Unistat TR401 Technical Data

Sommaire

  1. Overview p. 1
    Product name and general introduction to the Unistat TR401 unit
  2. Spec p. 2
    Technical data according to DIN 12876 standard
  3. p. 1 table, 34 rows

Technical data according to DIN 12876 — Unistat TR401 (Order-No. 1028.0007.01)

ParameterValue
Operating temperature range50...400 °C
Resolution of display0,01 K
Temperature stability at 70°C0,05 K
temperature set point / display5,7" colour Touchscreen
Absolute accuracy2-point-calibration
Internal temperature sensorPt100
External sensorPt100
Interface digitalEthernet, USB (Host u. Device), RS232
digital inputECS ONE
digital outputPOKO ONE
Alarm messageoptic, acoustic, relay
Safety classificationIII / FL
Heating power at 240V3 kW
Heating power at 230V3 kW
Heating power at 220V2,7 kW
Heating power at 208V2,4 kW
Heating power at 200V2,2 kW
Circulation pump: max. delivery31 l/min
max. delivery pressure0,9 bar
Pump connenction flowM24x1,5
Pump connection returnM24x1,5
Cooling water connectionG1/2 male
min. filling capacity2,3 l
Volume of expansion5 l
Overall dimensions WxDxH288x379x890 mm
Pressure equipment categoryArt. 4.3 PED
Degree of ProtectionIP20
Net weight55 kg
Power supply requirement200-240V 1~/2~ 50/60Hz
max. current14 A
Fuse16 A
min. ambient temperature5 °C
max. ambient temperature40 °C
from Serial-No.S408106

Technical details and dimensions are subject to change. No liability is accepted for errors or omissions. Illustrations can deviate from the original.

Engineering Facts

12 facts · every claim linked to a page
Selection

The thermostat covers a working temperature range of 50 to 400 °C, so it must be matched to the process temperature window.

Defines whether the unit can handle the required process temperatures at all.

Source
“Operating temperature range 50...400 °C”
Installation

The unit requires a 200–240 V single- or two-phase supply with a maximum current of 14 A and a 16 A fuse.

Determines electrical supply, breaker sizing and cabling for the installation.

Source
“Power supply requirement 200-240V 1~/2~ 50/60Hz max. current 14 A Fuse 16 A”
Installation

The cooling water connection is a G1/2 male thread, which must be provided for in the site piping.

Needed to plan the cooling water supply and return connections before installation.

Source
“Cooling water connection G1/2 male”
Installation

The pump connections are M24x1.5 on both flow and return, so matching hoses and fittings must be ordered.

Prevents mismatch of process hoses and adapters during commissioning.

Source
“Pump connenction flow M24x1,5 Pump connection return M24x1,5”
Installation

The unit has IP20 protection, so it must be installed in a dry indoor location protected from water and dust ingress.

Restricts the acceptable installation environment and enclosure requirements.

Source
“Degree of Protection IP20”
Operation

The ambient temperature at the installation site must stay between 5 °C and 40 °C for correct operation.

Room conditions outside this range can affect performance and reliability.

Source
“min. ambient temperature 5 °C max. ambient temperature 40 °C”
Installation

The controller is connected to the machine by a cable of approximately 5 m, which affects controller placement.

Determines where the Pilot ONE controller can be mounted relative to the unit.

Source
“Length of cable between machine and controller approx. 5m.”
Safety

The unit has adjustable overtemperature protection according to DIN 12876 and safety classification III / FL.

Confirms built-in safety protection relevant for risk assessment and compliance.

Source
“With adjustable overtemperature protection according to DIN 12876.”
Conformité

External branch circuit protection according to UL 489 is required for this equipment.

Mandates a specific upstream protective device in the electrical installation.

Source
“External branch circuit protection according to UL 489 required.”
Installation

Voltage and frequency tolerances are ±10 % only if they do not run in opposite directions, so supply quality must be checked.

Prevents operation outside the allowed supply envelope that could damage the unit.

Source
“Voltage + / - 10%, as long as the frequency tolerance does not run in the opposite direction.”
Installation

Single- and two-phase devices are delivered with a power cable and country-specific plug that must be specified at order.

Ensures the correct plug type is supplied for the destination country.

Source
“Single / two-phase devices (100V to 240V) --> with power cable and country-specific plug (please specify when ordering)”
Conformité

The equipment is classified as Class A, Group 1 for electromagnetic disturbance to EN55011.

Class A equipment may require EMC measures in residential or sensitive environments.

Source
“Classification (disturbance) to EN55011: Class A, Group 1”