Product introduction and general description of the Unistat P620w unit
Specp. 2
Technical data according to DIN 12876 standard
p. 2table, 17 rows
p. 1table, 31 rows
Technical data according to DIN 12876
Parameter
Value
Delivery at 5,0 bar
42 l/min
Pump connection
M38x1,5 male
max. permissible kin. viscosity
100 mm²/s
Cooling water connection
G3/4 male
min. cooling water differential pressure
1 bar
max. cooling water differential pressure
4 bar
max. cooling water pressure
6 bar
max. System pressure
6 bar
min. filling capacity
8,9 l
Filling capacity expansion tank
19,9 l
Overall dimensions WxDxH **
918x963x1771 mm
Power supply requirement
400V 3~ 50Hz
Pressure equipment category
II
Degree of Protection
IP20
min. ambient temperature
5 °C
max. ambient temperature
40 °C
from Serial-No.:
1.0/24
** Please respect space requirements. See operating conditions at www.huber-online.com
Unistat P620w — Technical data according to DIN 12876
Characteristic
Value
Order-No.
5016.0002.01
Operating temperature range
-60...200 °C
Temperature stability at -10°C
0,01 K
temperature set point / display
5,7" colour Touchscreen
Resolution of display
0,01 K
Internal temperature sensor
Pt100
Sensor external connection
Pt100
Interface digital
Ethernet, USB (Host u. Device), RS232
digital input
ECS ONE
digital output
POKO ONE
Interface analog
optional
Alarm message
optic, acoustic, relay
Safety classification
III / FL
Heating power
12 kW
Cooling power with Thermooil at 200°C
12 kW
Cooling power with Thermooil at 100°C
12 kW
Cooling power with Thermooil at 20°C
12 kW
Cooling power with Ethanol at 0°C
12 kW
Cooling power with Ethanol at -20°C
12 kW
Cooling power with Ethanol at -40°C
5,6 kW
Cooling power with Ethanol at -60°C
1,4 kW
Refrigeration machine
water-cooled, natural refrigerant
Refrigerant (ASHRAE, GHS)
R-1270 (A3, H220)
Global Warming Potential (GWP)
0
Circulation pump
MK pump
max. delivery
201 l/min
max. delivery pressure
5,3 bar
Delivery at 1,0 bar
179 l/min
Delivery at 2,0 bar
154 l/min
Delivery at 3,0 bar
124 l/min
Delivery at 4,0 bar
87 l/min
Engineering Facts
12 facts · every claim linked to a page
Installation
The unit requires a 400 V three-phase supply, so the electrical infrastructure must be verified before installation.
Determines whether the site has the correct three-phase power available and affects cabling and breaker sizing.
Source
“Power supply requirement
400V 3~ 50Hz”
Installation
Cooling water must be supplied at 1–4 bar differential pressure and no more than 6 bar, so the facility water loop must be checked against these limits.
Incorrect cooling water pressure can reduce cooling capacity or damage the water circuit.
Source
“min. cooling water differential pressure
1 bar
max. cooling water differential pressure
4 bar
max. cooling water pressure
6 bar”
Operation
The unit is designed for cooling water inlet temperatures between 5 °C and 20 °C, and lower temperatures require insulating the cooling water pipes.
Operating outside this range risks condensation and reduced performance.
Source
“This temperature control unit is designed to operate with cooling water intake temperature between 5°C and 20°C.”
Performance
Cooling capacity drops significantly at low temperatures with ethanol, from 12 kW at -20 °C to 5.6 kW at -40 °C and 1.4 kW at -60 °C.
Process design must account for reduced cooling power at the lowest operating temperatures.
Source
“at -40°C
5,6 kW
at -60°C
1,4 kW”
Installation
The pump connection is M38x1.5 male, so mating hoses or adapters must match this thread.
Ensures correct connection of external circuit plumbing and avoids leaks or adapters.
Source
“Pump connection
M38x1,5 male”
Safety
The refrigerant is R-1270 (A3, H220), a flammable natural refrigerant, which affects site safety classification and handling procedures.
Flammable refrigerant requires appropriate safety measures and may affect insurance or room classification.
Source
“R-1270 (A3, H220)”
Conformité
Voltage tolerance is ±5% only with simultaneous frequency tolerance of ±2%, and certain combinations are not allowed.
Power quality must be verified to avoid voiding warranty or causing malfunction.
Source
“Voltage + / - 5% with a simultaneous frequency tolerance of + / - 2%”
Installation
Three-phase devices with current consumption less than 63 A are supplied with cable but without a plug, so a suitable plug must be provided on site.
Affects electrical installation planning and cost.
Source
“Three-phase devices with current consumption less than 63A --> with cable, without plug”
Operation
The maximum permissible kinematic viscosity of the heat transfer fluid is 100 mm²/s, which limits fluid selection at low temperatures.
Using a fluid with too high viscosity can overload the pump and reduce flow.
Source
“max. permissible kin. viscosity
100 mm²/s”
Installation
The unit has IP20 protection, so it must be installed in a dry indoor environment protected from water ingress.
Prevents installation in wet or dusty environments that could damage electronics.
Source
“Degree of Protection
IP20”
Operation
The ambient temperature must be between 5 °C and 40 °C for proper operation.
Ensures the unit operates within its design envelope and maintains performance.
Source
“min. ambient temperature
5 °C
max. ambient temperature
40 °C”
Maintenance
The cooling water circuit contains copper, stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM, so suitable cooling water quality must be used to avoid corrosion.
Poor water quality can cause corrosion and fouling, reducing cooling efficiency and requiring maintenance.
Source
“Materiels used in the cooling water circuit include; copper, Stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM.”