Unistat P510w Temperature Control System Datasheet
Sommaire
Overviewp. 1
Introduces the Unistat P510w unit and its main features
Specpp. 1-2
Performance, temperature range, and technical data per DIN 12876
p. 2table, 22 rows
p. 1table, 29 rows
Technical data according to DIN 12876
Parameter
Value
Delivery at 1,5 bar
74 l/min
Delivery at 2,0 bar
55 l/min
Delivery at 2,5 bar
33 l/min
Pump connection
M30x1,5 male
max. permissible kin. viscosity
50 mm²/s
Cooling water connection
G1/2 male
Consumption at water 15°C, flow 0°C
355 l/h
min. cooling water differential pressure
1 bar
max. cooling water pressure
6 bar
min. filling capacity
4,1 l
Filling capacity expansion tank
3,6 l
Overall dimensions WxDxH **
460x554x1453 mm
Net weight
182 kg
sound pressure level +/- 4 dB(A)
60 dB(A)
Power supply requirement
400V 3~ 50Hz
max. current
13 A
Fuse
3x16A
Pressure equipment category
I
Degree of Protection
IP20
min. ambient temperature
5 °C
max. ambient temperature
40 °C
from Serial-No.:
S450083
** Please respect space requirements. See operating conditions at www.huber-online.com
Technical data according to DIN 12876 — Unistat P510w (Order-No.: 1070.0013.01)
Parameter
Value
Operating temperature range
-50...250 °C
Temperature stability at -10°C
0,01 K
temperature set point / display
5,7" colour Touchscreen
Resolution of display
0,01 K
Internal temperature sensor
Pt100
Sensor external connection
Pt100
Interface digital
Ethernet, USB (Host u. Device), RS232
digital input
ECS ONE
digital output
POKO ONE
Alarm message
optic, acoustic, relay
Safety classification
III / FL
Heating power
6 kW
Cooling power with Thermooil at 250°C
5,3 kW
Cooling power with Thermooil at 200°C
5,3 kW
Cooling power with Thermooil at 100°C
5,3 kW
Cooling power with Thermooil at 20°C
5,3 kW
Cooling power with Ethanol at 0°C
5,3 kW
Cooling power with Ethanol at -20°C
2,8 kW
Cooling power with Ethanol at -40°C
0,9 kW
Cooling power with Ethanol at -50°C
0,25 kW
Refrigeration machine
water-cooled, CFC- and HCFC-free
Refrigerant (ASHRAE, GHS)
R-452A (A1, H280)
Global Warming Potential (GWP)
2141
Refrigerant quantity
1,45 kg
CO2 equivalent
3,1 t
Circulation pump: MK pump max. delivery
119 l/min
Circulation pump: MK pump max. delivery pressure
3 bar
Circulation pump: MK pump Delivery at 0,5 bar
106 l/min
Circulation pump: MK pump Delivery at 1,0 bar
91 l/min
Таблица приведена как напечатана (характеристика — значение); строки Cooling power сгруппированы по теплоносителю (Thermooil / Ethanol) с указанием температур.
Engineering Facts
12 facts · every claim linked to a page
Installation
The unit requires a 400 V three-phase supply with a 3x16 A fuse, so the electrical installation must be sized accordingly.
Determines the site power supply, cable and breaker rating needed before installation.
Source
“Power supply requirement 400V 3~ 50Hz”
Installation
The cooling water circuit must be supplied at 1 bar minimum differential pressure and must not exceed 6 bar.
Ensures the water-cooled condenser receives adequate flow without overpressurizing the circuit.
Source
“min. cooling water differential pressure 1 bar”
Installation
Cooling water intake must stay between 5 °C and 20 °C, and pipes must be insulated if water can drop below the dew point.
Prevents condensation, corrosion and loss of cooling performance at the installation site.
Source
“This temperature control unit is designed to operate with cooling water intake temperature between 5°C and 20°C.”
Performance
Cooling capacity falls sharply at low temperatures, from 5.3 kW at 0 °C to only 0.25 kW at -50 °C with ethanol.
Critical for sizing the unit when the process requires deep cooling near the lower limit.
Source
“at -50°C 0,25 kW”
Selection
The pump delivers 119 l/min at zero head but only 33 l/min at 2.5 bar, so high-pressure-drop loops need careful pump curve review.
Matches pump capability to the pressure drop of the external application circuit.
Source
“max. delivery 119 l/min”
Safety
Leakage current exceeds 3.5 mA, which affects residual-current device selection and earthing requirements.
Needed to choose a suitable RCD type and ensure protective earthing at the installation.
Source
“Attention: leakage current > 3,5mA”
Conformité
Voltage and frequency tolerances are ±5 % and ±2 % but must not combine in the same direction, e.g. -5 % voltage with +2 % frequency is not allowed.
Defines acceptable supply quality and protects the compressor and electronics from out-of-tolerance operation.
Source
“-5% voltage and +2% frequency -> not allowed!”
Installation
Three-phase devices below 63 A are delivered with cable but without a plug, so a plug must be sourced and fitted on site.
Affects site electrical preparation and commissioning schedule.
Source
“Three-phase devices with current consumption less than 63A --> with cable, without plug”
Operation
The refrigerant R-452A has a GWP of 2141 and a CO2 equivalent of 3.1 t, which matters for leak monitoring and F-gas obligations.
Relevant for environmental compliance and refrigerant leak-check intervals.
Source
“Global Warming Potential (GWP) 2141”
Operation
The unit is rated IP20 and must be operated between 5 °C and 40 °C ambient, so it cannot be placed in a wet or hot environment.
Constrains the installation location and enclosure requirements.
Source
“Degree of Protection IP20”
Selection
The maximum permissible kinematic viscosity of the heat transfer fluid is 50 mm²/s, limiting fluid choice at low temperatures.
Prevents pump overload and ensures adequate circulation with the selected thermal fluid.
Source
“max. permissible kin. viscosity 50 mm²/s”
Maintenance
The cooling water circuit contains copper, stainless steel 1.4401, brass, PA, PPE, PTFE and EPDM, so only compatible water quality may be used.
Guides water treatment and avoids corrosion or scaling damage to the condenser.
Source
“Materiels used in the cooling water circuit include; copper, Stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM.”