Unistat T340 Temperature Control System Technical Data
Sommaire
Overviewp. 1
Product introduction and model identification for Unistat T340
Specp. 2
Technical data according to DIN 12876 standard
p. 1table, 25 rows
Technical data according to DIN 12876
Parameter
Value
Operating temperature range
65... 300 °C
Temperature stability at 70°C
0,01 K
temperature set point / display
5,7" colour Touchscreen
Resolution of display
0,01 K
Internal temperature sensor
Pt100
External sensor
Pt100
Interface digital
Ethernet, USB (Host u. Device), RS232
digital input
ECS ONE
digital output
POKO ONE
Alarm message
optic, acoustic, relay
Safety classification
III / FL
Heating power
48 kW
Circulation pump:
MK pump
max. delivery
200 l/min
max. delivery pressure
5,5 bar
Pump connection
M38x1,5 male
max. permissible kin. viscosity
50 mm²/s
Filling capacity expansion tank
56 l
Overall dimensions WxDxH **
800x1007x1607 mm
Net weight
373 kg
Power supply (3 Phase)
460V 3~ 60Hz
Degree of Protection
IP20
min. ambient temperature
5 °C
max. ambient temperature
40 °C
from Serial-No.:
1.0/25
** Technical details and dimensions are subject to change. No liability is accepted for errors or omissions. Illustrations can deviate from the original.
Engineering Facts
14 facts · every claim linked to a page
Selection
The unit has no active refrigeration, so reaching 65 °C requires heat loss from the application or suitable water cooling.
Determines whether the application needs external cooling or water supply to achieve low setpoints.
Source
“These devices do not have an active refrigeration system.”
Installation
For water-cooled operation, target temperatures down to 65 °C can only be achieved with appropriate heat loss from the application or suitable water cooling.
Requires planning of water cooling capacity and heat dissipation before installation.
Source
“Water cooled: Target temperatures (down to the minimum 65°C) can only be achieved with appropriate heat loss from the application or suitable water cooling.”
Installation
For air-cooled operation, the heat loss must be between 0.1 kW and 2.8 kW depending on the unit, and ambient temperature must be well below 60 °C to reach 65 °C.
Sets limits on application heat load and room conditions for air-cooled installations.
Source
“Air cooled: The heat loss must be between 0.1kW and 2.8kW, depending on the unit. In order to reach 65°C the ambient temperature must be well below 60°C.”
Installation
Three-phase devices drawing less than 63 A are supplied with cable but without a plug.
The installer must provide and fit a suitable plug on site.
Source
“Three-phase devices with current consumption less than 63A --> with cable, without plug”
Installation
Three-phase devices drawing more than 63 A are supplied without cable and without plug.
Requires on-site power cabling and connection hardware to be planned.
Source
“Three-phase devices with current consumption greater than 63A --> without cable, without plug”
Safety
The leakage current exceeds 3.5 mA, which affects protective earthing and RCD selection.
Needs proper earthing and appropriate residual current protection design.
Source
“Attention: leakage current > 3,5mA”
Conformité
Voltage tolerance is ±5% only with simultaneous frequency tolerance of ±2%, and certain combinations are not allowed.
Defines acceptable supply quality for reliable operation and warranty compliance.
Source
“Voltage + / - 5% with a simultaneous frequency tolerance of + / - 2%”
Conformité
The example states that -5% voltage with +2% frequency is not allowed, while -5% voltage with -2% frequency is allowed.
Clarifies which combined supply deviations are unacceptable for installation.
Source
“-5% voltage and +2% frequency -> not allowed!”
Conformité
The equipment is classified as Class A, Group 1 for electromagnetic disturbance to EN55011.
Affects EMC planning and suitability for certain environments.
Source
“Classification (disturbance) to EN55011: Class A, Group 1”
Installation
Pump connections use bore shape Y (60°) according to DIN 3863, which must match pipework or flexible hoses.
Ensures correct mating of hoses and piping to the pump connection.
Source
“Note: Pump connections: Bore shape Y (60°) according to DIN 3863, pipework/flexible”
Operation
The pump is a magnetically coupled stainless steel circulation pump with automatic switch-over and capacity adaptation for heating.
Informs about pump construction and automatic capacity behavior during operation.
Source
“Closed, magnetically coupled circulation pump made of stainless steel. Automatical switch-over and capacity adaption for heating.”
Operation
The expansion tank is not thermoregulated and is lockable for open baths.
Affects how the expansion tank is used and configured for closed versus open bath operation.
Source
“Expansion tank (not thermoregulated) for closed systems, lockable for open baths.”
Safety
The unit has adjustable overtemperature protection according to DIN 12876.
Requires setting and verifying the overtemperature limit during commissioning.
Source
“With adjustable overtemperature protection according to DIN 12876.”
Maintenance
The maximum permissible kinematic viscosity of the medium is 50 mm²/s.
Limits the choice of thermofluid to ensure pump performance and avoid damage.