Product introduction and general description of the Unistat T320 unit
Specp. 2
Technical data according to DIN 12876 including performance parameters
p. 1table, 29 rows
Technical data according to DIN 12876
Characteristic
Value
Operating temperature range
65... 300 °C
Temperature stability at 70°C
0,01 K
temperature set point / display
5,7" colour Touchscreen
Resolution of display
0,01 K
Internal temperature sensor
Pt100
External sensor
Pt100
Interface digital
Ethernet, USB (Host u. Device), RS232
digital input
ECS ONE
digital output
POKO ONE
Alarm message
optic, acoustic, relay
Safety classification
III / FL
Heating power
12 kW
Circulation pump:
MK pump
max. delivery
96 l/min
max. delivery pressure
3,5 bar
Pump connection
M30x1,5 male
max. permissible kin. viscosity
50 mm²/s
min. filling capacity
3 l
Filling capacity expansion tank
24 l
Overall dimensions WxDxH **
540x749x1184 mm
Net weight
168 kg
Power supply (3 Phase)
460V 3~ 60Hz
max. current (3 Phase)
17 A
Fuse (3 phase)
3x20A
Pressure equipment category
Art. 4.3 PED
Degree of Protection
IP20
min. ambient temperature
5 °C
max. ambient temperature
40 °C
from Serial-No.:
S603483
** Overall dimensions WxDxH
Engineering Facts
14 facts · every claim linked to a page
Selection
The unit has no active refrigeration, so reaching the 65 °C minimum requires adequate heat loss from the application or suitable water cooling.
Prevents specifying the T320 for cooling duties below ambient without an external heat sink or water cooling.
Source
“These devices do not have an active refrigeration system.”
Selection
For air-cooled operation the application heat loss must fall between 0.1 kW and 2.8 kW depending on the unit, and ambient must be well below 60 °C to reach 65 °C.
Defines the allowable process load window and ambient limit for air-cooled installations.
Source
“The heat loss must be between 0.1kW and 2.8kW, depending on the unit.”
Installation
Three-phase devices drawing less than 63 A are supplied with cable but without a plug, so the customer must provide the site plug.
Affects electrical installation planning and bill of materials for the 460 V 3~ supply.
Source
“Three-phase devices with current consumption less than 63A --> with cable, without plug”
Installation
Voltage and frequency tolerances are coupled: ±5 % voltage is only valid with ±2 % frequency, and a -5 % voltage with +2 % frequency combination is not allowed.
Must be checked against the site power quality before commissioning to avoid invalid supply conditions.
Source
“-5% voltage and +2% frequency -> not allowed!”
Installation
Pump connections use bore shape Y (60°) per DIN 3863 with ball socket and sleeve nut per DIN 3870, so mating hoses and fittings must match these standards.
Ensures correct selection of tempering hoses and adapters for the M30x1.5 pump connection.
Source
“Bore shape Y (60°) according to DIN 3863, pipework/flexible tempering hoses: Ball socket according to DIN”
Safety
The unit is classified III / FL per DIN 12876 and includes adjustable overtemperature protection, which governs the safety setup and fluid selection.
Determines the required safety category and the need to set and verify the overtemperature cut-out during commissioning.
Source
“With adjustable overtemperature protection according to DIN 12876.”
Operation
The expansion tank is not thermoregulated and is lockable for open baths, which affects how the tank is used in closed versus open configurations.
Influences setup of the expansion volume and whether the bath can be operated open.
Source
“Expansion tank (not thermoregulated) for closed systems, lockable for open baths.”
Operation
The pump is a magnetically coupled stainless-steel variable-speed pump with soft start and integrated pressure control, optionally with an external pressure sensor.
Guides pump configuration and the option to add external pressure control for process loops.
Source
“Powerful variable speed pump (soft start) with integrated pressure control with optional external pressure sensor.”
Operation
The maximum permissible kinematic viscosity of the circulating fluid is 50 mm²/s, limiting the choice of thermofluid at low temperatures.
Constrains thermofluid selection to keep pump load and flow within design limits.
Source
“max. permissible kin. viscosity”
Operation
The minimum filling capacity is 3 l and the expansion tank holds 24 l, setting the minimum fluid charge and expansion allowance.
Used to size the initial fill and verify expansion volume for the application loop.
Source
“min. filling capacity”
Installation
The unit is rated IP20 and must be operated between 5 °C and 40 °C ambient, restricting placement in plant rooms.
Limits installation location to dry, protected areas within the stated ambient range.
Source
“Degree of Protection”
Conformité
The pressure equipment falls under Article 4.3 PED, which affects the conformity documentation required for the installation.
Determines the PED documentation and marking to verify at delivery.
Source
“Pressure equipment category”
Performance
Heating power is 12 kW with a circulation pump delivering up to 96 l/min at 3.5 bar, defining the achievable heat-up rate and loop pressure budget.
Used to check whether the unit can meet the required heat-up time and pressure drop of the process loop.
Source
“Heating power”
Selection
The unit requires a 460 V 3~ 60 Hz supply with 17 A max current and 3x20 A fusing, which must match the site electrical infrastructure.
Confirms site voltage, frequency and breaker sizing before ordering.