Product introduction and general description of the unit
Specpp. 1-2
Technical data according to DIN 12876 and performance figures
p. 2table, 2 rows
p. 1table, 34 rows
Technical data according to DIN 12876
Parameter
Value
min. ambient temperature
5 °C
from Serial-No.:
1.1/26
Unistat T345w HT — Technical data according to DIN 12876
Parameter
Value
Order-No.
1042.0011.01
Operating temperature range
65...300 °C
Minimum temperature with water cooling
15 °C
temperature set point / display
5,7" colour Touchscreen
Resolution of display
0,01 K
Internal temperature sensor
Pt100
Sensor external connection
Pt100
Interface digital
Ethernet, USB (Host u. Device), RS232
digital input
ECS ONE
digital output
POKO ONE
Alarm message
optic, acoustic, relay
Safety classification
III / FL
Heating power
72 kW
Cooling power at 300°C
40 kW
Cooling power at 200°C
40 kW
Cooling power at 100°C
24 kW
Circulation pump
MK pump
max. delivery
200 l/min
max. delivery pressure
5,5 bar
Pump connection
M38x1,5 male
max. permissible kin. viscosity
50 mm²/s
Cooling water consumption at water temp. 15°C
1600 l/h
Cooling water connection
G1 1/4 male
min. cooling water differential pressure
1 bar
max. cooling water pressure
6 bar
min. filling capacity
33 l
Filling capacity expansion tank
56 l
Overall dimensions WxDxH **
800x1007x1607 mm
Net weight
445 kg
Power supply (3 Phase)
400V 3~ 50Hz
max. current (3 Phase)
115 A
Fuse (3 phase)
3x125A
Degree of Protection
IP20
max. ambient temperature
40 °C
** Overall dimensions WxDxH
Engineering Facts
12 facts · every claim linked to a page
Installation
The unit requires a three-phase 400 V supply with a 125 A fuse per phase, so the electrical installation must be sized for a 115 A maximum current.
Determines the site power distribution, cable cross-section and protection device rating before ordering.
Source
“Power supply (3 Phase) 400V 3~ 50Hz”
Installation
The cooling water circuit needs a minimum differential pressure of 1 bar and can withstand a maximum of 6 bar, with G1 1/4 male connections.
Defines the required cooling water supply pressure and the fittings/adapters needed for connection.
Source
“min. cooling water differential pressure 1 bar”
Performance
Cooling water consumption is 1600 l/h at 15 °C water temperature, so the facility must provide that flow rate continuously.
Sizing of the cooling water supply and drain, and verification that the building loop can deliver this flow.
Source
“Cooling water consumption at water temp. 15°C 1600 l/h”
Operation
The unit has no active refrigeration system, so reaching the 65 °C minimum target depends on adequate heat loss from the application or suitable water cooling.
Prevents specifying the unit for cooling duties it cannot perform without external heat removal.
Source
“These devices do not have an active refrigeration system.”
Installation
The cooling water intake temperature must be between 5 °C and 20 °C, and pipes must be insulated if the water temperature falls below the condensation point.
Ensures the site cooling water temperature is compatible and that condensation is avoided.
Source
“This temperature control unit is designed to operate with cooling water intake temperature between 5°C and 20°C.”
Conformité
Voltage and frequency tolerances are ±5 % and ±2 % respectively, but they must not be applied in opposite directions simultaneously.
Confirms whether the site power quality is acceptable for warranty and safe operation.
Source
“Voltage + / - 5% with a simultaneous frequency tolerance of + / - 2%”
Safety
The leakage current exceeds 3.5 mA, which requires appropriate protective earthing and possibly a higher-rated residual current device.
Affects the electrical protection concept and RCD selection during installation.
Source
“Attention: leakage current > 3,5mA”
Installation
Three-phase devices drawing more than 63 A are delivered without cable and without plug, so the customer must provide the power connection.
Clarifies the scope of supply and the need to plan for on-site cabling and a suitable isolator.
Source
“Three-phase devices with current consumption greater than 63A --> without cable, without plug”
Installation
The pump connection is M38x1.5 male with a bore shape Y (60°) per DIN 3863, so hoses must use a matching ball socket and sleeve nut.
Ensures the correct mating hose fittings are ordered to avoid leaks or delays.
Source
“Pump connection M38x1,5 male”
Operation
The maximum permissible kinematic viscosity of the thermofluid is 50 mm²/s, which limits the choice of heat transfer fluid at low temperatures.
Prevents selecting a fluid that is too viscous for the pump at the operating temperature.
Source
“max. permissible kin. viscosity 50 mm²/s”
Installation
The unit has IP20 protection and a maximum ambient temperature of 40 °C, so it must be installed in a dry, ventilated indoor location.
Rules out outdoor or wet/dusty environments and informs room ventilation requirements.
Source
“Degree of Protection IP20”
Maintenance
The cooling water circuit contains copper, stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM, so only compatible cooling water may be used.
Guides water quality specification to avoid corrosion or scaling of the heat exchanger.
Source
“Materiels used in the cooling water circuit include; copper, Stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM.”