Unistat P520w Technical Data

Sommaire

  1. Overview p. 1
    Product introduction and general description of the Unistat P520w unit
  2. Spec pp. 1-2
    Technical performance data and operating parameters per DIN 12876
  3. p. 2 table, 22 rows
  4. p. 1 table, 30 rows

Technical data according to DIN 12876

ParameterValue
Delivery at 1,5 bar57 l/min
Delivery at 2,0 bar46 l/min
Delivery at 2,5 bar32 l/min
Pump connectionM30x1,5 male
max. permissible kin. viscosity50 mm²/s
Cooling water connectionG1/2 male
Consumption at water 15°C, flow 20°C492 l/h
min. cooling water differential pressure1 bar
max. cooling water pressure6 bar
min. filling capacity4,9 l
Filling capacity expansion tank16 l
Overall dimensions WxDxH **540x604x1332 mm
Net weight208 kg
sound pressure level +/- 4 dB(A)61 dB(A)
Power supply requirement400V 3~ 50Hz
max. current14,5 A
Fuse3x16A
Pressure equipment categoryI
Degree of ProtectionIP20
min. ambient temperature5 °C
max. ambient temperature40 °C
from Serial-No.:S412416

** Please respect space requirements. See operating conditions at www.huber-online.com

Technical data according to DIN 12876

CharacteristicUnistat P520w
Order-No.1072.0007.01
Operating temperature range-55...250 °C
Temperature stability at -10°C0,01 K
temperature set point / display5,7" colour Touchscreen
Resolution of display0,01 K
Internal temperature sensorPt100
Sensor external connectionPt100
Interface digitalEthernet, USB (Host u. Device), RS232
digital inputECS ONE
digital outputPOKO ONE
Alarm messageoptic, acoustic, relay
Safety classificationIII / FL
Heating power6 kW
Cooling power with Thermooil at 250°C6 kW
Cooling power with Thermooil at 200°C6 kW
Cooling power with Thermooil at 100°C6 kW
Cooling power with Thermooil at 20°C6 kW
Cooling power with Ethanol at 0°C6 kW
Cooling power with Ethanol at -20°C4,2 kW
Cooling power with Ethanol at -40°C1,5 kW
Cooling power with Ethanol at -50°C0,65 kW
Refrigeration machinewater-cooled, CFC- and HCFC-free
Refrigerant (ASHRAE, GHS)R-452A (A1, H280)
Global Warming Potential (GWP)2141
Refrigerant quantity1,44 kg
CO2 equivalent3,1 t
Circulation pump: MK pump max. delivery82 l/min
Circulation pump: MK pump max. delivery pressure3 bar
Circulation pump: MK pump Delivery at 0,5 bar75 l/min
Circulation pump: MK pump Delivery at 1,0 bar66 l/min

Engineering Facts

14 facts · every claim linked to a page
Installation

The unit requires a 400 V three-phase supply with a 3x16 A fuse, so the electrical installation must be sized accordingly.

Determines the site power supply, cable and breaker rating needed before installation.

Source
“Power supply requirement 400V 3~ 50Hz max. current 14,5 A Fuse 3x16A”
Installation

The cooling water connection is G1/2 male and needs at least 1 bar differential pressure and no more than 6 bar, with about 492 l/h at 15 °C.

Sizes the cooling water supply, piping and pressure regulation for reliable heat rejection.

Source
“Cooling water connection G1/2 male Consumption at water 15°C, flow 20°C 492 l/h min. cooling water differential pressure 1 bar max. cooling water pressure 6 bar”
Installation

The unit is designed for cooling water intake between 5 °C and 20 °C, and pipes must be insulated if the water can fall below the condensation point.

Prevents condensation and ensures the specified cooling performance is available.

Source
“This temperature control unit is designed to operate with cooling water intake temperature between 5°C and 20°C. When cooling water temperatures fall below the condensation point, it is essential to insulate the cooling water pipes in the unit.”
Performance

Cooling capacity drops significantly at low temperatures with ethanol, from 6 kW at 0 °C to 0.65 kW at -50 °C, so low-temperature duty must be checked against the load.

Confirms whether the unit can hold the required low temperature under the actual process heat load.

Source
“Cooling power with Ethanol at 0°C 6 kW at -20°C 4,2 kW at -40°C 1,5 kW at -50°C 0,65 kW”
Performance

The circulation pump delivers 82 l/min at zero head but only 32 l/min at 2.5 bar, so system pressure drop must be matched to the pump curve.

Ensures adequate flow through high-pressure-drop external circuits such as flow-through chemistry.

Source
“max. delivery 82 l/min max. delivery pressure 3 bar Delivery at 0,5 bar 75 l/min Delivery at 1,0 bar 66 l/min”
Selection

The P models are specifically intended for applications with high pressure drops, such as flow-through chemistry or semiconductor industry.

Guides model selection when the external loop has high hydraulic resistance.

Source
“Unistat „P“ Models: Circulating pumps with a high discharge pressure for applications with high pressure drops, e.g. in the Flow-Through chemistry or in the Semicon industry.”
Safety

The refrigerant R-452A has a GWP of 2141 and a charge of 1.44 kg, giving 3.1 t CO2 equivalent that must be considered for environmental compliance and leak management.

Affects refrigerant handling, leak detection obligations and environmental permitting.

Source
“Refrigerant (ASHRAE, GHS) R-452A (A1, H280) Global Warming Potential (GWP) 2141 Refrigerant quantity 1,44 kg CO2 equivalent 3,1 t”
Safety

Leakage current exceeds 3.5 mA, so protective earthing and residual current protection must be designed for this.

Requires appropriate earth connection and RCD selection to avoid nuisance tripping or shock hazard.

Source
“Attention: leakage current > 3,5mA”
Installation

Voltage and frequency tolerances are ±5% and ±2% but must not combine in the same direction, e.g. -5% voltage with +2% frequency is not allowed.

Defines the acceptable site power quality for safe and reliable operation.

Source
“Voltage + / - 5% with a simultaneous frequency tolerance of + / - 2% Example -5% voltage and +2% frequency -> not allowed!”
Installation

Three-phase devices below 63 A are supplied with cable but without a plug, so the customer must provide the site plug and connection.

Clarifies what electrical termination work is required on site.

Source
“Three-phase devices with current consumption less than 63A --> with cable, without plug”
Operation

The unit is rated for ambient temperatures from 5 °C to 40 °C, so the installation room must stay within this range.

Prevents derating or faults caused by an unsuitable ambient environment.

Source
“min. ambient temperature 5 °C max. ambient temperature 40 °C”
Installation

The unit weighs 208 kg and measures 540x604x1332 mm, so floor loading and access routes must be planned.

Determines handling equipment, doorway clearance and floor load capacity.

Source
“Overall dimensions WxDxH ** 540x604x1332 mm Net weight 208 kg”
Conformité

The equipment is EMC class A group 1, which is intended for industrial environments and may cause interference in residential areas.

Affects installation location and EMC coordination with sensitive equipment nearby.

Source
“Classification (disturbance) to EN55011: Class A, Group 1”
Operation

The pump connections use bore shape Y (60°) per DIN 3863 with ball socket and sleeve nut per DIN 3870, so matching hoses and fittings must be ordered.

Ensures correct mating fittings and avoids leaks at the M30x1.5 pump connections.

Source
“Pump connections: Bore shape Y (60°) according to DIN 3863, pipework/flexible tempering hoses: Ball socket according to DIN 3863, sleeve nut according to DIN 3870.”