Unistat P635w Technical Data

Sommaire

  1. Overview p. 1
    Introduces the Unistat P635w unit and its main features
  2. Spec pp. 1-2
    Lists performance, temperature, and physical specifications per DIN 12876
  3. p. 2 table, 17 rows
  4. p. 1 table, 31 rows

Technical data according to DIN 12876

ParameterValue
Delivery at 5,0 bar42 l/min
Pump connectionM38x1,5 male
max. permissible kin. viscosity100 mm²/s
Cooling water connectionG3/4 male
min. cooling water differential pressure1 bar
max. cooling water differential pressure4 bar
max. cooling water pressure6 bar
max. System pressure6 bar
min. filling capacity8,9 l
Filling capacity expansion tank19,9 l
Overall dimensions WxDxH **918x963x1771 mm
Power supply requirement400V 3~ 50Hz
Pressure equipment categoryI
Degree of ProtectionIP20
min. ambient temperature5 °C
max. ambient temperature40 °C
from Serial-No.:1.0/24

** Please respect space requirements. See operating conditions at www.huber-online.com

Technical data according to DIN 12876

CharacteristicUnistat P635w
Order-No.5019.0101.01
Operating temperature range-60...200 °C
Temperature stability at -10°C0,01 K
temperature set point / display5,7" colour Touchscreen
Resolution of display0,01 K
Internal temperature sensorPt100
Sensor external connectionPt100
Interface digitalEthernet, USB (Host u. Device), RS232
digital inputECS ONE
digital outputPOKO ONE
Interface analogoptional
Alarm messageoptic, acoustic, relay
Safety classificationIII / FL
Heating power24 kW
Cooling power with Thermooil at 200°C27 kW
Cooling power with Thermooil at 100°C27 kW
Cooling power with Thermooil at 20°C27 kW
Cooling power with Ethanol at 0°C27 kW
Cooling power with Ethanol at -20°C25 kW
Cooling power with Ethanol at -40°C14 kW
Cooling power with Ethanol at -60°C3,5 kW
Refrigeration machinewater-cooled, CFC- and HCFC-free
Refrigerant (ASHRAE, GHS)R-452A (A1, H280)
Global Warming Potential (GWP)2141
Circulation pumpMK pump
max. delivery201 l/min
max. delivery pressure5,3 bar
Delivery at 1,0 bar179 l/min
Delivery at 2,0 bar154 l/min
Delivery at 3,0 bar124 l/min
Delivery at 4,0 bar87 l/min

Engineering Facts

14 facts · every claim linked to a page
Installation

The unit requires a 400 V three-phase supply and must be hard-wired without a plug because its current exceeds 63 A.

Determines electrical supply design, cable sizing and whether a plug/socket is needed at the installation site.

Source
“Three-phase devices with current consumption greater than 63A --> without cable, without plug”
Installation

Cooling water must be supplied at 1–4 bar differential pressure and no more than 6 bar static pressure, with inlet temperature between 5 °C and 20 °C.

Defines the facility cooling-water loop requirements and protects the water-cooled refrigeration circuit.

Source
“min. cooling water differential pressure 1 bar max. cooling water differential pressure 4 bar”
Installation

If cooling water is colder than the condensation point, the cooling water pipes inside the unit must be insulated.

Prevents condensation, corrosion and electrical risk when using cold facility water.

Source
“When cooling water temperatures fall below the condensation point, it is essential to insulate the cooling water pipes in the unit.”
Performance

Cooling capacity is rated at 20 °C room temperature, 15 °C cooling water inlet and 1 bar differential pressure, and drops as cooling water temperature rises.

Allows correct comparison of quoted cooling power against actual site conditions.

Source
“Output data valid for: Room temperature 20°C, cooling water inlet 15°C and 1 bar differential pressure between cooling water inlet and -”
Installation

Voltage and frequency tolerances are ±5 % and ±2 % respectively, but the combination of −5 % voltage with +2 % frequency is not permitted.

Guides power quality assessment and generator/UPS compatibility checks.

Source
“-5% voltage and +2% frequency -> not allowed!”
Safety

The unit is classified as pressure equipment category I and has safety classification III / FL according to DIN 12876.

Relevant for risk assessment, permissible fluids and compliance documentation.

Source
“Safety classification III / FL”
Safety

The refrigerant R-452A is classified A1 with GHS hazard H280, and has a GWP of 2141.

Affects refrigerant handling, leak detection, F-gas reporting and environmental compliance.

Source
“R-452A (A1, H280)”
Operation

The pump delivers 201 l/min maximum and 42 l/min at 5.0 bar, with a maximum delivery pressure of 5.3 bar.

Needed to match the circulator to external circuit pressure drop and required flow.

Source
“max. delivery 201 l/min max. delivery pressure 5,3 bar”
Operation

The maximum permissible kinematic viscosity of the heat transfer fluid is 100 mm²/s.

Limits fluid selection at low temperatures and protects the pump.

Source
“max. permissible kin. viscosity 100 mm²/s”
Installation

The unit needs a minimum filling capacity of 8.9 l and has a 19.9 l expansion tank.

Determines heat transfer fluid volume to procure and expansion space planning.

Source
“min. filling capacity 8,9 l Filling capacity expansion tank 19,9 l”
Installation

The unit is IP20 and must be operated between 5 °C and 40 °C ambient temperature.

Defines room requirements and protection against dust and water ingress.

Source
“Degree of Protection IP20 min. ambient temperature 5 °C max. ambient temperature 40 °C”
Installation

Pump connections are M38x1.5 male and cooling water connections are G3/4 male.

Needed to order matching hoses, adapters and fittings before delivery.

Source
“Pump connection M38x1,5 male”
Conformité

The equipment is EMC class A, group 1, so it is intended for industrial environments and may cause radio interference in residential areas.

Affects installation location and EMC planning in laboratory or mixed-use buildings.

Source
“Classification (disturbance) to EN55011: Class A, Group 1”
Maintenance

The cooling water circuit contains copper, stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM, so only suitable cooling water may be used.

Guides water quality and material compatibility to avoid corrosion and fouling.

Source
“Materiels used in the cooling water circuit include; copper, Stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM.”