Introduces the Unistat P610 unit and its main features
Specpp. 1-2
Performance, temperature, pressure and connection data per DIN 12876
p. 2table, 20 rows
p. 1table, 26 rows
Technical data according to DIN 12876
Parameter
Value
Delivery at 1,0 bar
68 l/min
Delivery at 1,5 bar
59 l/min
Delivery at 2,0 bar
49 l/min
Delivery at 2,5 bar
36 l/min
Pump connection
M30x1,5 male
max. permissible kin. viscosity
50 mm²/s
min. filling capacity
6,5 l
Filling capacity expansion tank
26 l
Overall dimensions Temperature Control Unit WxDxH
1290x1005x1540 mm
Dimensions of External Transformer WxDxH
364x740x682 mm
Power supply factory configured (3 Phase)
208V 3~ 60Hz
Input Voltage External Transformer
208V 3~ 60Hz
Output Voltage External Transformer
440V 3~ 60Hz
Power Supply for Temperature Control Unit
460V 3~ 60HZ
Max. current temp. control unit with external transformer
90 A
Fuse input voltage external transformer
3x100A
Degree of Protection
IP20
min. ambient temperature
5 °C
max. ambient temperature
40 °C
from Serial-No.:
1.0/24
Delivery at 1,0 bar; Technical details and dimensions are subject to change. No liability is accepted for errors or omissions. Illustrations can deviate from the original.
Technical data according to DIN 12876
Parameter
Unistat P610
Order-No.
1052.0040.01
Operating temperature range
-60...200 °C
Temperature stability at -10°C
0,01 K
temperature set point / display
5,7" colour Touchscreen
Resolution of display
0,01 K
Internal temperature sensor
Pt100
Sensor external connection
Pt100
Interface digital
Ethernet, USB (Host u. Device), RS232
digital input
ECS ONE
digital output
POKO ONE
Alarm message
optic, acoustic, relay
Safety classification
III / FL
Heating power
6 kW
Cooling power with Thermooil at 200°C
7 kW
Cooling power with Thermooil at 100°C
7 kW
Cooling power with Ethanol at 0°C
7 kW
Cooling power with Ethanol at -20°C
6,4 kW
Cooling power with Ethanol at -40°C
2,6 kW
Cooling power with Ethanol at -60°C
0,05 kW
Refrigeration machine
air-cooled, natural refrigerant
Refrigerant (ASHRAE, GHS)
R-1270 (A3, H220)
Global Warming Potential (GWP)
0
UN-number
UN 3358
Circulation pump: MK pump max. delivery
82 l/min
Circulation pump: MK pump max. delivery pressure
3 bar
Circulation pump: MK pump Delivery at 0,5 bar
75 l/min
Engineering Facts
14 facts · every claim linked to a page
Safety
The unit uses refrigerant R-1270 (A3, H220), a flammable natural refrigerant, so the installation location must be assessed for A3 flammability zoning and ignition sources.
A3 refrigerants are flammable; this drives room ventilation, ignition-source control and local code compliance at the installation site.
Source
“Refrigerant (ASHRAE, GHS) R-1270 (A3, H220)”
Installation
The unit is supplied for 460 V 3~ 60 Hz operation but requires an external transformer fed from 208 V 3~ 60 Hz, so the electrical design must include the transformer footprint and its 3x100 A input fusing.
The transformer is a separate large cabinet (364x740x682 mm) that must be accommodated and wired by the electrical contractor.
Source
“Power Supply for Temperature Control Unit 460V 3~ 60HZ”
Installation
Maximum current of the temperature control unit with external transformer is 90 A, so feeder, disconnect and protection must be sized for at least this current.
Undersizing the branch circuit or protection will cause nuisance tripping or non-compliance with UL 489 requirements.
Source
“Max. current temp. control unit with external transformer 90 A”
Safety
Leakage current exceeds 3.5 mA, so a high-leakage-current protective device or reinforced earthing per local code is required.
Standard 30 mA RCDs may trip; the earthing and protection scheme must be designed accordingly.
Source
“Attention: leakage current > 3,5mA”
Conformité
External branch circuit protection according to UL 489 is required, so UL-listed molded-case or similar breakers must be specified in the panel.
This is a mandatory listing requirement for the North American installation and affects the choice of breaker type.
Source
“External branch circuit protection according to UL 489 required.”
Installation
Three-phase devices above 63 A are delivered without cable and without plug, so the customer must supply and terminate the power cable on site.
Cable, lugs and termination labor must be included in the installation scope and budget.
Source
“Three-phase devices with current consumption greater than 63A --> without cable, without plug”
Installation
Voltage and frequency tolerances are +/−5% and +/−2% but must not combine in the unfavorable direction, so the site power quality must be verified before commissioning.
A weak or unstable supply can push the unit outside its EN60034-1 tolerance and void reliable operation.
Source
“-5% voltage and +2% frequency -> not allowed!”
Installation
Pump connections are M30x1.5 male with bore shape Y (60°) per DIN 3863, so mating hoses must use ball socket DIN 3863 and sleeve nut DIN 3870.
Wrong fittings will not seal; specifying the correct DIN 3863/3870 hose ends avoids leaks and rework.
Source
“Pump connection M30x1,5 male”
Performance
Cooling capacity collapses from 2.6 kW at −40 °C to only 0.05 kW at −60 °C, so the −60 °C end of the range is only suitable for low-heat-load holding, not for pull-down.
Applications needing meaningful cooling near −60 °C will be undersized; the process heat load must be checked against this curve.
Source
“at -60°C 0,05 kW”
Performance
Cooling capacity is rated at 20 °C room temperature and drops as ambient rises, so the installed cooling capacity must be derated for the actual plant room temperature.
Sizing on catalog values in a hot room leads to insufficient cooling; the ambient range is 5–40 °C.
Source
“If the ambient temperature rises, the cooling capacity may drop.”
Installation
The unit is IP20 and rated for 5–40 °C ambient, so it must be installed indoors in a clean, non-condensing environment with adequate clearance.
IP20 offers no water protection; washdown or outdoor locations are excluded and space requirements must be respected.
Source
“Degree of Protection IP20”
Operation
Minimum filling capacity is 6.5 l with a 26 l expansion tank, so the external loop volume and fluid expansion must be checked against these figures.
Too little fluid causes pump cavitation and poor control; too much expansion volume overflows the tank.
Source
“min. filling capacity 6,5 l”
Operation
Maximum permissible kinematic viscosity is 50 mm²/s, so the heat transfer fluid must be selected to stay below this at the lowest operating temperature.
A fluid that is too viscous at low temperature overloads the pump and degrades heat transfer.
Source
“max. permissible kin. viscosity 50 mm²/s”
Selection
The P-series pump delivers up to 3 bar and 82 l/min but only 36 l/min at 2.5 bar, so high-pressure-drop loops must be checked against the actual pump curve.
A loop with high pressure drop will see sharply reduced flow; the system curve must intersect the pump curve acceptably.