 3~ 50Hz”
Installation

Three-phase devices below 63 A are supplied with cable but without a plug, so the customer must provide and fit the site plug.

Avoids a commissioning delay caused by a missing mains plug.

Source
“Three-phase devices with current consumption less than 63A --> with cable, without plug”
Installation

The pump connection is M30x1.5 male with DIN 3863 ball-socket and DIN 3870 sleeve-nut mating, so hoses and fittings must match this standard.

Ensures correct mating of process hoses and avoids leaks or adapters.

Source
“Pump connection M30x1,5 male”
Installation

The unit needs 560x754x1457 mm of space and the manufacturer explicitly requires respecting space requirements, so the installation bay must be planned accordingly.

Prevents installation in an undersized bay and ensures access for operation and service.

Source
“Please respect space requirements.”
Operation

Cooling capacity is rated at 20 °C room temperature and drops if ambient rises, so the machine room must be kept cool to preserve performance.

Explains why performance may fall short in a hot plant room and supports HVAC planning.

Source
“If the ambient temperature rises, the cooling capacity may drop.”
Operation

The permitted ambient range is 5 to 40 °C, so operation outside this window is not allowed.

Sets the environmental envelope for reliable operation and warranty compliance.

Source
“max. ambient temperature 40 °C”
Safety

Leakage current exceeds 3.5 mA, so protective earthing and RCD selection must account for high leakage current.

Affects earth connection and residual-current device choice to avoid nuisance tripping.

Source
“Attention: leakage current > 3,5mA”
Safety

The refrigerant is R-452A with GHS hazard H280 and GWP 2141, so refrigerant handling and leak response must follow pressurized-gas rules.

Informs site risk assessment, ventilation and refrigerant leak procedures.

Source
“R-452A (A1, H280)”
Conformité

Voltage and frequency tolerances are ±5% and ±2% but not simultaneously in the unfavorable combination, so supply quality must be checked against this rule.

Prevents operation on a supply that is out of the permitted combined tolerance.

Source
“-5% voltage and +2% frequency -> not allowed!”
Conformité

The equipment is EMC Class A Group 1, so it is intended for industrial environments and may cause interference in residential areas.

Guides EMC planning and restricts use in sensitive or residential locations.

Source
“Classification (disturbance) to EN55011: Class A, Group 1”
Maintenance

The refrigeration machine is air-cooled, so adequate air flow and heat rejection space around the unit are required.

Determines ventilation and clearance requirements in the installation room.

Source
“air-cooled, CFC- and HCFC-free”
Performance

The circulation pump delivers 119 l/min at zero head but only 33 l/min at 2.5 bar, so external pressure drop must be checked against the pump curve.

Ensures the pump can overcome the process loop resistance at the required flow.

Source
“Delivery at 2,5 bar 33 l/min”